GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1100Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of GASTROINTESTINAL VIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus colonoscope is the subject of 1,053 adverse event reports, which mention no deaths and 11 injuries. Many reports describe scopes that were not adequately cleaned or left residue, and others describe poor or missing images. Reports rose from 96 in the previous 12 months to 670 in the latest 12 months. 1 open recall, started in August 2024, concerns camera color settings programmed incorrectly.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
990
Deaths reported
0
Injuries reported
11
Recalls naming it
1

Summary

The FDA's MAUDE database contains 990 adverse event reports that list this exact device (DI 04953170420191), received between May 9, 2023 and Sep 30, 2026 — classified as 11 injuries, 979 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (407 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 18 reportsDec 2023: 10 reportsJan 2024: 18 reportsFeb 2024: 19 reportsMar 2024: 22 reportsApr 2024: 15 reportsMay 2024: 22 reportsJun 2024: 17 reportsJul 2024: 19 reportsAug 2024: 27 reportsSep 2024: 24 reportsOct 2024: 34 reportsNov 2024: 12 reportsDec 2024: 1 reportsJan 2025: 8 reportsFeb 2025: 9 reportsMar 2025: 14 reportsApr 2025: 11 reportsMay 2025: 11 reportsJun 2025: 6 reportsJul 2025: 7 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 13 reportsNov 2025: 39 reportsDec 2025: 159 reportsJan 2026: 91 reportsFeb 2026: 45 reportsMar 2026: 52 reportsApr 2026: 59 reportsMay 2026: 63 reportsJun 2026: 42 reportsJul 2026: 22 reportsAug 2026: 4 reportsSep 2026: 31 reportsOct 2026: 0 reports202420252026159
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 21, 2024Class IIOpen, Classified

    It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.

    Z-0181-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately407
  • Communication or Transmission Problem153
  • Residue After Decontamination95
  • No Display/Image88
  • Poor Quality Image80
  • Device Reprocessing Problem63
  • Microbial Contamination of Device40
  • Thermal Decomposition of Device29

Patient problems

  • No Clinical Signs, Symptoms or Conditions977
  • Insufficient Information5
  • Foreign Body In Patient4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Fungal Infection2
  • Drug Resistant Bacterial Infection1

Recent adverse event reports

The 25 most recent of 990 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Communication or Transmission Problem

    The customer returned the gastrointestinal videoscope to the service center for a separate issue. During the device analysis an E226 error was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952101 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951622 ↗
  • MalfunctionSep 30, 2026Communication or Transmission Problem

    The customer reported the Gastrointestinal videoscope scope communication error (E226) during a Inspection for use. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947966 ↗
  • MalfunctionSep 30, 2026Erratic or Intermittent Display; No Display/Image

    The customer reported the gastrointestinal videoscope had a flickering display during inspection for use. During the initial evaluation of the device, a blackout image and no image was observed. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945851 ↗
  • MalfunctionSep 30, 2026Communication or Transmission Problem; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941472 ↗
  • MalfunctionSep 29, 2026No Display/Image; Poor Quality Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939526 ↗
  • MalfunctionSep 29, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934777 ↗
  • MalfunctionSep 29, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931692 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924923 ↗
  • MalfunctionSep 28, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923435 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the gastrointestinal videoscope for a separate issue. No patient involvement. During inspection of the returned device, it was found white residue in the air/water supply channels.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922763 ↗
  • MalfunctionSep 28, 2026Corroded

    The customer returned the gastrointestinal videoscope for a separate issue. No patient involvement. During inspection of the returned device, it was found the air/water port inlet was rusty.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921823 ↗
  • MalfunctionSep 28, 2026Erratic or Intermittent Display; No Display/Image; Poor Quality Image

    The customer reported Black image and image Intermittent during the reprocessing involving an Olympus Gastrointestinal videoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50913873 ↗
  • MalfunctionSep 24, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890361 ↗
  • MalfunctionSep 24, 2026Communication or Transmission Problem

    The customer returned the Gastrointestinal Videoscope for a separate issue. During the initial evaluation of the returned device, scope communication error code E226 was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884986 ↗
  • MalfunctionSep 24, 2026Failure to Clean Adequately

    The customer reported that the gastrointestinal videoscope had a separate issue. During the initial evaluation of the device, white residues were observed on both sides of the air/water channel. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884280 ↗
  • MalfunctionSep 24, 2026Failure to Clean Adequately

    The customer reported that the gastrointestinal videoscope had a separate issue. During the initial evaluation of the device, white residue was observed in the air/water channel on both sides. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883591 ↗
  • MalfunctionSep 24, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883208 ↗
  • MalfunctionSep 24, 2026Material Integrity Problem

    The customer reported that the gastrointestinal videoscope exhibited fibers protruding from the channel. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883037 ↗
  • MalfunctionSep 24, 2026Communication or Transmission Problem; Failure to Clean Adequately

    The customer reported an E226 communication error during inspection for an unspecified procedure involving an Olympus gastrointestinal videoscope. Additionally, during the initial evaluation of the returned device a white residue was found at the air/water supply and auxiliary water tube. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882722 ↗
  • MalfunctionSep 24, 2026Failure to Clean Adequately

    The customer reported the gastrointestinal videoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed in the air/water supply cylinder and channel. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882700 ↗
  • MalfunctionSep 23, 2026Poor Quality Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876213 ↗
  • MalfunctionSep 22, 2026Residue After Decontamination; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862145 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50861483 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860784 ↗

Other Colonoscope And Accessories, Flexible/Rigid devices