PENTAX EG-3490KRecalls & reported problems
Hoya Corporation · PENTAX VIDEO UPPER G.I.SCOPE EG-3490K · Class 2
All models of PENTAX (72)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04961333227044), received between Jan 7, 2021 and Jan 16, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K131902); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component3
- Device Damaged by Another Device1
- Erratic or Intermittent Display1
- Failure to Disconnect1
- Fluid/Blood Leak1
- Material Split, Cut or Torn1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21165050 ↗MalfunctionSep 10, 2024No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20178326 ↗MalfunctionSep 10, 2024Erratic or Intermittent Display; Erratic or Intermittent Display
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20178231 ↗MalfunctionMay 1, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. INSERTION FLEXIBLE TUBE (IFT) STEEL BRAID PIERCING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16841921 ↗MalfunctionJan 12, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16143595 ↗MalfunctionNov 14, 2022Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15786683 ↗MalfunctionNov 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15705615 ↗MalfunctionOct 20, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE PEELED OFF (BLACK GLUE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15636877 ↗MalfunctionOct 14, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15603118 ↗MalfunctionJan 7, 2021Partial Blockage; Break; Failure to Disconnect; Device Damaged by Another Device; Partial Blockage; Break; Failure to Disconnect; Device Damaged by Another Device
PENTAX MEDICAL WAS MADE AWARE OF A COMPLAINT ON 08-DEC-2020 UNDER RMA NOTIFICATION (B)(4) FOR AN "ACCESSORY STUCK IN SCOPE" INVOLVING THE PENTAX MEDICAL VIDEO GASTROSCOPE, MODEL EG-3490K, SERIAL NUMBER (B)(4). THE USER ALSO REPORTED THE STENT HAD BROKEN OFF. THE USER FACILITY RESPONDED TO A GFE REQUEST VIA EMAIL ON 15-DEC-2020 AND AGAIN ON 23-DEC-2020, AND STATED THE ENDOSCOPE WAS BEING USED FOR TREATMENT. A BOSTON SCIENTIFIC PANCREATIC PIGTAIL STENT AND WAS TAKEN OUT BY A ENDOCHOICE RAT TOOTH GRASPER. NO SERIOUS INJURY OR DEATH OF A PATIENT OR USER WAS REPORTED. THE ENDOSCOPE WAS REMOVED FROM CIRCULATION IMMEDIATELY AFTER THE FAILURE/EVENT OCCURRED AND SUBSEQUENTLY CALLED IN FOR SERVICE/REPLACEMENT THE CUSTOMER OWNED ENDOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 16-DEC-2020. DURING EVALUATION OF THE ENDOSCOPE UNDER SERVICE ORDER (B)(4), THE PENTAX MEDICAL SERVICE REPAIR TECHNICIAN FOUND AN "ACCESSORY STUCK IN PRIMARY OPERATION CHANNEL" CONFIRMING THE CUSTOMER'S COMPLAINT AND DOCUMENTING THE FOLLOWING ADDITIONAL FINDINGS ON 21-DEC-2020: PRIMARY OPERATION CHANNEL BLOCKED, PASSED DRY LEAK TEST, PASSED WET LEAK TEST, DISTAL BODY CHIP AT CHANNEL OPENING AT THINNEST PART, U
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11135699 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131902. One submission often covers many models, so check the product and lot details in each record.
- Z-0318-2017 · Aug 15, 2016 · TerminatedPentax Video Gastroscope is intended to be used with a Pentax video processor (including light source), documentation equipment, monitor, EndoTherapy Device such as a Biopsy Forcep
- Z-2071-2016 · Apr 6, 2016 · TerminatedGastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.