PENTAX VSB-3430KRecalls & reported problems
Hoya Corporation · PVE VIDEO SMALL BOWEL ENTEROSCOPE 30K SERIES · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04961333227686), received between Jul 15, 2021 and Jul 15, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Reprocessing Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K962897); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Reprocessing Problem1
- Partial Blockage1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2021Partial Blockage; Device Reprocessing Problem; Partial Blockage; Device Reprocessing Problem
THE USER FACILITY CONTACTED PENTAX MEDICAL CUSTOMER SERVICE ON (B)(6) 2021 FOR AN ENDOSCOPE THAT "CANNOT PASS ACCESSORY TO INLET PORT,AND CUSTOMER STATED CAN'T PASS CLEANING BRUSH ON BIOPSY CHANNEL" INVOLVING PENTAX MEDICAL SMBOWEL ENTEROSCOPE MODEL VSB-3430K, SERIAL NUMBER (B)(4). THREE GOOD FAITH EFFORT(GFE) ATTEMPTS WERE MADE WITH NO RESPONSE FROM THE USER FACILITY. THE CUSTOMER OWNED COLONOSCOPE WAS RETURNED TO PENTAX MEDICAL FOR EVALUATION ON 14-JUN-2021. THE ENDOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE UNDER SERVICE ORDER (B)(4) AND THE TECHNICIAN CONFIRMED THE CUSTOMERS EXPERIENCE WHEN THEY FOUND A "CHECK VALVE STUCK INSIDE BIOPSY INLET T-PIECE" AND DOCUMENTED THE FOLLOWING INSPECTION FINDINGS ON 15-JUN-2021: UMBILICAL CABLE BUMP AT PVE SIDE, PASSED DRY LEAK TEST, PASSED WET LEAK TEST. DURING QC TESTING THE "GROUND TERMINAL FAIL BIOPSY INLET CONTINUITY TEST" ON 29-JUN-2021. THE DEVICE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: BENDING RUBBER, ADJUSTING COLLAR, UD PULLEY ASSY, RL PULLEY ASSY, ANGLE GUIDE, O-RINGS AND SEALS. MODEL VSB-3430K, SERIAL NUMBER (B)(4), HAS BEEN ROUTINELY SERVICED AT A PENTAX FACILITY SINCE THE DEVICE WAS PUT INTO SERVICE ON 07-AUG-20
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12173345 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K962897. One submission often covers many models, so check the product and lot details in each record.
- Z-2071-2016 · Apr 6, 2016 · TerminatedGastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page