PENTAX EG-1690KRecalls & reported problems
Hoya Corporation · PENTAX VIDEO UPPER G.I.SCOPE EG-1690K · Class 2
All models of PENTAX (72)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Pentax video upper GI scope (EG-1690K) is a flexible gastroscope used to look inside the upper digestive tract and take tissue samples. Most of its 54 reports describe the scope breaking or coming apart during use, along with weak suction and air-flow problems and cloudy or lost video images; in 6 reports patients were injured, with one describing a piece of the scope's bending rubber coming loose inside the patient and not being retrievable. No deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04961333225743), received between Nov 13, 2020 and Oct 1, 2025 — classified as 6 injuries, 48 malfunctions.
The most frequently reported specific device problem is “Break” (14 reports).
2 other recalls cover devices cleared under the same FDA submission (K131902); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break14
- Detachment of Device or Device Component12
- Obstruction of Flow8
- Insufficient Flow or Under Infusion4
- Suction Problem4
- Optical Distortion3
- Fluid/Blood Leak2
- No Display/Image2
Patient problems
- No Clinical Signs, Symptoms or Conditions52
- Device Embedded In Tissue or Plaque2
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 1, 2025Improper Flow or Infusion; Improper Flow or Infusion
PENTAX MEDICAL RECEIVED A VOLUNTARY MEDWATCH REPORT (B)(4) FROM AN FDA MEDSUN PROGRAM USER FACILITY VIA EMAIL ON SEPTEMBER 12, 2025. ACCORDING TO THE USER REPORT, THE INSUFFLATION BECAME INADEQUATE DURING THE ENDOSCOPIC PROCEDURE. AS A RESULT, THE INSIDE OF THE BODY COULD NOT BE VISUALIZED, AND THE SPECIMEN COLLECTION PROCEDURE WAS DISCONTINUED. AFTER THE PROCEDURE, THE ENDOSCOPE CONCERNED WAS INSPECTED, AND THE INSUFFLATION FAILURE WAS REPRODUCED. THE DISTRIBUTOR WAS CONTACTED REGARDING THIS EVENT, AND THE ENDOSCOPE WAS SENT FOR INVESTIGATION. THERE WAS NO REPORT OF PATIENT HARM. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY; HOWEVER, SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23188720 ↗MalfunctionMar 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. BENDING RUBBER MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21501520 ↗MalfunctionFeb 19, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21407854 ↗MalfunctionFeb 13, 2025Optical Obstruction; Optical Obstruction
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE CLOUDY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21371182 ↗InjuryOct 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
DUE TO AN EMERGENCY, AN EXAMINATION WAS CONDUCTED WITH EG-1690K THAT HAD NOT BEEN USED FOR A LONG TIME AND HAD BEEN STORED IN A WAREHOUSE, AND THE BENDING RUBBER FELL OUT INSIDE THE PATIENT'S BODY. THE BENDING RUBBER THAT HAD FALLEN OFF INSIDE THE PATIENT'S BODY COULD NOT BE RETRIEVED. AFTER THE EXAMINATION, THE DOCTOR DETERMINED THAT THERE WAS NO HEALTH HAZARD TO THE PATIENT AND NO EXPLANATION WAS GIVEN TO THE PATIENT. THE HOSPITAL DIRECTOR ALSO SAID THAT THE EVENT WILL NOT BE REPORTED TO THE ADMINISTRATION. THIS EVENT OCCURRED AT THE TIME OF DURING USE. THERE WAS NO REPORT OF PATIENT HARM. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY; HOWEVER, SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20553454 ↗InjuryOct 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
MalfunctionOct 10, 2024Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20414427 ↗MalfunctionJun 23, 2023No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17189161 ↗InjuryJun 6, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion
PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, ENDOSCOPE USED DURING PROCEDURE FAILED TO BLOW AIR AND ALLOW INSUFFLATION OF THE UPPER GI TRACT PRECLUDING COMPLETION OF THE PROCEDURE SAFELY. THE PROCEDURE WAS THEREFORE SUSPENDED, AND FOLLOWED BY MULTIPLE TROUBLESHOOTING STEPS THAT FAILED UNTIL THE SCOPE WAS CHANGED (LARGER SCOPE AVAILABLE ON SITE). THEREFORE, TEAM HAD TO REQUEST EQUIVALENT 17 SERIES SCOPE, WHICH, AFTER INHERENT DELAY, WAS USED TO COMPLETE THE ENDOSCOPY UNEVENTFULLY. PATIENT HAD INCREASED DURATION OF PROCEDURE OVER ROUTINE 5 MINS. THERE ARE SEVERAL EVENTS, AND THIS ISSUE HAS BEEN EXPERIENCED BY SEVERAL PHYSICIANS. IN ALL INSTANCES, THE PROCEDURE (USUALLY BUT NOT EXCLUSIVELY A COLONOSCOPY) STARTS UNEVENTFULLY. DURING THE PROCEDURE, MORE LIKELY IF PROLONGED / POOR CLEANOUT, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED / RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17076222 ↗InjuryJun 6, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion
PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, ENDOSCOPE USED DURING PROCEDURE FAILED TO BLOW AIR AND ALLOW INSUFFLATION OF THE UPPER GI TRACT PRECLUDING COMPLETION OF THE PROCEDURE SAFELY. THE PROCEDURE WAS THEREFORE SUSPENDED, AND FOLLOWED BY MULTIPLE TROUBLESHOOTING STEPS THAT FAILED UNTIL THE SCOPE WAS CHANGED (LARGER SCOPE AVAILABLE ON SITE). THEREFORE, TEAM HAD TO REQUEST EQUIVALENT 17 SERIES SCOPE, WHICH, AFTER INHERENT DELAY, WAS USED TO COMPLETE THE ENDOSCOPY UNEVENTFULLY. PATIENT HAD INCREASED DURATION OF PROCEDURE OVER ROUTINE 5 MINS. THERE ARE SEVERAL EVENTS, AND THIS ISSUE HAS BEEN EXPERIENCED BY SEVERAL PHYSICIANS. IN ALL INSTANCES, THE PROCEDURE (USUALLY BUT NOT EXCLUSIVELY A COLONOSCOPY) STARTS UNEVENTFULLY. DURING THE PROCEDURE, MORE LIKELY IF PROLONGED / POOR CLEANOUT, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED / RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17076221 ↗InjuryJun 5, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion
InjuryJun 5, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion
PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, ENDOSCOPE USED DURING PROCEDURE FAILED TO BLOW AIR AND ALLOW INSUFFLATION OF THE UPPER GI TRACT PRECLUDING COMPLETION OF THE PROCEDURE SAFELY. THE PROCEDURE WAS THEREFORE SUSPENDED, AND FOLLOWED BY MULTIPLE TROUBLESHOOTING STEPS THAT FAILED UNTIL THE SCOPE WAS CHANGED (LARGER SCOPE AVAILABLE ON SITE). THEREFORE, TEAM HAD TO REQUEST EQUIVALENT 17 SERIES SCOPE, WHICH, AFTER INHERENT DELAY, WAS USED TO COMPLETE THE ENDOSCOPY UNEVENTFULLY. PATIENT HAD INCREASED DURATION OF PROCEDURE OVER ROUTINE 5 MINS. THERE ARE SEVERAL EVENTS, AND THIS ISSUE HAS BEEN EXPERIENCED BY SEVERAL PHYSICIANS. IN ALL INSTANCES, THE PROCEDURE (USUALLY BUT NOT EXCLUSIVELY A COLONOSCOPY) STARTS UNEVENTFULLY. DURING THE PROCEDURE, MORE LIKELY IF PROLONGED / POOR CLEANOUT, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED / RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17061627 ↗MalfunctionMay 29, 2023Obstruction of Flow; Obstruction of Flow
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17016271 ↗MalfunctionMay 16, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ANGLE WIRE CUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16936741 ↗MalfunctionApr 6, 2023No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16688058 ↗MalfunctionApr 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16662038 ↗MalfunctionMar 16, 2023Obstruction of Flow; Obstruction of Flow
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16551313 ↗MalfunctionFeb 27, 2023Optical Obstruction; Optical Obstruction
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16441636 ↗MalfunctionJan 31, 2023Component Missing; Component Missing
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16275465 ↗MalfunctionDec 7, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15929713 ↗MalfunctionDec 2, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ANGLE WIRE CUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15903364 ↗MalfunctionNov 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15705632 ↗MalfunctionOct 14, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ANGLE WIRE CUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15602615 ↗MalfunctionSep 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15388556 ↗MalfunctionAug 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15308289 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131902. One submission often covers many models, so check the product and lot details in each record.
- Z-0318-2017 · Aug 15, 2016 · TerminatedPentax Video Gastroscope is intended to be used with a Pentax video processor (including light source), documentation equipment, monitor, EndoTherapy Device such as a Biopsy Forcep
- Z-2071-2016 · Apr 6, 2016 · TerminatedGastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page