Skip to content

COLONOVIDEOSCOPE OLYMPUS CF-HQ1100DIRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of COLONOVIDEOSCOPE (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus colonoscope is an endoscope used to look inside the colon. Most reports describe the scope not being cleaned adequately, and some describe microbial contamination inside it or foreign objects found in its air and water tubes during evaluation. There are 260 reports, with no deaths and no injuries. Reports fell to 1 in the last 12 months from 16 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
261
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 261 adverse event reports that list this exact device (DI 04953170424298), received between Jun 20, 2023 and Sep 8, 2026 — classified as 261 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (188 reports).

1 other recall cover devices cleared under the same FDA submission (K222584); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 14 reportsJan 2024: 22 reportsFeb 2024: 24 reportsMar 2024: 9 reportsApr 2024: 13 reportsMay 2024: 9 reportsJun 2024: 15 reportsJul 2024: 19 reportsAug 2024: 14 reportsSep 2024: 18 reportsOct 2024: 31 reportsNov 2024: 8 reportsDec 2024: 3 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202631
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately188
  • Microbial Contamination of Device58
  • Device Reprocessing Problem15
  • Communication or Transmission Problem6
  • Thermal Decomposition of Device6
  • Contamination /Decontamination Problem3
  • Partial Blockage2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions261

Recent adverse event reports

The 25 most recent of 261 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2026Communication or Transmission Problem

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, B30 (scope communication error). No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50675747 ↗
  • MalfunctionJul 30, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50080545 ↗
  • MalfunctionMar 10, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21569051 ↗
  • MalfunctionFeb 5, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE AIR/WATER CYLINDER, AIR/WATER TUBE, AND IN THE JET TUBE (J-TUBE). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21315607 ↗
  • MalfunctionJan 30, 2025Device Reprocessing Problem; Obstruction of Flow; Device Reprocessing Problem; Obstruction of Flow

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE AIR/WATER CYLINDER, TUBE AND AUXILIARY JET TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21277796 ↗
  • MalfunctionJan 17, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE FOLLOWING: THERE WAS FOREIGN MATERIAL IN AIR/WATER CYLINDER AND TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21181592 ↗
  • MalfunctionJan 16, 2025Microbial Contamination of Device; Microbial Contamination of Device

    THE SUBJECT DEVICE TESTED POSITIVE FOR 215 COLONY FORMING UNITS (CFUS) OF ENTEROBACTERIA, PSEUDOMONAS, AND STAPHYLOCOCCUS AUREUS. ALL CHANNELS WERE SAMPLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21169733 ↗
  • MalfunctionDec 13, 2024Partial Blockage; Partial Blockage

    NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20927631 ↗
  • MalfunctionDec 6, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20865736 ↗
  • MalfunctionDec 3, 2024Partial Blockage; Failure to Clean Adequately; Partial Blockage; Failure to Clean Adequately

    NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20834239 ↗
  • MalfunctionNov 22, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT, DURING REPROCESSING, THE SCOPE TESTED POSITIVE FOR 17 COLONY FORMING UNITS (CFUS) OF PSEUDOMONAS AERUGINOSA. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20758856 ↗
  • MalfunctionNov 21, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT, DURING REPROCESSING, THE COLONOVIDEOSCOPE TESTED POSITIVE FOR MORE THAN 100 COLONY FORMING UNITS (CFUS) OF ENTEROBACTER SP.. AND KLEBSIELLA SP.. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20752061 ↗
  • MalfunctionNov 14, 2024Defective Device; Defective Device

    IT WAS REPORTED, THE COLONOVIDEOSCOPE ASSET CHECK FAILED REQUIRING REPAIR. THE ISSUE OCCURRED DURING PREVENTIVE MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20687959 ↗
  • MalfunctionNov 11, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE COLONOVIDEOSCOPE EXHIBITED NOZZLE HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20656462 ↗
  • MalfunctionNov 8, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE JET TUBE OF THE COLONOVIDEOSCOPE WAS CLOGGED WITH FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20639312 ↗
  • MalfunctionNov 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED A CLOGGED NOZZLE BY A DARK RUBBERY MATERIAL THAT COULDN'T BE REMOVED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20633053 ↗
  • MalfunctionNov 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE COLONOVIDEOSCOPE EXHIBITED A WHITISH FOREIGN MATERIAL WAS FOUND INSIDE OF THE AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20632029 ↗
  • MalfunctionNov 6, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS WITHIN THE AIR/WATER TUBE AND CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20616439 ↗
  • MalfunctionOct 31, 2024Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED HAVING USED THE COLONOVIDEOSCOPE ON SIX PATIENTS WHILE AWAITING THE RESULTS FOR ROUTINE CULTURE TESTING. POSITIVE CULTURE TEST RESULTS WERE SUBSEQUENTLY RECEIVED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20582983 ↗
  • MalfunctionOct 31, 2024Microbial Contamination of Device; Failure to Clean Adequately; Microbial Contamination of Device; Failure to Clean Adequately

    CORRECTION TO B5 OF THE INITIAL REPORT: THE CUSTOMER REPORTED HAVING USED THE COLONOVIDEOSCOPE ON 6 OF PATIENTS WHILE AWAITING THE RESULTS FOR ROUTINE CULTURE TESTING. POSITIVE CULTURE TEST RESULTS WERE SUBSEQUENTLY RECEIVED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20579323 ↗
  • MalfunctionOct 31, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE AIR-WATER TUBE, AND FOREIGN OBJECTS IN THE JET-TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20577214 ↗
  • MalfunctionOct 30, 2024Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED HAVING USED THE COLONOVIDEOSCOPE ON SIX PATIENTS WHILE AWAITING THE RESULTS FOR ROUTINE CULTURE TESTING. POSITIVE CULTURE TEST RESULTS WERE SUBSEQUENTLY RECEIVED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20571333 ↗
  • MalfunctionOct 30, 2024Microbial Contamination of Device; Failure to Clean Adequately; Microbial Contamination of Device; Failure to Clean Adequately

    THE CUSTOMER REPORTED HAVING USED THE COLONOVIDEOSCOPE ON 6 OF PATIENTS WHILE AWAITING THE RESULTS FOR ROUTINE CULTURE TESTING. POSITIVE CULTURE TEST RESULTS WERE SUBSEQUENTLY RECEIVED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20566496 ↗
  • MalfunctionOct 30, 2024Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED HAVING USED THE COLONOVIDEOSCOPE ON SIX PATIENTS WHILE AWAITING THE RESULTS FOR ROUTINE CULTURE TESTING. POSITIVE CULTURE TEST RESULTS WERE SUBSEQUENTLY RECEIVED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20565712 ↗
  • MalfunctionOct 29, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT, DURING REPROCESSING, THE COLONOVIDEOSCOPE TESTED POSITIVE FOR 6 COLONY FORMING UNITS (CFUS) OF PSEUDOMONAS AERUGINOSA. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20558573 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K222584. One submission often covers many models, so check the product and lot details in each record.

  • Z-0181-2025 · Aug 21, 2024 · Open, Classified
    Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100

Other Colonoscope And Accessories, Flexible/Rigid devices

All Colonoscope And Accessories, Flexible/Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page