Quick summary

This is an Olympus EVIS EXERA III colonoscope, a flexible endoscope. Many of the 84 reports describe the scope not being cleaned adequately during reprocessing, while others describe no image, communication errors, stained images and foreign objects found during checks. The reports include no deaths and no injuries, and none describe patient involvement. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
86
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 86 adverse event reports that list this exact device (DI 04953170436659), received between May 5, 2023 and Sep 29, 2026 — classified as 86 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (51 reports).

8 other recalls cover devices cleared under the same FDA submission (K131780); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 8 reportsJan 2024: 5 reportsFeb 2024: 8 reportsMar 2024: 8 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 8 reportsSep 2024: 6 reportsOct 2024: 7 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately51
  • Communication or Transmission Problem12
  • Device Reprocessing Problem9
  • Microbial Contamination of Device4
  • No Display/Image4
  • Break2
  • Poor Quality Image2
  • Contamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions86

Recent adverse event reports

The 25 most recent of 86 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026No Display/Image; Poor Quality Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941334 ↗
  • MalfunctionSep 18, 2026No Display/Image; Communication or Transmission Problem

    It was reported that during a diagnostic colonoscopy, the colonovideoscope experienced a B30 error message and the image disappeared. The reported event resulted in a procedural prolongation of less than 30 minutes. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50826888 ↗
  • MalfunctionDec 27, 2025No Display/Image; Communication or Transmission Problem; No Display/Image; Communication or Transmission Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23903259 ↗
  • MalfunctionDec 8, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD AN E237 ERROR (SCOPE ERROR) AND AN E226 ERROR (SCOPE COMMUNICATION ERROR). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23743527 ↗
  • MalfunctionNov 12, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NOT DISPLAY IMAGE. THE EVENT OCCURRED DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23538506 ↗
  • MalfunctionOct 31, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE EXHIBITED A STAINED IMAGE; THE ISSUE OCCURRED DURING THE SERVICE OR MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441230 ↗
  • MalfunctionMar 11, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS ON THE AIR/WATER CYLINDER AND TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21575394 ↗
  • MalfunctionDec 11, 2024Failure to Clean Adequately; Failure to Clean Adequately

    NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903124 ↗
  • MalfunctionDec 6, 2024Contamination; Contamination

    IT WAS REPORTED THAT PRESENCE OF A WHITE POWDER SLIPPED UNDER THE RUBBER ON THE COLONOVIDEOSCOPE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20870789 ↗
  • MalfunctionDec 4, 2024Device Reprocessing Problem; Obstruction of Flow; Device Reprocessing Problem; Obstruction of Flow

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20849958 ↗
  • MalfunctionNov 22, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD AN UNSUPPORTED SCOPE AND THE SCOPE WAS NOT TAKING ANY PICTURES. THE EVENT OCCURRED DURING A COLONOSCOPY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20758072 ↗
  • MalfunctionNov 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE HAD AN UNSPECIFIED AIR/WATER FLOW ISSUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20631767 ↗
  • MalfunctionNov 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT FOREIGN MATERIAL WAS FOUND IN THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20630462 ↗
  • MalfunctionOct 29, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT EXHIBITED WHITE RESIDUE IN THE AIR/WATER CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20555824 ↗
  • MalfunctionOct 15, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE JET TUBE OF THE COLONOVIDEOSCOPE HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20454290 ↗
  • MalfunctionOct 11, 2024Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THAT THE ADHESIVE WAS STUCK INSIDE THE BIOPSY CHANNEL OF THE COLONOVIDEOSCOPE. THIS OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20427750 ↗
  • MalfunctionOct 10, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE AIR/ WATER NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20421066 ↗
  • MalfunctionOct 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED WHITE FOREIGN MATERIAL INSIDE THE JET TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20385184 ↗
  • MalfunctionOct 4, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE COLONOVIDEOSCOPE HAD AN ERROR CODE E315 (INCOMPATIBLE SCOPE). THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20367751 ↗
  • MalfunctionOct 2, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED, THE COLONOVIDEOSCOPE HAD E315 SCOPE ERROR. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20353347 ↗
  • MalfunctionSep 25, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONO VIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON THE BENDING SECTION COVER AND THE CONNECTING TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20303908 ↗
  • MalfunctionSep 25, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONO VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE AIR/WATER TUBE AND THE AIR/WATER CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20298503 ↗
  • MalfunctionSep 17, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE COLONOVIDEOSCOPE HAD E216 AND E315 COMMUNICATION PROBLEM ERRORS. THE ISSUE OCCURRED WHEN TAKING THE DEVICES OUT OF THE BOXES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20242592 ↗
  • MalfunctionSep 17, 2024Thermal Decomposition of Device; Failure to Clean Adequately; Thermal Decomposition of Device; Failure to Clean Adequately

    CORRECTION TO B5 OF THE INITIAL REPORT. THERE WAS NO BURN ON THE PLASTIC DISTAL END COVER OBSERVED DURING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20235907 ↗
  • MalfunctionSep 16, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL AT THE AIR/WATER TUBE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20223594 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131780. One submission often covers many models, so check the product and lot details in each record.

  • Z-1210-2024 · Jan 11, 2024 · Open, Classified
    Colonoscope, Model Number PCF-H190DL.
  • Z-0418-2024 · Oct 16, 2023 · Open, Classified
    PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0419-2024 · Oct 16, 2023 · Open, Classified
    PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0416-2024 · Oct 16, 2023 · Open, Classified
    GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
  • Z-0417-2024 · Oct 16, 2023 · Open, Classified
    GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
  • Z-0415-2024 · Oct 16, 2023 · Open, Classified
    CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0341-2024 · Sep 25, 2023 · Open, Classified
    Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-
  • Z-0056-2025 · Sep 14, 2023 · Open, Classified
    EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancil

Other Colonoscope And Accessories, Flexible/Rigid devices

All Colonoscope And Accessories, Flexible/Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page