taperflo™ W40T0406SRecalls & reported problems
Vascutek Ltd · taperflo Gelatin Sealed ePTFE Vascular Prosthesis Short Taper with Wrap · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05037881009599), received between Mar 22, 2018 and Aug 22, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Puncture/Hole2
- Material Integrity Problem1
- Material Perforation1
- Occlusion Within Device1
Patient problems
- Hematoma1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2018Material Puncture/Hole; Material Perforation; Material Integrity Problem; Material Puncture/Hole; Material Perforation; Material Integrity Problem
THE EVENT WAS REPORTED TO VASCUTEK ON (B)(6) 2018 AS FOLLOWS: "BLOOD LEAKAGE: ON (B)(6) 2018: THE TAPERFLO WAS IMPLANTED AT THE (B)(6) HOSPITAL. AFTER THE SHUNT WAS MONITORED FOR ABOUT THREE WEEKS AT THE HOSPITAL, THE PATIENT WAS TRANSFERRED TO A DIALYSIS FACILITY AND CONTINUED THE DIALYSIS TREATMENT. ON (B)(6) 2018: A HEMATOMA WAS OBSERVED DUE TO BLOOD LEAKAGE FROM THE LOOP PORTION OF THE TAPERFLO. THE PATIENT WAS TRANSFERRED TO THE (B)(6) HOSPITAL AND THE TAPERFLO WAS REMOVED FROM THE PATIENT. THE PATIENT IS CURRENTLY BEING MONITORED AND THE CONDITION IS STABLE AFTER THE PROCEDURE. OPERATION TYPE: FEMORAL ARTERY ARTIFICIAL BLOOD VESSEL REPLACEMENT SURGERY." FURTHER INFORMATION: "SINCE THE PATIENT WAS RECEIVING DIALYSIS WITHOUT PROBLEMS AFTER IMPLANTING THE TAPERFLO, THE HOSPITAL CONFIRMED THAT THE DIALYSIS FACILITY DID NOT PUNCTURE THE LOOP PORTION. HOWEVER, THE HOSPITAL WOULD LIKE TO KNOW THE CONDITION ON THE LOOP, SUCH AS PUNCTURE MARKS ON THE LOOP." IT WAS REPORTED THAT THERE WAS NO IMPACT ON PATIENT HEALTH AND EVENT WAS POSSIBLY RELATED TO DEVICE.
Patient problems: Hematoma
MAUDE report 7804183 ↗MalfunctionMar 22, 2018Occlusion Within Device; Material Puncture/Hole; Occlusion Within Device; Material Puncture/Hole
TERUMO (B)(4) REPORTED GRAFT OCCLUSION: THE 10TH DAY AFTER IMPLANT, PUNCTURING WAS PERFORMED SINCE THE CONDITION OF THE PATIENT LOOKED GOOD WITH NO SWELLING. THE PATIENT WAS THEN DISCHARGED FROM THE HOSPITAL AND SENT BACK TO THE CLINIC IN WHICH THE PATIENT HAD BEEN HOSPITALIZED ORIGINALLY. THE 37TH DAY AFTER IMPLANT, THE PATIENT WAS SENT BACK TO THE HOSPITAL AGAIN BECAUSE THE GRAFT WAS OCCLUDED. WHEN THE GRAFT WAS EXPLANTED, THE FOLLOWING WERE FOUND: JELLY-LIKE SUBSTANCES SURROUNDING THE GRAFT DID NOT ADHERE TO THE GRAFT. THE GRAFT WAS FLAT, AND WHEN THE GRAFT WAS PUNCTURED WITH A NEEDLE, THE NEEDLE PERFORATED THE FRONT AND REAR WALLS OF THE GRAFT. A LOW-ECHOIC AREA OF SEROMA-LIKE SUBSTANCES WAS OBSERVED AROUND THE ENTIRE LENGTH OF THE GRAFT WHICH WAS CRUSHED FLATLY. WHEN THE GRAFT WAS REPLACED WITH A THORATEC GRAFT, THE IMPLANTED GRAFT WAS EASILY REMOVED WITHOUT ADHERING TO THE TISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7361901 ↗
Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices
- Cardioroot HEWROOT0030Intervascular Sas · Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 recall2 deaths13 reports
- FLIXENE SLIDER GDS 25058Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall13 reports
- ADVANTA SLIDER GDS 22114Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall11 reports
- Cardioroot HEWROOT0032Intervascular Sas · Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 recall2 deaths10 reports
- FLIXENE SLIDER GDS 25129Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall9 reports
- FLIXENE SLIDER GDS 25137Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall8 reports
- ADVANTA SLIDER GDS 22222Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall8 reports
- FLIXENE SLIDER GDS 25138Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page