VIDEO GASTROSCOPE EG-2990IRecalls & reported problems
Pentax Of America, Inc. · HD VIDEO GASTROSCOPE 2.8C 9.8T 1050L (REFURBISHED) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04961333232512), received between Jun 23, 2021 and Jun 23, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Reprocessing Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K131902); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Reprocessing Problem1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2021Device Reprocessing Problem; Physical Resistance/Sticking; Device Reprocessing Problem; Physical Resistance/Sticking
PENTAX MEDICAL WAS MADE AWARE OF AN EVENT ON (B)(6) 2021 THAT OCCURRED AFTER REPROCESSING IN THE UNITED STATES. THE REPORTED COMPLAINT THAT THE ENDOSCOPE FAILED ADENOSINE TRIPHOSPHATE(ATP) TEST AFTER REPROCESSING INVOLVING PENTAX MEDICAL VIDEO GASTROSCOPE, MODEL EG-2990I, SERIAL NUMBER (B)(4). ALTHOUGH THE USER FACILITY WAS ABLE TO CLEAN THE ENDOSCOPE, THERE WAS RESISTANCE EXPERIENCED AND THEY WERE NOT SURE IF SOMETHING WAS STUCK IN THE CHANNEL. NO PATIENT INVOLVEMENT WAS REPORTED. THE USER FACILITY RESPONDED TO GOOD FAITH EFFORT EMAILS ON 10-JUN-2021 DETAILING THE METHOD FOR REPROCESSING, CHEMICALS USED FOR REPROCESSING, AS WELL AS TESTING AND STORAGE DETAILS. THE CUSTOMER OWNED ENDOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 03-JUN-2021. THE ENDOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE UNDER SERVICE ORDER 3133570 AND THE TECHNICIAN DID NOT IDENTIFY ANYTHING STUCK WITHIN THE ENDOSCOPE BUT DID NOTE "UMBILICAL CABLE SEVERE CRUSH" AND "SUCTION TUBE RESISTANCE" AND ALSO DOCUMENTED THE FOLLOWING INSPECTION FINDINGS ON (B)(6) 2021: PASSED DRY LEAK TEST, BENDING RUBBER GLUE CRACKING AT INSERTION TUBE SIDE, BENDING RUBBER GLUE CRACKING AT DISTAL SIDE, # 1 REMOTE CONTROL BU
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12051941 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131902. One submission often covers many models, so check the product and lot details in each record.
- Z-0318-2017 · Aug 15, 2016 · TerminatedPentax Video Gastroscope is intended to be used with a Pentax video processor (including light source), documentation equipment, monitor, EndoTherapy Device such as a Biopsy Forcep
- Z-2071-2016 · Apr 6, 2016 · TerminatedGastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, ti
Other Gastroscope And Accessories, Flexible/Rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page