Skip to content

PENTAX Medical EG29-i10c (UDI 04961333248483)Recalls & reported problems

Hoya Corporation · PENTAX Medical Video Upper GI Scope EG29-i10c (Made in Malaysia) · Class 2

All models of PENTAX Medical (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Pentax upper GI video scope is an endoscope used to look inside the upper digestive tract. The FDA records hold 96 reports, with no deaths and no injuries; many describe video image failures such as blackouts or cloudy images, and some describe fluid damage. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
95
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 95 adverse event reports that list this exact device (DI 04961333248483), received between Jul 1, 2021 and May 8, 2026 — classified as 95 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image15
  • Optical Distortion14
  • Fluid/Blood Leak13
  • Loss of or Failure to Bond12
  • Improper or Incorrect Procedure or Method10
  • Electrical Overstress7
  • Leak/Splash6
  • Moisture or Humidity Problem6

Patient problems

  • No Clinical Signs, Symptoms or Conditions94
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 95 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25120208 ↗
  • MalfunctionApr 6, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24787279 ↗
  • MalfunctionMar 11, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24565972 ↗
  • MalfunctionJan 7, 2026Poor Quality Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23986886 ↗
  • MalfunctionJul 23, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22573321 ↗
  • MalfunctionJul 15, 2025Poor Quality Image; Poor Quality Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22503977 ↗
  • MalfunctionJul 14, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22491555 ↗
  • MalfunctionJun 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER NOZZLE MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22164901 ↗
  • MalfunctionJun 2, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22121421 ↗
  • MalfunctionMay 21, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22058205 ↗
  • MalfunctionMar 3, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21501549 ↗
  • MalfunctionFeb 6, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21318885 ↗
  • MalfunctionOct 3, 2024Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20356865 ↗
  • MalfunctionSep 24, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20289356 ↗
  • MalfunctionSep 10, 2024Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20178206 ↗
  • MalfunctionAug 28, 2021Fluid/Blood Leak; Fluid/Blood Leak

    PENTAX MEDICAL WAS MADE AWARE OF A COMPLAINT WHICH OCCURRED IN THE EMEA REGION DURING INSPECTION IN THE WORKSHOP WITH THE REPORTING "THE ELECTRICAL CONNECTING PINS ON THE LG-PLUG ARE LEAKY. ELECTRICAL SAFETY TEST FAILED." INVOLVING PENTAX MEDICAL VIDEO GASTROSCOPE MODEL EG29-I10C, SERIAL NUMBER (B)(4). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR OTHER SIGNIFICANT/IMPORTANT MEDICAL EVENT. THE DEFECTIVE PARTS WILL BE REPLACED AS PART OF THE SERVICE REQUEST. THIS EVENT IS FDA REPORTABLE DUE TO THE LACK OF INFORMATION TO JUSTIFY THAT THIS EVENT WOULD NOT CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR OTHER ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12381165 ↗
  • MalfunctionAug 27, 2021Fluid/Blood Leak; Poor Quality Image; Electrical Overstress; Fluid/Blood Leak; Poor Quality Image; Electrical Overstress

    MEDICAL WAS MADE AWARE OF THE FOLLOWING COMPLAINT THAT OCCURRED IN THE EMEA REGION. (B)(6) FOUND MANY BLACK DOTS IN THE IMAGE AND WHEN INSPECTED IN THE WORKSHOP THE ELECTRICAL SAFETY TEST FAIL AND THEY FOUND A LEAKY TUBE INVOLVING PENTAX MEDICAL VIDEO GASTROSCOPE MODEL EG29-I10C, SERIAL NUMBER (B)(4). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR OTHER SIGNIFICANT/ IMPORTANT MEDICAL EVENT. THE ENDOSCOPE WILL BE REPAIRED AS PART OF THE SERVICE REPAIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12379819 ↗
  • MalfunctionAug 21, 2021Loss of or Failure to Bond; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Improper or Incorrect Procedure or Method

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ELECTRICAL SAFETY TEST FAIL. DETECTED AT INCOMING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351829 ↗
  • MalfunctionAug 21, 2021Electrical /Electronic Property Problem; Optical Distortion; Electrical /Electronic Property Problem; Optical Distortion

    THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM. ISO FAILED ; MISTY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351828 ↗
  • MalfunctionAug 21, 2021Loss of or Failure to Bond; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Improper or Incorrect Procedure or Method

    THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM. DIELECTRIC STRENGTH TEST FAILED. DETECTED AT INCOMING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351827 ↗
  • MalfunctionAug 21, 2021Loss of or Failure to Bond; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Improper or Incorrect Procedure or Method

    THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM. ISOLATION TEST - FAILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351825 ↗
  • MalfunctionAug 21, 2021Electrical /Electronic Property Problem; Optical Distortion; Material Split, Cut or Torn; Electrical /Electronic Property Problem; Optical Distortion; Material Split, Cut or Torn

    THIS EVENT OCCURRED AT THE TIME OF BEFORE USE. THERE WAS NO REPORT OF PATIENT HARM. ISO FAILED, MISTY IMAGE, LED COVER GLASS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351824 ↗
  • MalfunctionAug 21, 2021Loss of or Failure to Bond; Break; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Break; Improper or Incorrect Procedure or Method

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. OBJECTIVE LENS BROKEN. ELECTRICAL SAFETY TEST FAILED (DETECTED AT WORKSHOP INSPECTION).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351823 ↗
  • MalfunctionAug 21, 2021Loss of or Failure to Bond; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Improper or Incorrect Procedure or Method

    THIS EVENT OCCURRED DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM. ELECTRICAL SAFETY TEST FAIL. DETECTED AT INCOMING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351822 ↗
  • MalfunctionAug 21, 2021Loss of or Failure to Bond; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Improper or Incorrect Procedure or Method

    THIS EVEN OCCURRED DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM. ELECTRICAL SAFETY TEST FAILED, FOUND DURING THE INCOMING INSPECTION BY VARIMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351820 ↗

Other Gastroscope And Accessories, Flexible/Rigid devices

All Gastroscope And Accessories, Flexible/Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page