EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XT190Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA III BRONCHOVIDEOSCOPE (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III bronchoscope has 122 adverse event reports, with 2 deaths and 11 injuries. Many reports describe blank screens, burned or chipped plastic at the tip, and communication failures. Five open recalls name this device, mainly for fire risk during laser or high-frequency therapy, possible infection from wet channels, and the scope getting stuck in a connector. Reports rose from 28 in the previous year to 47 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
127
Deaths reported
2
Injuries reported
11
Recalls naming it
5

Summary

The FDA's MAUDE database contains 127 adverse event reports that list this exact device (DI 04953170402470), received between Mar 4, 2022 and Sep 25, 2026 — classified as 2 deaths, 11 injuries, 114 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (34 reports).

5 FDA recall records name this device's identifier, 5 still open.

3 other recalls cover devices cleared under the same FDA submission (K250862); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 4 reportsJan 2024: 3 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 10 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 6 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 4 reportsDec 2025: 7 reportsJan 2026: 6 reportsFeb 2026: 8 reportsMar 2026: 2 reportsApr 2026: 6 reportsMay 2026: 5 reportsJun 2026: 1 reportsJul 2026: 2 reportsAug 2026: 7 reportsSep 2026: 4 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2025Class IOpen, Classified

    Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.

    Z-0066-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Oct 12, 2023Class IOpen, Classified

    There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

    Z-0193-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0327-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2498-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  5. Jun 8, 2023Class IOpen, Classified

    There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.

    Z-2017-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Display/Image34
  • Thermal Decomposition of Device22
  • Communication or Transmission Problem14
  • Adverse Event Without Identified Device or Use Problem10
  • Detachment of Device or Device Component9
  • Melted6
  • Poor Quality Image6
  • Electrical /Electronic Property Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions115
  • Unspecified Infection6
  • Cardiac Arrest2
  • Hemorrhage/Blood Loss/Bleeding2
  • Pneumothorax2
  • Burn(s)1
  • Insufficient Information1
  • Laceration(s)1

Recent adverse event reports

The 25 most recent of 127 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Crack; No Display/Image

    The customer reported no image during reprocessing involving an Olympus Bronchovideoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897217 ↗
  • MalfunctionSep 23, 2026Corroded

    The customer reported the Bronchovideoscope had a separate issue. During the initial evaluation of the device, it was observed that the LG lens is corroded. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870946 ↗
  • MalfunctionSep 8, 2026Communication or Transmission Problem

    The customer returned the Bronchovideoscope for a separate issue. During the initial evaluation of the device, it was observed no communication between the plug unit and the processor. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50668179 ↗
  • MalfunctionSep 1, 2026Image Display Error/Artifact

    The customer reported the Bronchovideoscope when turned on in any direction lines were on the monitor during setup. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50535634 ↗
  • MalfunctionAug 28, 2026Break; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50503820 ↗
  • MalfunctionAug 27, 2026Thermal Decomposition of Device; Electrical /Electronic Property Problem; Material Fragmentation

    The customer returned the BronchoVideoscope for a separate issue. During the device analysis, it was observed that Distal end Plastic Cover burn at channel opening and Distal end Plastic Cover chip was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50479960 ↗
  • MalfunctionAug 26, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50458963 ↗
  • MalfunctionAug 24, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418497 ↗
  • MalfunctionAug 13, 2026Material Puncture/Hole

    The customer returned the Bronchovideoscope to the service center for a separate issue. During the device analysis, the service repair technician indicated that the distal end plastic cover was found chipped inside channel. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50237854 ↗
  • MalfunctionAug 5, 2026Thermal Decomposition of Device

    The customer returned the Bronchovideoscope to the service center for a separate issue. During the device analysis, the Service Repair Technician found that the plastic distal end cover was burned. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50141205 ↗
  • MalfunctionAug 4, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50134703 ↗
  • MalfunctionJul 31, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50105007 ↗
  • MalfunctionJul 28, 2026Thermal Decomposition of Device

    The customer returned the Bronchovideoscope to the service center for a separate issue. During the device analysis, the Service Repair Technician indicates that a burnt plastic distal end cover was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50056597 ↗
  • MalfunctionJun 19, 2026Thermal Decomposition of Device

    THE CUSTOMER RETURNED THE BRONCHOVIDEOSCOPE, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING INSPECTION IT WAS FOUND THAT THE DISTAL END PLASTIC COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25566336 ↗
  • MalfunctionMay 28, 2026Thermal Decomposition of Device; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE HAD A BURN AND A CHIP ON THE C-COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25323346 ↗
  • MalfunctionMay 28, 2026Thermal Decomposition of Device

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF MINOR LEAK. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, BURNT C-COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25322056 ↗
  • MalfunctionMay 4, 2026Electrical /Electronic Property Problem; Flare or Flash

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF IT JUST COLLAPSED. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, BURNT PLASTIC DISTAL END COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25067965 ↗
  • MalfunctionMay 1, 2026Obstruction of Flow

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF INTERNAL DAMAGE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, SUCTION FLOW WAS LOW DUE TO OBSTRUCTION WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056730 ↗
  • MalfunctionMay 1, 2026Material Fragmentation

    THE CUSTOMER DID NOT REPORT A MALFUNCTION. DURING ONSITE INSPECTION OF BRONCHOVIDEOSCOPE, PLASTIC DISTAL END COVER CHIPPED WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056404 ↗
  • MalfunctionApr 29, 2026Degraded; Image Display Error/Artifact

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE OBJECTIVE LENS HAD GLUE MISSING AND THE IMAGE HAD STATIC, WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25038411 ↗
  • MalfunctionApr 28, 2026Peeled/Delaminated

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF THAT THE TIP OF SCOPE WAS PEELING. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE BENDING RUBBER GLUE WAS MISSING WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25024128 ↗
  • MalfunctionApr 23, 2026Detachment of Device or Device Component

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF TIP OF SCOPE BROKEN. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, DISTAL END PLASTIC COVER IS PARTIALLY MISSING WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24967416 ↗
  • MalfunctionApr 3, 2026Break

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF LEAK. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, OBJECTIVE LENS HAS CHIPPED GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24779458 ↗
  • MalfunctionApr 2, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE HAD THE CAMERA DETACH FROM THE DISTAL END. THE ISSUE WAS OBSERVED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765491 ↗
  • MalfunctionApr 2, 2026Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED INTERMITTENT FLICKERING IN IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765079 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K250862. One submission often covers many models, so check the product and lot details in each record.

  • Z-0048-2026 · Sep 11, 2025 · Open, Classified
    BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100
  • Z-0057-2026 · Sep 11, 2025 · Open, Classified
    EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190
  • Z-0052-2026 · Sep 11, 2025 · Open, Classified
    BRONCHOVIDEOSCOPE OLYMPUS BF-H1100

Other Bronchoscope (flexible or rigid) devices

All Bronchoscope (flexible or rigid) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page