PENTAX EC-3490LKRecalls & reported problems
Hoya Corporation · VIDEO COLONOSCOPE EC-3490K SERIES · Class 2
Quick summary
This Pentax video colonoscope is a flexible scope used to examine the colon. The 42 reports mostly describe video image problems such as cloudy, blacked-out or failed images, and parts breaking or peeling off, with no patient harm mentioned in recent reports. The reports include 2 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 04961333223695), received between Mar 1, 2022 and Nov 19, 2024 — classified as 2 injuries, 40 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
6 other recalls cover devices cleared under the same FDA submission (K131855); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Detachment of Device or Device Component8
- Erratic or Intermittent Display8
- Obstruction of Flow4
- Optical Obstruction4
- Fluid/Blood Leak3
- No Display/Image3
- Suction Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions42
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2024Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20720763 ↗MalfunctionOct 15, 2024Optical Obstruction; Optical Obstruction
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20448325 ↗MalfunctionOct 10, 2024No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20414424 ↗MalfunctionOct 10, 2024No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20414377 ↗MalfunctionJun 27, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19621414 ↗MalfunctionMay 24, 2024Optical Obstruction; Optical Obstruction
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19389291 ↗MalfunctionMay 13, 2024Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19296036 ↗MalfunctionApr 8, 2024Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19057232 ↗MalfunctionJan 12, 2024Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18500886 ↗MalfunctionJan 9, 2024Optical Obstruction; Optical Obstruction
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18472447 ↗MalfunctionNov 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18203728 ↗MalfunctionNov 13, 2023Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18117602 ↗MalfunctionSep 12, 2023No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17734014 ↗MalfunctionSep 12, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17727283 ↗MalfunctionSep 6, 2023Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17688384 ↗MalfunctionAug 25, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. BIOPSY INLET PIECE LOOSENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17621212 ↗MalfunctionAug 23, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17600989 ↗MalfunctionJun 22, 2023Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17178754 ↗MalfunctionJun 8, 2023Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17084520 ↗InjuryJun 6, 2023Suction Problem; Suction Problem
PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, AN EVENT OCCURRED DURING A PATIENT'S COLONOSCOPY PROCEDURE. THE DEVICE MALFUNCTIONED AND THERE WAS A CONCERN FOR POTENTIAL PERFORATION DUE TO THE MALFUNCTION. IN REGARD TO THIS EVENT, AND OTHERS, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED/RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNORMALITIES, THEREFORE THE POSSIBILITY OF MISSING PATHOLOGY; RECOGNIZING LANDMARKS NECESSARY TO COMPLETE THE PROCEDURE, AND PROLONGING THE PROCEDURE OR RENDERING IT HARDER TO COMPLETE. EVENTUALLY, AND IN THE MORE SERIOUS EVENTS, SUCTION OVERCAME INSUFFLATION RESULTING IN SUCTION INTO THE MUCOSA. THIS LEADS TO EXTREME DIFFICULTY COMPLETING THE PROCEDURE, MUCOSAL INJURY THROUGH CREATION OF 'SUCTION POLYPS', AND MUCOSAL TRAUMA AS WAS DOCUMENTED. THE INABILITY TO MAINTAIN LUMINAL INSUFFLATION ALSO GIVES RISE TO THE SUSPICION OF PERFORATION, WHICH THEN HAS RA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17068703 ↗InjuryJun 5, 2023Suction Problem; Suction Problem
PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, AN EVENT OCCURRED DURING A PATIENT'S COLONOSCOPY PROCEDURE. THE DEVICE MALFUNCTIONED AND THERE WAS A CONCERN FOR POTENTIAL PERFORATION DUE TO THE MALFUNCTION. IN REGARD TO THIS EVENT, AND OTHERS, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED/RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS:(1) THE ABILITY TO IDENTIFY ABNORMALITIES, THEREFORE THE POSSIBILITY OF MISSING PATHOLOGY; (2) RECOGNIZING LANDMARKS NECESSARY TO COMPLETE THE PROCEDURE, AND (3) PROLONGING THE PROCEDURE OR RENDERING IT HARDER TO COMPLETE. EVENTUALLY, AND IN THE MORE SERIOUS EVENTS, SUCTION OVERCAME INSUFFLATION RESULTING IN SUCTION INTO THE MUCOSA. THIS LEADS TO EXTREME DIFFICULTY COMPLETING THE PROCEDURE, MUCOSAL INJURY THROUGH CREATION OF 'SUCTION POLYPS', AND MUCOSAL TRAUMA AS WAS DOCUMENTED. THE INABILITY TO MAINTAIN LUMINAL INSUFFLATION ALSO GIVES RISE TO THE SUSPICION OF PERFORATION, WHICH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17058165 ↗MalfunctionMay 30, 2023Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17028371 ↗MalfunctionFeb 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16406714 ↗MalfunctionFeb 21, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16406712 ↗MalfunctionFeb 21, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16406700 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131855. One submission often covers many models, so check the product and lot details in each record.
- Z-2449-2021 · Jul 14, 2021 · Open, ClassifiedGastroscope Family # 1-Gastroscopes without a Water Jet Channel Models: EG-2790i, EG-1690K, EG-2490K, EG-2790K, EG27-i10; Gastroscope Family # 2- Gastroscopes with a Water Jet Chan
- Z-2448-2021 · Jul 14, 2021 · Open, ClassifiedColonoscope Family # 1-Pentax Video Colonoscopes (EC Family) with One Instrument Channel and a Water Jet Channel Models: EC-2990Li, EC-3490Li, EC-3890Li, EC-3490TLi, EC-3490LK, EC
- Z-1268-2020 · Jan 3, 2020 · TerminatedPentax Video Colonoscope Model: EC38-i10L
- Z-1267-2020 · Jan 3, 2020 · TerminatedPentax Video Colonoscope Model: EC34-i10L
- Z-0636-2017 · Aug 15, 2016 · TerminatedPentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forcep
- Z-2068-2016 · Apr 6, 2016 · TerminatedColonoscope Intended to provide optical visualization of and therapeutic access to the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.