GLIDEWIRE GS3510Recalls & reported problems
Terumo Corporation · RADIFOCUS GLIDEWIRE · Class 2
All models of GLIDEWIRE (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04987350773678), received between Jan 7, 2022 and Feb 3, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Advance” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance2
- Material Deformation2
- Material Separation2
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 3, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation
THE USER FACILITY REPORTED THEY WERE UNABLE TO BACKLOAD A CATHETER ONTO THE WIRE. WHEN TRYING TO BACKLOAD THE CATHETER ON TO THE WIRE, THE HYDROPHILIC MATERIAL AT THE BACK OF THE WIRE WOULD BUNCH UP AND PREVENT THE CATHETER FROM BEING ADVANCED. THE WIRE WAS REMOVED AND A NEW GLIDEWIRE WAS INSERTED. THE CASE WAS COMPLETED WITH NO FURTHER ISSUES. NO BLOOD LOSS. THERE WAS NO PATIENT INJURY AND/OR REQUIRE MEDICAL OR SURGICAL INTERVENTION. NO BLOOD LOSS WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 09JAN2025: THE PROCEDURE PERFORMED WAS AN AORTAGRAM WITH RUNOFF OF LOWER LEG. THERE WAS NO IMPACT TO THE PATIENT FROM THE HYDROPHILIC MATERIAL BUNCHING UP. THE WIRE WAS EXPOSED WHEN THE HYDROPHILIC MATERIAL BUNCHED UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293987 ↗InjuryFeb 3, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation
THE USER FACILITY REPORTED THEY WERE UNABLE TO BACKLOAD A CATHETER ONTO THE WIRE. WHEN TRYING TO BACKLOAD THE CATHETER ON TO THE WIRE, THE HYDROPHILIC MATERIAL AT THE BACK OF THE WIRE WOULD BUNCH UP AND PREVENT THE CATHETER FROM BEING ADVANCED. THE WIRE WAS REMOVED AND A NEW GLIDEWIRE WAS INSERTED. THE CASE WAS COMPLETED WITH NO FURTHER ISSUES. NO BLOOD LOSS. THERE WAS NO PATIENT INJURY AND/OR REQUIRE MEDICAL OR SURGICAL INTERVENTION. NO BLOOD LOSS WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 09JAN2025: THE PROCEDURE PERFORMED WAS AN AORTAGRAM WITH RUNOFF OF LOWER LEG. THERE WAS NO IMPACT TO THE PATIENT FROM THE HYDROPHILIC MATERIAL BUNCHING UP. THE WIRE WAS EXPOSED WHEN THE HYDROPHILIC MATERIAL BUNCHED UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293331 ↗MalfunctionNov 7, 2024Peeled/Delaminated; Peeled/Delaminated
THE USER FACILITY REPORTED THAT DURING A PERIPHERAL INTERVENTION USING A 6FR SHEATH IN THE RIGHT COMMON FEMORAL ARTERY (CFA), IT WAS OBSERVED THAT THE WIRE BEGAN TO DELAMINATE AFTER INITIAL ACCESS. THE WIRE WAS SUBSEQUENTLY REMOVED FROM THE SHEATH. ADDITIONAL INFORMATION RECEIVED ON 09 OCT 2024: IT WAS CONFIRMED THAT NO WIRE FRAGMENTS WERE LEFT IN THE PATIENT, AND THE PATIENT REMAINED STABLE. A NEW STIFF WIRE WAS USED TO REPLACE THE DEFECTIVE ONE, ALLOWING THE PROCEDURE TO CONTINUE SUCCESSFULLY. THE SAMPLE WAS CONTAMINATED BY BLOOD AND BODY FLUID BUT WAS NOT CONTAMINATED BY INFECTIOUS AGENTS OR ANTI-CANCER AGENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20627135 ↗InjuryJan 7, 2022Material Separation; Material Separation
THE USER FACILITY REPORTED THAT THE GLIDEWIRE SEVERED IN HALF AS THE DOCTOR WAS SPINNING THE WIRE THRU A TIGHT LESION. HE WAS ABLE TO SNARE THE WIRE WITHOUT CAUSING HARM TO THE PATIENT. THE PATIENT WAS STABLE. THE PROCEDURE WAS SUCCESSFUL, HOWEVER WAS A PROLONGED CASE. ADDITIONAL INFORMATION WAS RECEIVED ON 16DECEMBER 2021: THE PROCEDURE WAS AN AORTOGRAM, PELVIC ANGIOGRAM, AND LEFT SUPERFICIAL FEMORAL ARTERY (L SFA) WITH LITHOTRIPSY AND SELF-EXPANDING STENT PLACEMENT. ADDITIONAL INFORMATION WAS RECEIVED ON 17DECEMBER 2021: NO FRAGMENT WAS LEFT IN THE PATIENT. THE DOCTOR ONLY HAD TO SNARE THE WIRE TO REMOVE IT. IT IS UNKNOWN IF THERE WAS ANY OTHER EQUIPMENT OR DEVICES USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13196512 ↗InjuryJan 7, 2022Material Separation; Material Separation
THE USER FACILITY REPORTED THAT THE GLIDEWIRE SEVERED IN HALF AS THE DOCTOR WAS SPINNING THE WIRE THRU A TIGHT LESION. HE WAS ABLE TO SNARE THE WIRE WITHOUT CAUSING HARM TO THE PATIENT. THE PATIENT WAS STABLE. THE PROCEDURE WAS SUCCESSFUL, HOWEVER WAS A PROLONGED CASE. ADDITIONAL INFORMATION WAS RECEIVED ON 16DECEMBER 2021: THE PROCEDURE WAS AN AORTOGRAM, PELVIC ANGIOGRAM, AND LEFT SUPERFICIAL FEMORAL ARTERY (L SFA) WITH LITHOTRIPSY AND SELF-EXPANDING STENT PLACEMENT. ADDITIONAL INFORMATION WAS RECEIVED ON 17DECEMBER 2021: NO FRAGMENT WAS LEFT IN THE PATIENT. THE DOCTOR ONLY HAD TO SNARE THE WIRE TO REMOVE IT. IT IS UNKNOWN IF THERE WAS ANY OTHER EQUIPMENT OR DEVICES USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13196485 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.