PENTAX EG-1690KRecalls & reported problems

Hoya Corporation · PENTAX VIDEO UPPER G.I.SCOPE EG-1690K · Class 2

All models of PENTAX (72)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Pentax video upper GI scope (EG-1690K) is a flexible gastroscope used to look inside the upper digestive tract and take tissue samples. Most of its 54 reports describe the scope breaking or coming apart during use, along with weak suction and air-flow problems and cloudy or lost video images; in 6 reports patients were injured, with one describing a piece of the scope's bending rubber coming loose inside the patient and not being retrievable. No deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04961333225743), received between Nov 13, 2020 and Oct 1, 2025 — classified as 6 injuries, 48 malfunctions.

The most frequently reported specific device problem is “Break” (14 reports).

2 other recalls cover devices cleared under the same FDA submission (K131902); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break14
  • Detachment of Device or Device Component12
  • Obstruction of Flow8
  • Insufficient Flow or Under Infusion4
  • Suction Problem4
  • Optical Distortion3
  • Fluid/Blood Leak2
  • No Display/Image2

Patient problems

  • No Clinical Signs, Symptoms or Conditions52
  • Device Embedded In Tissue or Plaque2

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 1, 2025Improper Flow or Infusion; Improper Flow or Infusion

    PENTAX MEDICAL RECEIVED A VOLUNTARY MEDWATCH REPORT (B)(4) FROM AN FDA MEDSUN PROGRAM USER FACILITY VIA EMAIL ON SEPTEMBER 12, 2025. ACCORDING TO THE USER REPORT, THE INSUFFLATION BECAME INADEQUATE DURING THE ENDOSCOPIC PROCEDURE. AS A RESULT, THE INSIDE OF THE BODY COULD NOT BE VISUALIZED, AND THE SPECIMEN COLLECTION PROCEDURE WAS DISCONTINUED. AFTER THE PROCEDURE, THE ENDOSCOPE CONCERNED WAS INSPECTED, AND THE INSUFFLATION FAILURE WAS REPRODUCED. THE DISTRIBUTOR WAS CONTACTED REGARDING THIS EVENT, AND THE ENDOSCOPE WAS SENT FOR INVESTIGATION. THERE WAS NO REPORT OF PATIENT HARM. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY; HOWEVER, SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23188720 ↗
  • MalfunctionMar 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. BENDING RUBBER MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21501520 ↗
  • MalfunctionFeb 19, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21407854 ↗
  • MalfunctionFeb 13, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21371182 ↗
  • InjuryOct 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    DUE TO AN EMERGENCY, AN EXAMINATION WAS CONDUCTED WITH EG-1690K THAT HAD NOT BEEN USED FOR A LONG TIME AND HAD BEEN STORED IN A WAREHOUSE, AND THE BENDING RUBBER FELL OUT INSIDE THE PATIENT'S BODY. THE BENDING RUBBER THAT HAD FALLEN OFF INSIDE THE PATIENT'S BODY COULD NOT BE RETRIEVED. AFTER THE EXAMINATION, THE DOCTOR DETERMINED THAT THERE WAS NO HEALTH HAZARD TO THE PATIENT AND NO EXPLANATION WAS GIVEN TO THE PATIENT. THE HOSPITAL DIRECTOR ALSO SAID THAT THE EVENT WILL NOT BE REPORTED TO THE ADMINISTRATION. THIS EVENT OCCURRED AT THE TIME OF DURING USE. THERE WAS NO REPORT OF PATIENT HARM. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY; HOWEVER, SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 20553454 ↗
  • InjuryOct 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    REFER TO H11.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 20546262 ↗
  • MalfunctionOct 10, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414427 ↗
  • MalfunctionJun 23, 2023No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17189161 ↗
  • InjuryJun 6, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion

    PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, ENDOSCOPE USED DURING PROCEDURE FAILED TO BLOW AIR AND ALLOW INSUFFLATION OF THE UPPER GI TRACT PRECLUDING COMPLETION OF THE PROCEDURE SAFELY. THE PROCEDURE WAS THEREFORE SUSPENDED, AND FOLLOWED BY MULTIPLE TROUBLESHOOTING STEPS THAT FAILED UNTIL THE SCOPE WAS CHANGED (LARGER SCOPE AVAILABLE ON SITE). THEREFORE, TEAM HAD TO REQUEST EQUIVALENT 17 SERIES SCOPE, WHICH, AFTER INHERENT DELAY, WAS USED TO COMPLETE THE ENDOSCOPY UNEVENTFULLY. PATIENT HAD INCREASED DURATION OF PROCEDURE OVER ROUTINE 5 MINS. THERE ARE SEVERAL EVENTS, AND THIS ISSUE HAS BEEN EXPERIENCED BY SEVERAL PHYSICIANS. IN ALL INSTANCES, THE PROCEDURE (USUALLY BUT NOT EXCLUSIVELY A COLONOSCOPY) STARTS UNEVENTFULLY. DURING THE PROCEDURE, MORE LIKELY IF PROLONGED / POOR CLEANOUT, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED / RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNOR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17076222 ↗
  • InjuryJun 6, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion

    PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, ENDOSCOPE USED DURING PROCEDURE FAILED TO BLOW AIR AND ALLOW INSUFFLATION OF THE UPPER GI TRACT PRECLUDING COMPLETION OF THE PROCEDURE SAFELY. THE PROCEDURE WAS THEREFORE SUSPENDED, AND FOLLOWED BY MULTIPLE TROUBLESHOOTING STEPS THAT FAILED UNTIL THE SCOPE WAS CHANGED (LARGER SCOPE AVAILABLE ON SITE). THEREFORE, TEAM HAD TO REQUEST EQUIVALENT 17 SERIES SCOPE, WHICH, AFTER INHERENT DELAY, WAS USED TO COMPLETE THE ENDOSCOPY UNEVENTFULLY. PATIENT HAD INCREASED DURATION OF PROCEDURE OVER ROUTINE 5 MINS. THERE ARE SEVERAL EVENTS, AND THIS ISSUE HAS BEEN EXPERIENCED BY SEVERAL PHYSICIANS. IN ALL INSTANCES, THE PROCEDURE (USUALLY BUT NOT EXCLUSIVELY A COLONOSCOPY) STARTS UNEVENTFULLY. DURING THE PROCEDURE, MORE LIKELY IF PROLONGED / POOR CLEANOUT, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED / RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNOR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17076221 ↗
  • InjuryJun 5, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion

    REFER TO H10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17066893 ↗
  • InjuryJun 5, 2023Suction Problem; Insufficient Flow or Under Infusion; Suction Problem; Insufficient Flow or Under Infusion

    PATIENT INVOLVED - NO KNOWN ADVERSE EVENT. PER THE INITIAL REPORT, ENDOSCOPE USED DURING PROCEDURE FAILED TO BLOW AIR AND ALLOW INSUFFLATION OF THE UPPER GI TRACT PRECLUDING COMPLETION OF THE PROCEDURE SAFELY. THE PROCEDURE WAS THEREFORE SUSPENDED, AND FOLLOWED BY MULTIPLE TROUBLESHOOTING STEPS THAT FAILED UNTIL THE SCOPE WAS CHANGED (LARGER SCOPE AVAILABLE ON SITE). THEREFORE, TEAM HAD TO REQUEST EQUIVALENT 17 SERIES SCOPE, WHICH, AFTER INHERENT DELAY, WAS USED TO COMPLETE THE ENDOSCOPY UNEVENTFULLY. PATIENT HAD INCREASED DURATION OF PROCEDURE OVER ROUTINE 5 MINS. THERE ARE SEVERAL EVENTS, AND THIS ISSUE HAS BEEN EXPERIENCED BY SEVERAL PHYSICIANS. IN ALL INSTANCES, THE PROCEDURE (USUALLY BUT NOT EXCLUSIVELY A COLONOSCOPY) STARTS UNEVENTFULLY. DURING THE PROCEDURE, MORE LIKELY IF PROLONGED / POOR CLEANOUT, THE ENDOSCOPE SUCTIONS THE LUMINAL CONTENTS WITHOUT THE SUCTION BUTTON BEING PRESSED. ONCE THIS STARTS, IT TENDS TO GET WORSE. IN SOME CASES, CHANGING THE VALVE TO A NON-DISPOSABLE VALVE IMPROVED / RESOLVED THE ISSUE. THIS LEADS TO POOR AND NON-SUSTAINED INSUFFLATION OF THE INTESTINAL LUMEN, RESULTING IN POOR VISUALIZATION OF THE LUMEN. THIS IMPACTS: THE ABILITY TO IDENTIFY ABNOR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17061627 ↗
  • MalfunctionMay 29, 2023Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17016271 ↗
  • MalfunctionMay 16, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ANGLE WIRE CUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16936741 ↗
  • MalfunctionApr 6, 2023No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16688058 ↗
  • MalfunctionApr 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16662038 ↗
  • MalfunctionMar 16, 2023Obstruction of Flow; Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16551313 ↗
  • MalfunctionFeb 27, 2023Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16441636 ↗
  • MalfunctionJan 31, 2023Component Missing; Component Missing

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16275465 ↗
  • MalfunctionDec 7, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15929713 ↗
  • MalfunctionDec 2, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ANGLE WIRE CUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15903364 ↗
  • MalfunctionNov 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15705632 ↗
  • MalfunctionOct 14, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ANGLE WIRE CUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15602615 ↗
  • MalfunctionSep 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15388556 ↗
  • MalfunctionAug 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15308289 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131902. One submission often covers many models, so check the product and lot details in each record.

  • Z-0318-2017 · Aug 15, 2016 · Terminated
    Pentax Video Gastroscope is intended to be used with a Pentax video processor (including light source), documentation equipment, monitor, EndoTherapy Device such as a Biopsy Forcep
  • Z-2071-2016 · Apr 6, 2016 · Terminated
    Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →