VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE V (REF N6007430)Recalls & reported problems
Olympus Medical Systems Corp. · VISERA HYSTEROVIDEOSCOPE · Class 2
All models of VISERA HYSTEROVIDEOSCOPE (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 04953170411564), received between Mar 9, 2023 and Jul 8, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (10 reports).
2 other recalls cover devices cleared under the same FDA submission (K221557); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component10
- Peeled/Delaminated3
- Material Split, Cut or Torn2
- Degraded1
- Gas/Air Leak1
- Material Deformation1
- Material Separation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE LIGHT GUIDE LENS ADHESIVE WAS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25753277 ↗MalfunctionJun 25, 2026Detachment of Device or Device Component
THE CUSTOMER RETURNED THE HYSTEROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE LIGHT GUIDE LENS HAD CHIPPED CEMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25631316 ↗MalfunctionJun 22, 2026Peeled/Delaminated
IT WAS REPORTED THE VIDEOSCOPE BLACK PLASTIC COTING WAS COMING OFF/PEELED OFF OF THE INSERTION TUBE. THE ISSUE WAS DISCOVERED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25585449 ↗MalfunctionMay 26, 2026Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTERO VIDEOSCOPE EXHIBITED A CHIP ON LIGHT GUIDE LENS ADHESIVE, AND RUST ON THE COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25280222 ↗MalfunctionMay 19, 2026Material Separation; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROVIDEOSCOPE EXHIBITED A DETACHED OBJECTIVE LENS, AND MISSING ADHESIVE FROM THE BENDING SECTION COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25217562 ↗MalfunctionMay 13, 2026Degraded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CUSTOMER'S VIDEOSCOPE DEVICE HAD CHIPPED BENDING SECTION COVER ADHESIVE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25162263 ↗MalfunctionMay 6, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTERO VIDEOSCOPE HAD A CHIPPED DISTAL SHEATH ADHESIVE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25095202 ↗MalfunctionMar 24, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTERO VIDEOSCOPE EXHIBITED DETACHED BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24679488 ↗MalfunctionNov 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROVIDEOSCOPE EXHIBITED MISSING ADHESIVE AROUND LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23647753 ↗MalfunctionNov 13, 2025Detachment of Device or Device Component; Gas/Air Leak; Material Deformation; Material Split, Cut or Torn; Detachment of Device or Device Component; Gas/Air Leak; Material Deformation; Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED A LEAK ON DISTAL END (A-RUBBER) CAUSED BY A CUT AND DETACHED BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23539439 ↗MalfunctionOct 31, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE EXHIBITED EXCESSIVE PEELING AT SHEATH ON INSERTION TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23437331 ↗MalfunctionOct 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED MISSING BENDING SECTION COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426650 ↗MalfunctionOct 27, 2025Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VISERA HYSTEROVIDEOSCOPE EXHIBITED A PEELING AND MISSING ADHESIVE ON THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23391806 ↗MalfunctionOct 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VISERA HYSTEROVIDEOSCOPE HAD A MISSING BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374670 ↗MalfunctionMar 9, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO OLYMPUS, THAT DURING AN UNSPECIFIED PROCEDURE THE VISERA HYSTEROVIDEOSCOPE INSERTION SECTION WAS DAMAGED. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING EVALUATION OF THE RETURNED DEVICE, THE EVALUATION FOUND THAT THE FORCEP STOPPER CAP WAS SCRAPING. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16518920 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221557. One submission often covers many models, so check the product and lot details in each record.
- Z-0336-2024 · Sep 25, 2023 · Open, ClassifiedHysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
- Z-2480-2023 · Jul 20, 2023 · Open, ClassifiedVisera Hysterovideoscope Olympus HYF Type V
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page