EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190 (REF N4505842)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus GIF-1TH190 gastrointestinal videoscope has 24 adverse event reports, all describing malfunctions, with no deaths and no injuries. Reports mostly describe foreign material or residue left inside the scope's channels after cleaning, a damaged tip with plastic strands, and poor or missing image display. No recalls name this device. Reports rose sharply, from 3 in the earlier 12-month period to 20 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 04953170452475), received between Oct 17, 2024 and Jun 25, 2026 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K232997); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately3
- Poor Quality Image3
- Flare or Flash2
- No Display/Image2
- Residue After Decontamination2
- Thermal Decomposition of Device2
- Device Contamination with Chemical or Other Material1
- Device Reprocessing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 25, 2026Failure to Clean Adequately
THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD WHITE FOREIGN MATERIAL IN THE AIR/WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25619939 ↗MalfunctionMay 29, 2026Failure to Clean Adequately
THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE ANALYSIS, IT WAS OBSERVED THAT THE AIR/WATER CHANNEL HAD FOREIGN MATERIAL. THE ROOT CAUSE OF THE FOREIGN MATERIAL COULD NOT BE IDENTIFIED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25334151 ↗MalfunctionApr 16, 2026Material Integrity Problem
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24897137 ↗MalfunctionApr 6, 2026Thermal Decomposition of Device
IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A BURNT DISTAL END PLASTIC COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790901 ↗MalfunctionMar 30, 2026Thermal Decomposition of Device
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE DISTAL SHEATH OF THE GASTROINTESTINAL VIDEOSCOPE WAS MELTED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24730362 ↗MalfunctionMar 20, 2026No Display/Image
IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE IS SHOWING NO IMAGE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24648727 ↗MalfunctionMar 6, 2026Obstruction of Flow
IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A CLIP COMING OUT OF IT DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24536769 ↗MalfunctionFeb 25, 2026Melted
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED BURNT PLASTIC DISTAL END COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24445534 ↗MalfunctionFeb 13, 2026Poor Quality Image; Overheating of Device
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED BURNT COVER INSULATION AND A BURNT DISTAL END OF THE PLASTIC COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24335441 ↗MalfunctionJan 30, 2026Residue After Decontamination
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED WHITE RESIDUE INSIDE THE NOZZLE AND THE CONTROL BODY CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24225975 ↗MalfunctionJan 22, 2026No Display/Image
IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAD NO IMAGE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24144440 ↗MalfunctionDec 23, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED DUE TO CLOG WITH WHITE RESIDUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23878211 ↗MalfunctionDec 16, 2025Flare or Flash; Flare or Flash
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURNED COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821965 ↗MalfunctionDec 12, 2025Poor Quality Image; Poor Quality Image
IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE IMAGE WAS GRAINY WHEN HOOKED UP TO THE PROCESSOR. THIS OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23786486 ↗MalfunctionDec 5, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED WHITE RESIDUES IN THE AIR/WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23731664 ↗MalfunctionOct 15, 2025Flare or Flash; Flare or Flash
IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAD A BURNT DISTAL TIP. THE ISSUE OCCURRED DURING AN UNKNOWN PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23299295 ↗MalfunctionAug 26, 2025Poor Quality Image; Residue After Decontamination; Poor Quality Image; Residue After Decontamination
IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAD BLURRY VISION. IT IS UNKNOWN WHEN THIS MALFUNCTION OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22896560 ↗MalfunctionJun 6, 2025Loss of or Failure to Bond; Device Contamination with Chemical or Other Material; Loss of or Failure to Bond; Device Contamination with Chemical or Other Material
IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAS GLUE IN IT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22159423 ↗MalfunctionOct 17, 2024Microbial Contamination of Device; Microbial Contamination of Device
IT WAS REPORTED THAT, DURING REPROCESSING, AFTER A MICROBIOLOGICAL ROUTINE CONTROL, THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR >200 COLONY FORMING UNITS (CFUS) OF ENTEROBACTER CLOACAE COMPLEX MICROORGANISMS AND STENOTROPHOMONAS MALTOPHILIA. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20472658 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K232997. One submission often covers many models, so check the product and lot details in each record.
- Z-0620-2025 · Oct 31, 2024 · Open, ClassifiedEVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
- Z-0498-2025 · Sep 18, 2024 · Open, ClassifiedBrand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This
Other Gastroscope And Accessories, Flexible/Rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page