VIDEO GASTROSCOPE EG27-V10CRecalls & reported problems
Pentax Of America, Inc. · VIDEO GASTROSCOPE - PTAH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04961333244676), received between Nov 13, 2020 and Nov 13, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Failure to Disconnect” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Disconnect1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2020Failure to Disconnect; Failure to Disconnect
DURING EVALUATION BY THE PENTAX MEDICAL SERVICE REPAIR DEPARTMENT ON 06-FEB-2020, AN "ACCESSORY STUCK IN PRIMARY OPERATION CHANNEL" WAS IDENTIFIED INVOLVING MODEL EG27-V10C, SERIAL NUMBER (B)(4), AND THE COMPLAINT HANDLING UNIT WAS NOTIFIED. ON 26-MAY-2020, COMMUNICATIONS BETWEEN PENTAX MEDICAL COMPLAINT HANDLING UNIT AND MARKETING PRODUCT MANAGER ALONG WITH THE ASSETS AND USED EQUIPMENT COORDINATOR ESTABLISHED THE ENDOSCOPE WAS DELIVERED TO (B)(6) ON (B)(6) 2019 AS A DEMO UNIT AND RETURNED ON 08-AUG-2019. MARKETING CONFIRMED THE DEVICE WAS NOT AVAILABLE FOR SALES IN THE UNITED STATES AND THAT THERE IS NO KNOWN PATIENT INVOLVEMENT INVOLVING THIS DEMO UNIT, AND NO PATIENT RISK. THE PENTAX MEDICAL SERVICE REPAIR TECHNICIAN DOCUMENTED AN "ACCESSORY STUCK IN PRIMARY OPERATION CHANNEL" DOCUMENTED UNDER SERVICE ORDER (B)(4) ON 06-FEB-2020. OTHER INSPECTIONAL FINDINGS INCLUDED: PRIMARY OPERATION CHANNEL RESISTANCE, PASSED WET LEAK TEST, PASSED DRY LEAK TEST. THE INVESTIGATION AND REPAIR IS IN-PROCESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10840098 ↗
Other Gastroscope And Accessories, Flexible/Rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.