Skip to content

COLONOVIDEOSCOPE OLYMPUS CF-EZ1500D SERIES (REF N6022440)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of COLONOVIDEOSCOPE (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170425103), received between May 23, 2026 and Jul 27, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem1
  • Failure to Clean Adequately1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50018432 ↗
  • MalfunctionMay 23, 2026Failure to Clean Adequately

    THE COLONOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF INTERNAL BLOCKAGE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, WHITE RESIDUE AROUND NOZZLE OPENING WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25265159 ↗

Other Colonoscope And Accessories, Flexible/Rigid devices

All Colonoscope And Accessories, Flexible/Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page