Skip to content

PENTAX EG-2490KRecalls & reported problems

Hoya Corporation · PENTAX VIDEO UPPER G.I.SCOPE EG-2490K · Class 2

All models of PENTAX (72)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This PENTAX video gastroscope is used to examine the upper digestive tract. Its 26 reports mostly describe parts coming loose, loss of the video image, fluid or blood leaks, and occasional clogged channels or missing components. The reports include no deaths or injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 04961333225651), received between Jun 26, 2020 and May 14, 2026 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (8 reports).

2 other recalls cover devices cleared under the same FDA submission (K131902); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component8
  • No Display/Image6
  • Fluid/Blood Leak4
  • Component Missing2
  • Obstruction of Flow2
  • Break1
  • Device Reprocessing Problem1
  • Erratic or Intermittent Display1

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. OPERATION/SUCTION CHANNEL CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25167937 ↗
  • MalfunctionMar 26, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (SHADOW IN IMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21694833 ↗
  • MalfunctionMar 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21501500 ↗
  • MalfunctionDec 3, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20828669 ↗
  • MalfunctionOct 28, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20542159 ↗
  • MalfunctionOct 10, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414400 ↗
  • MalfunctionJul 26, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN.THERE WAS NO REPORT OF PATIENT HARM.VIDEO IMAGE FAILURE(BLACKOUT )

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19834747 ↗
  • MalfunctionJul 1, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19641400 ↗
  • MalfunctionJun 26, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19612254 ↗
  • MalfunctionApr 12, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19094365 ↗
  • MalfunctionMar 11, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18872774 ↗
  • MalfunctionNov 15, 2023No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18137378 ↗
  • MalfunctionNov 1, 2023Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. OPERATION/SUCTION CHANNEL CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18045316 ↗
  • MalfunctionOct 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17937530 ↗
  • MalfunctionFeb 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16406719 ↗
  • MalfunctionNov 10, 2022Component Missing; Component Missing

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15768123 ↗
  • MalfunctionSep 19, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15452465 ↗
  • MalfunctionSep 16, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15436041 ↗
  • MalfunctionJul 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14882519 ↗
  • MalfunctionJun 30, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14869124 ↗
  • MalfunctionJun 17, 2022Component Missing; Component Missing

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14728433 ↗
  • MalfunctionJun 8, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14629493 ↗
  • MalfunctionJun 7, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14612609 ↗
  • MalfunctionFeb 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13589628 ↗
  • MalfunctionJun 26, 2020Break; Device Reprocessing Problem; Break; Device Reprocessing Problem

    PENTAX MEDICAL WAS MADE AWARE OF AN EVENT ON (B)(6) 2020 THAT OCCURRED AT THE (B)(6) INVOLVING PENTAX MEDICAL VIDEO GASTROSCOPE, MODEL EG-2490K, SERIAL NUMBER (B)(4). THE USER FACILITY CALLED IN AN ENDOSCOPE FOR EVALUATION REPORTING "POSSIBLE METAL SHAVINGS IN THE CHANNEL." BASED ON GOOD FAITH EFFORT RESPONSES RECEIVED FROM THE USER FACILITY ON 14-MAY-2020 AND 18-MAY-2020 VIA EMAIL, PENTAX LEARNED THAT THERE WERE NO ACCESSORIES USED DURING AN INTESTINAL BIOPSY PROCEDURE. THE USER FLUSHED THE PRIMATE WITH WATER AND SUCTIONED BACK AND FORTH A FEW TIMES TO TRY TO REMOVE ANY AND AS MUCH OF THE REPORTED DEBRIS AS POSSIBLE. THERE WERE NO REPORTED SERIOUS INJURY OR DEATH OF A PATIENT OR USER, OR DELAY IN THE PROCEDURE WHICH WOULD REQUIRE MEDICAL INTERVENTION; AND THE ENDOSCOPE WAS USED TO COMPLETE THE PROCEDURE. THE ENDOSCOPE WAS THEN REMOVED FROM CIRCULATION AND SUBSEQUENTLY CALLED IN FOR SERVICE. THE USER FACILITY ALSO ALERTED US THAT THE ENDOSCOPE WAS USED ON NON HUMAN PRIMATES. THE DATE OF OCCURRENCE IS UNKNOWN BUT THE USER DID STATE THAT THE PROCEDURE WAS A FEW WEEKS PRIOR AND THE ANIMAL IS FINE. THE GASTROSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 11-JUN-2020. THE GASTRO

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10203188 ↗

FDA clearance and classification

Cleared through 510(k) K131902 on Apr 10, 2014 (decision: Substantially Equivalent). Submitted by Pentax Medical Company as “Pentax Video Upper G.I. Scopes (Eg Family)”. The FDA review took 289 days. FDA record ↗

The FDA regulates product code FDS (gastroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131902. One submission often covers many models, so check the product and lot details in each record.

  • Z-0318-2017 · Aug 15, 2016 · Terminated
    Pentax Video Gastroscope is intended to be used with a Pentax video processor (including light source), documentation equipment, monitor, EndoTherapy Device such as a Biopsy Forcep
  • Z-2071-2016 · Apr 6, 2016 · Terminated
    Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti

Common questions about the PENTAX EG-2490K

Has the PENTAX EG-2490K been recalled?

No FDA recall record in our data names this exact device identifier (DI 04961333225651), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the PENTAX EG-2490K?

The FDA's MAUDE database lists 25 adverse event reports for this device, received between Jun 26, 2020 and May 14, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (8), “No Display/Image” (6) and “Fluid/Blood Leak” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PENTAX EG-2490K?

None of the reports are classified as deaths, 0 as injuries and 25 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PENTAX EG-2490K FDA approved?

It was cleared through 510(k) K131902 on Apr 10, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the PENTAX EG-2490K, and is it still sold?

The FDA's device database lists Hoya Corporation as the company responsible for it (brand PENTAX, model EG-2490K). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page