EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190VRecalls & reported problems across all models
Olympus Medical Systems Corp. · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
4 open recalls name at least one EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5,281
- Deaths reported
- 10
- Injuries reported
- 550
- Recalls
- 4
We found 2 models of EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 5,281 adverse event reports in the FDA's MAUDE database, received between Jan 6, 2021 and Sep 30, 2026 — classified as 10 deaths, 550 injuries, 4,721 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (1,055 reports).
4 FDA recall records name at least one of these models by its device identifier, 4 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 14, 2025Class IIOpen, Classified
Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
Z-0579-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗
- May 5, 2025Class IIOpen, Classified
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
Z-1911-2025 · Root cause: Labeling design · Recall details · FDA record ↗
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0346-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 27, 2023Class IIOpen, Classified
Recent reports of patient infection.
Z-2467-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V problems
Device problems
- Failure to Clean Adequately1,055
- Crack871
- Detachment of Device or Device Component783
- Peeled/Delaminated764
- Microbial Contamination of Device397
- Material Fragmentation309
- Degraded265
- Device Reprocessing Problem213
Patient problems
- No Clinical Signs, Symptoms or Conditions4,965
- Foreign Body In Patient97
- Unspecified Tissue Injury89
- Hemorrhage/Blood Loss/Bleeding28
- Laceration(s)23
- Unspecified Infection18
- Bacterial Infection15
- Insufficient Information12
EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190VDuodenoscope and accessories, flexible/rigid4 recalls10 deaths4,900 reports
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190VDuodenoscope And Accessories, Flexible/Rigid1 recall381 reports
Recent EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Detachment of Device or Device Component
The customer reported the duodenovideoscope had a separate issue. During the initial evaluation of the device, it was observed that objective and light guide lens had chipped glue. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950101 ↗MalfunctionSep 30, 2026Degraded
The customer returned the Duodenovideoscope to the service center for a annual maintenance. During the initial evaluation of the device, lifting of the lens adhesive was observed. No patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949265 ↗MalfunctionSep 30, 2026Material Fragmentation
The customer returned the duodenovideoscope to the service center for a separate issue. During the initial evaluation of the returned device chipped adhesive was found at the objective and light guide lens. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949249 ↗MalfunctionSep 30, 2026Detachment of Device or Device Component
The customer reported the duodenovideoscope, which is an Olympus asset, with no reported complaint. During the initial evaluation of the device, it was observed that there was peeling on the cement on the objective and light guide lens. There was no patient involvement
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948951 ↗MalfunctionSep 30, 2026Material Fragmentation
The customer returned the duodenovideoscope to the service center for annual inspection. During the device analysis, chipped adhesive was found around the objective and light guide lens. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948880 ↗MalfunctionSep 30, 2026Detachment of Device or Device Component; Material Integrity Problem
During the onsite inspection of the duodenovideoscope, areas of missing or damaged sealing agent were observed around the objective lens and the light guide lens. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947942 ↗MalfunctionSep 30, 2026Device Reprocessing Problem
The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis a foreign material was found inside the forceps passage. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947916 ↗MalfunctionSep 30, 2026Break; Failure to Clean Adequately
The customer returned the duodenovideoscope for a separate issue. During the device evaluation, the air/water channel was found to have white residue. A cracked c-cover was also found. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945608 ↗MalfunctionSep 30, 2026Material Fragmentation
During inspection of the Duodenovideoscope, chipped adhesive around the objective lens and light guide lens was observed. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945533 ↗MalfunctionSep 30, 2026Crack
The customer reported the Duodenovideoscope for a separate issue. During the initial evaluation of the returned device, chipped glue was observed on the objective (OB) lens. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945503 ↗MalfunctionSep 30, 2026Crack
The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis chipped glue was found at the objective lens. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945256 ↗MalfunctionSep 30, 2026Detachment of Device or Device Component
The customer returned the Duodenovideoscope for a separate issue. During the initial evaluation of the device, the objective lens and the light guide lens had chipped adhesive. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944920 ↗MalfunctionSep 29, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940744 ↗MalfunctionSep 29, 2026Loss of or Failure to Bond; Crack
The customer returned the Duodenovideoscope to the service center for a separate issue. During the device analysis, the service repair technician indicates light guide lens and objective lens had chip, deep scratches, and the adhesive around the lens had a gap and the adhesive around the objective and light guide lens had lifted off and chipped. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936159 ↗MalfunctionSep 29, 2026Peeled/Delaminated
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935708 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- HD CAMERA HEAD2 models · 5,217 reports
- 4K CAMERA HEAD3 models · 4,926 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page