Vantage Orian MRT-1550Recalls & reported problems

Canon Inc. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 10, 2024Open, Classified

    The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

    Z-2041-2024 · Root cause: Device Design · FDA record ↗

  2. Jan 29, 2024Open, Classified

    For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

    Z-1248-2024 · Root cause: Component change control · FDA record ↗

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →