EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V (REF N5992940)Recalls & reported problems

Olympus Medical Systems Corp. · EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V · Class 2

All models of EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus duodenoscope is a flexible scope used to examine the duodenum. Reports most often describe cleaning failures, cracks and peeling, and some mention microbial contamination. The device appears in 4,675 reports, including 10 deaths and 499 injuries. It has 4 open recalls about cleaning and reprocessing, including water left in the channel after repair. Reports rose to 2,291 in the last 12 months from 321 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,900
Deaths reported
10
Injuries reported
498
Recalls naming it
4

Summary

The FDA's MAUDE database contains 4,900 adverse event reports that list this exact device (DI 04953170405563), received between Jan 6, 2021 and Sep 30, 2026 — classified as 10 deaths, 498 injuries, 4,392 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (955 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 93 reportsDec 2023: 80 reportsJan 2024: 223 reportsFeb 2024: 88 reportsMar 2024: 106 reportsApr 2024: 83 reportsMay 2024: 72 reportsJun 2024: 65 reportsJul 2024: 46 reportsAug 2024: 66 reportsSep 2024: 80 reportsOct 2024: 99 reportsNov 2024: 35 reportsDec 2024: 5 reportsJan 2025: 18 reportsFeb 2025: 11 reportsMar 2025: 18 reportsApr 2025: 15 reportsMay 2025: 10 reportsJun 2025: 22 reportsJul 2025: 19 reportsAug 2025: 26 reportsSep 2025: 54 reportsOct 2025: 88 reportsNov 2025: 117 reportsDec 2025: 249 reportsJan 2026: 283 reportsFeb 2026: 134 reportsMar 2026: 95 reportsApr 2026: 227 reportsMay 2026: 388 reportsJun 2026: 266 reportsJul 2026: 240 reportsAug 2026: 253 reportsSep 2026: 264 reportsOct 2026: 0 reports202420252026388
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 14, 2025Class IIOpen, Classified

    Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

    Z-0579-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. May 5, 2025Class IIOpen, Classified

    Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

    Z-1911-2025 · Root cause: Labeling design · Recall details · FDA record ↗

  3. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0346-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Jul 27, 2023Class IIOpen, Classified

    Recent reports of patient infection.

    Z-2467-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately955
  • Crack838
  • Peeled/Delaminated749
  • Detachment of Device or Device Component715
  • Microbial Contamination of Device337
  • Material Fragmentation288
  • Degraded251
  • Device Reprocessing Problem196

Patient problems

  • No Clinical Signs, Symptoms or Conditions4,617
  • Unspecified Tissue Injury88
  • Foreign Body In Patient84
  • Hemorrhage/Blood Loss/Bleeding27
  • Laceration(s)23
  • Unspecified Infection13
  • Bacterial Infection12
  • Laceration(s) of Esophagus11

Recent adverse event reports

The 25 most recent of 4,900 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer reported the duodenovideoscope had a separate issue. During the initial evaluation of the device, it was observed that objective and light guide lens had chipped glue. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950101 ↗
  • MalfunctionSep 30, 2026Degraded

    The customer returned the Duodenovideoscope to the service center for a annual maintenance. During the initial evaluation of the device, lifting of the lens adhesive was observed. No patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949265 ↗
  • MalfunctionSep 30, 2026Material Fragmentation

    The customer returned the duodenovideoscope to the service center for a separate issue. During the initial evaluation of the returned device chipped adhesive was found at the objective and light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949249 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer reported the duodenovideoscope, which is an Olympus asset, with no reported complaint. During the initial evaluation of the device, it was observed that there was peeling on the cement on the objective and light guide lens. There was no patient involvement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948951 ↗
  • MalfunctionSep 30, 2026Material Fragmentation

    The customer returned the duodenovideoscope to the service center for annual inspection. During the device analysis, chipped adhesive was found around the objective and light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948880 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component; Material Integrity Problem

    During the onsite inspection of the duodenovideoscope, areas of missing or damaged sealing agent were observed around the objective lens and the light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947942 ↗
  • MalfunctionSep 30, 2026Device Reprocessing Problem

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis a foreign material was found inside the forceps passage. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947916 ↗
  • MalfunctionSep 30, 2026Break; Failure to Clean Adequately

    The customer returned the duodenovideoscope for a separate issue. During the device evaluation, the air/water channel was found to have white residue. A cracked c-cover was also found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945608 ↗
  • MalfunctionSep 30, 2026Material Fragmentation

    During inspection of the Duodenovideoscope, chipped adhesive around the objective lens and light guide lens was observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945533 ↗
  • MalfunctionSep 30, 2026Crack

    The customer reported the Duodenovideoscope for a separate issue. During the initial evaluation of the returned device, chipped glue was observed on the objective (OB) lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945503 ↗
  • MalfunctionSep 30, 2026Crack

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis chipped glue was found at the objective lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945256 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer returned the Duodenovideoscope for a separate issue. During the initial evaluation of the device, the objective lens and the light guide lens had chipped adhesive. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944920 ↗
  • MalfunctionSep 29, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940744 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Crack

    The customer returned the Duodenovideoscope to the service center for a separate issue. During the device analysis, the service repair technician indicates light guide lens and objective lens had chip, deep scratches, and the adhesive around the lens had a gap and the adhesive around the objective and light guide lens had lifted off and chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936159 ↗
  • MalfunctionSep 29, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935708 ↗
  • MalfunctionSep 29, 2026Crack; Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935475 ↗
  • MalfunctionSep 29, 2026Crack; Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935318 ↗
  • MalfunctionSep 29, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934998 ↗
  • MalfunctionSep 29, 2026Crack; Material Fragmentation

    The customer returned the duodenovideoscope, which is an Olympus asset with no reported complaint. During the initial evaluation of the returned device, the light guide lens and objective lens had chipped glue, and a chipped lens was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933179 ↗
  • MalfunctionSep 29, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932156 ↗
  • MalfunctionSep 29, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931987 ↗
  • MalfunctionSep 29, 2026Crack; Material Fragmentation

    The customer returned the Duodenovideoscope to the service center for a separate issue. During initial evaluation of the device, the objective lens and the light guide lens had chipped glue and chipped lenses. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927600 ↗
  • MalfunctionSep 29, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926948 ↗
  • MalfunctionSep 29, 2026Material Fragmentation; Naturally Worn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926864 ↗
  • MalfunctionSep 28, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926708 ↗

Other Duodenoscope and accessories, flexible/rigid devices

All Duodenoscope and accessories, flexible/rigid devices →