Skip to content

PENTAX EG29-i10 (UDI 04961333223909)Recalls & reported problems

Hoya Corporation · PENTAX VIDEO UPPER G.I.SCOPE EG29-i10 · Class 2

All models of PENTAX (72)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Pentax upper GI scope has 130 adverse event reports on record, with no deaths and no injuries. The reports mostly describe video image failures, such as blackouts or cloudy images, and some mention parts coming loose, clogged air or water channels, or leaks. Reports note no patient harm. There are no recalls naming this device. No reports came in the last 12 months, compared with 12 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
130
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 130 adverse event reports that list this exact device (DI 04961333223909), received between Sep 8, 2021 and Mar 3, 2025 — classified as 130 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (37 reports).

2 other recalls cover devices cleared under the same FDA submission (K131902); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 8 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component37
  • Break34
  • Obstruction of Flow22
  • Fluid/Blood Leak14
  • No Display/Image8
  • Optical Distortion8
  • Component Missing2
  • Erratic or Intermittent Display2

Patient problems

  • No Clinical Signs, Symptoms or Conditions130

Recent adverse event reports

The 25 most recent of 130 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21501502 ↗
  • MalfunctionFeb 20, 2025No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21418061 ↗
  • MalfunctionFeb 13, 2025No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21371301 ↗
  • MalfunctionFeb 5, 2025No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21309852 ↗
  • MalfunctionFeb 5, 2025Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21309749 ↗
  • MalfunctionJan 28, 2025No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21249497 ↗
  • MalfunctionJan 16, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21164892 ↗
  • MalfunctionJan 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21088542 ↗
  • MalfunctionNov 27, 2024Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20793920 ↗
  • MalfunctionNov 25, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20769023 ↗
  • MalfunctionNov 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20731231 ↗
  • MalfunctionNov 7, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. INSERTION FLEXIBLE TUBE (IFT) STEEL BRAID PIERCING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20627709 ↗
  • MalfunctionOct 28, 2024Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(SHADOW IN IMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20542270 ↗
  • MalfunctionOct 25, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20530448 ↗
  • MalfunctionOct 10, 2024Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414449 ↗
  • MalfunctionOct 10, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414444 ↗
  • MalfunctionOct 10, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414398 ↗
  • MalfunctionOct 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20356976 ↗
  • MalfunctionOct 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20356965 ↗
  • MalfunctionOct 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20356882 ↗
  • MalfunctionSep 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20264715 ↗
  • MalfunctionSep 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20254396 ↗
  • MalfunctionSep 10, 2024Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. OPERATION/SUCTION CHANNEL CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20178190 ↗
  • MalfunctionAug 26, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20062369 ↗
  • MalfunctionAug 23, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20051568 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131902. One submission often covers many models, so check the product and lot details in each record.

  • Z-0318-2017 · Aug 15, 2016 · Terminated
    Pentax Video Gastroscope is intended to be used with a Pentax video processor (including light source), documentation equipment, monitor, EndoTherapy Device such as a Biopsy Forcep
  • Z-2071-2016 · Apr 6, 2016 · Terminated
    Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page