EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V (REF N5992941)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III duodenoscope is a flexible endoscope. Its 366 reports include 314 malfunctions and 52 injuries, with no deaths. Many describe cleaning that fails, parts that detach, microbial contamination and cracks, and 13 mention a foreign body in the patient. Reports rose from 28 in the 12 months before to 102 in the latest 12 months. One recall is open, begun in October 2025, over updated reprocessing materials after continued reports of positive cultures and infections.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
381
Deaths reported
0
Injuries reported
52
Recalls naming it
1

Summary

The FDA's MAUDE database contains 381 adverse event reports that list this exact device (DI 04953170452024), received between Nov 10, 2021 and Sep 29, 2026 — classified as 52 injuries, 329 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (100 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 18 reportsDec 2023: 17 reportsJan 2024: 9 reportsFeb 2024: 4 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 5 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 5 reportsSep 2024: 9 reportsOct 2024: 7 reportsNov 2024: 5 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 1 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 8 reportsFeb 2026: 3 reportsMar 2026: 5 reportsApr 2026: 11 reportsMay 2026: 14 reportsJun 2026: 17 reportsJul 2026: 19 reportsAug 2026: 19 reportsSep 2026: 18 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 14, 2025Class IIOpen, Classified

    Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

    Z-0579-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately100
  • Detachment of Device or Device Component68
  • Microbial Contamination of Device60
  • Crack33
  • Material Fragmentation21
  • Device Reprocessing Problem17
  • Contamination16
  • Peeled/Delaminated15

Patient problems

  • No Clinical Signs, Symptoms or Conditions348
  • Foreign Body In Patient13
  • Airway Obstruction5
  • Unspecified Infection5
  • Bacterial Infection3
  • Insufficient Information3
  • Device Embedded In Tissue or Plaque2
  • Fungal Infection2

Recent adverse event reports

The 25 most recent of 381 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Crack; Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935674 ↗
  • MalfunctionSep 22, 2026Crack

    The customer returned the duodenovideoscope for a separate issue. During the initial evaluation of the returned device, Objective and light guide lens had crack glue was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50852613 ↗
  • MalfunctionSep 22, 2026Material Fragmentation

    During inspection of the Duodenovideoscope, chipping of the adhesive around the objective lens and the light guide lens was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50851955 ↗
  • MalfunctionSep 21, 2026Crack; Peeled/Delaminated; Material Puncture/Hole

    The customer returned the Duodenovideoscope to the service center for a separate issue. During the device analysis, the service repair technician indicates the objective and light guide lens had pinholes on the glue and peeling on the cement and the glue was missing from the light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50843847 ↗
  • MalfunctionSep 18, 2026Detachment of Device or Device Component

    The customer returned the duodenovideoscope, which is an Olympus asset with no reported complaint. During the initial evaluation of the returned device, Objective lens and light guide lens had chipped adhesive was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820394 ↗
  • MalfunctionSep 18, 2026Material Fragmentation

    The customer reported that the Duodenovideoscope had a separate issue. During the initial evaluation of the device, chipped adhesive around the objective lens was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820327 ↗
  • MalfunctionSep 18, 2026Degraded

    During the initial evaluation of the Duodenovideoscope device, it was observed that the Distal tip rubber adhesive was irregular. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50817886 ↗
  • MalfunctionSep 17, 2026Material Fragmentation

    The customer reported the Duodenovideoscope had an irregular adhesive on the distal tip rubber. The event occurred during maintenance. During the evaluation of the returned device, adhesive on the rubber (tip rubber) has a chip was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808839 ↗
  • MalfunctionSep 17, 2026Detachment of Device or Device Component

    During onsite inspection of duodenovideoscope, it was found missing adhesive at objective lens and light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808706 ↗
  • MalfunctionSep 16, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50796522 ↗
  • MalfunctionSep 10, 2026Corroded; Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721151 ↗
  • MalfunctionSep 10, 2026Flaked

    The customer returned the duodenovideoscope, which is an Olympus asset with no reported complaint. During the initial evaluation of the device, there was missing glue around the light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50714056 ↗
  • MalfunctionSep 9, 2026Residue After Decontamination; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50696337 ↗
  • MalfunctionSep 9, 2026Corroded; Degraded

    The customer reported the duodenovideoscope had uneven adhesive on distal tip during maintenance. During the initial evaluation of the device, it was observed there was corrosion in the suction chamber. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50694999 ↗
  • MalfunctionSep 8, 2026Material Protrusion/Extrusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50674942 ↗
  • MalfunctionSep 1, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50554082 ↗
  • MalfunctionSep 1, 2026Detachment of Device or Device Component; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50554064 ↗
  • MalfunctionSep 1, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50553795 ↗
  • MalfunctionAug 31, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50532153 ↗
  • MalfunctionAug 31, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50532053 ↗
  • MalfunctionAug 26, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459549 ↗
  • MalfunctionAug 26, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50454455 ↗
  • MalfunctionAug 21, 2026Crack; Peeled/Delaminated

    No additional information received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50379365 ↗
  • MalfunctionAug 18, 2026Crack

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis, there was chipped glue on the objective lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50299240 ↗
  • MalfunctionAug 18, 2026Degraded; Component Missing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50299146 ↗

Other Duodenoscope And Accessories, Flexible/Rigid devices

All Duodenoscope And Accessories, Flexible/Rigid devices →