OES HYSTEROFIBERSCOPE OLYMPUS HYF TYPE XP (UDI 04953170411571)Recalls & reported problems
Olympus Medical Systems Corp. · OES HYSTEROFIBERSCOPE · Class 2
All models of OES HYSTEROFIBERSCOPE (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 04953170411571), received between Oct 30, 2020 and Nov 17, 2025 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Microbial Contamination of Device” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Microbial Contamination of Device8
- Failure to Clean Adequately3
- Device Reprocessing Problem2
- Collapse1
- Degraded1
- Material Deformation1
- Material Discolored1
- Material Protrusion/Extrusion1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 17, 2025Collapse; Collapse
IT WAS REPORTED THE SUBJECT DEVICE DISTAL END "EXPLODED". THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23575855 ↗MalfunctionAug 18, 2025Degraded; Material Discolored; Degraded; Material Discolored
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SCOPE EXHIBITED CORROSION ON THE DISTAL END AND FORCEPS CHANNEL PORT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22826014 ↗MalfunctionOct 15, 2024Material Protrusion/Extrusion; Material Twisted/Bent; Material Protrusion/Extrusion; Material Twisted/Bent
IT WAS REPORTED, THE VIDEOSCOPE HAD A DEFORMATION AND PROTRUSION NEAR THE CURVED PART. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20455172 ↗MalfunctionOct 9, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE FLEXIBLE SCOPE EXHIBITED THE TIP OF THE COATING WAS PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20413260 ↗MalfunctionOct 8, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE HYSTEROFIBERSCOPE EXHIBITED FOREIGN MATERIAL IN THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20400347 ↗MalfunctionAug 8, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE HYSTEROFIBERSCOPE CONNECTING TUBE AND A-RUBBER (BENDING SECTION COVER) HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19945708 ↗MalfunctionJun 6, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SCOPE EXHIBITED FOREIGN MATERIAL ON THE BENDING SECTION COVER AND THE ADHESIVE OF THE BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19476627 ↗MalfunctionJun 4, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT THE HYSTEROFIBERSCOPE, IT WAS DIFFICULT TO ATTACH THE CAP TO THE STERILIZATION NOZZLE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19459668 ↗MalfunctionMar 18, 2024Material Deformation; Material Deformation
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROFIBERSCOPE FORCEPS CHANNEL PORT WAS DEFORMED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18921606 ↗MalfunctionFeb 3, 2024Microbial Contamination of Device; Microbial Contamination of Device
IT WAS REPORTED THAT, DURING REPROCESSING, THE HYSTEROFIBERSCOPE TESTED POSITIVE FOR 1 COLONY FORMING UNITS (CFU) OF STAPHYLOCOCCUS HOMINIS ON (B)(6) 2023. THE DEVICE WAS TESTED POSITIVE AGAIN ON (B)(6) 2024, FOR 5 COLONY FORMING UNITS (CFUS) OF NEISSERIA SPP. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18637347 ↗MalfunctionDec 12, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, DURING REPROCESSING, THE HYSTEROFIBERSCOPE TESTED POSITIVE FOR 1 COLONY FORMING UNITS (CFUS) OF UNSPECIFIED MICROORGANISMS. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18313311 ↗MalfunctionOct 23, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED THAT THE OES HYSTEROFIBERSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUES WERE FOUND DURING REGULAR EXAMINATION IN THE HOSPITAL. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17985256 ↗MalfunctionAug 28, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, DURING REPROCESSING, THE OES HYSTEROFIBERSCOPE TESTED POSITIVE FOR 1 COLONY FORMING UNITS (CFUS) OF GRAM-POSITIVE BACTERIA. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17630867 ↗MalfunctionJan 17, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE USER DID NOT SHARE THEIR NON-CONFORMING CULTURE RESULTS HOWEVER, THE SUBJECT DEVICE WAS REPROCESSED AND RE-CULTURE TESTED ON 06APRIL2022 BY THE USER FACILITY WHERE THE RESULTS WERE <1 CFU IN ALL CHANNELS AND CONFORMED TO THE REGULATION'S RECOMMENDATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16185380 ↗MalfunctionNov 18, 2022Microbial Contamination of Device; Microbial Contamination of Device
AS REPORTED BY THE CUSTOMER, AFTER A MICROBIOLOGICAL ROUTINE SAMPLING OF THIS DEVICE, CARRIED OUT AS REQUIRED BY FRENCH REGULATION, UNEXPECTED CONTAMINATION WAS DETECTED WHICH WAS ABOVE THE ACCEPTABLE THRESHOLD FOR ALL CHANNELS. THE TEST WAS PERFORMED PRIOR TO USE. THERE WAS NO CONTAMINATION OR ANY OTHER DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE. PER CUSTOMER HYGIENE MICROBIOLOGICAL INVESTIGATION (HMI) SAMPLING TAKEN, THE MICROBIAL RESULTS OBTAINED ARE AS FOLLOWS: DATE OF ANALYSIS (B)(6) 2022, SAMPLE TAKEN (B)(6) 2022: MESOPHILIC AEROBIC FLORA IS GREATER THAN 80 CFU/VOLUME FOR ALL CHANNELS. DECLARATION OF CONFORMITY: THE SAMPLE PRESENTED DURING ITS SAMPLING CHARACTERISTICS THAT DID NOT COMPLY WITH THE GUIDES AND RECOMMENDATIONS OF THE MINISTRY OF HEALTH. DATE OF ANALYSIS (B)(6) 2022, SAMPLE TAKEN (B)(6) 2022: MESOPHILIC AEROBIC FLORA IS 27 CFU/VOLUME FOR ALL CHANNELS. DATE OF ANALYSIS (B)(6) 2022, SAMPLE TAKEN (B)(6) 2022: MESOPHILIC AEROBIC FLORA IS 7 CFU/VOLUME FOR ALL CHANNELS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15824267 ↗MalfunctionJan 12, 2022Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, THE OES HYSTEROFIBERSCOPE TESTED POSITIVE FOR CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. ALL CHANNELS WERE SAMPLED. THE SAMPLING OCCURRED DURING REPROCESSING, BEFORE USE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13237238 ↗MalfunctionNov 20, 2020Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MULTIPLE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE DEVICE. FIRST TIME; (B)(6) 2020. UNSPECIFIED MICROBES. SECOND TIME; (B)(6) 2020. UNSPECIFIED MICROBES. THIRD TIME; (B)(6) 2020. UNSPECIFIED MICROBES. FOURTH TIME; (B)(6) 2020. UNSPECIFIED MICROBES. THE DEVICE HAD BEEN MANUALLY REPROCESSED USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10873622 ↗MalfunctionOct 30, 2020Device Reprocessing Problem; Pressure Problem; Device Reprocessing Problem; Pressure Problem
THE SERVICE CENTER WAS INFORMED THAT DURING REPROCESSING, TINY SPECKS ARE SHOWING UP ON THE LENS OF THE SCOPE AND FOREIGN MATERIAL IS FALLING OFF FROM THE CHANNEL. THERE WAS A SMALL LEAK OBSERVED DURING AN AIR LEAK TEST. NO PATIENT INJURY OR HARM WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10766548 ↗
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page