LAMINAR OPOS20LRecalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · AMO 20 GAUGE IRRIGATION SLEEVE AND TEST · Class 2
All models of LAMINAR (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474500389), received between Dec 11, 2019 and Nov 21, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Flushing Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Flushing Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2022Material Deformation; Material Deformation
IT WAS REPORTED THAT PHACO TIP WAS LOOSENED DURING PHACO PROCEDURE. CUSTOMER WAS ABLE TO DETACH THE TIP FROM HANDPIECE BY HAND. THE ALLEGED PRODUCT WAS REPLACED WITH A BRAND NEW PHACO TIP. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE BACKUP. THERE WAS NO PATIENT INJURY. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15840902 ↗InjuryDec 11, 2019Flushing Problem; Flushing Problem
A HOSPITAL REPORTED A CATARACT PATIENT WAS PRESENTED WITH TOXIC ANTERIOR SEGMENT SYNDROME (TASS) WHEN USING A PHACO HANDPIECE, TIPS AND SLEEVE. A BRIEF DESCRIPTION FROM THE HOSPITAL THAT AT A ONE-DAY FOLLOW UP EXAM, IT WAS NOTED THE PATIENT HAD INFLAMMATION LASTING 24 HOURS. ANTIBACTERIAL INFUSION AND AN ANTERIOR CHAMBER WASH WERE PERFORMED. A CULTURE TEST WAS PERFORMED. THE TEST FOUND NO BACTERIA WAS PRESENT, HENCE, THE SURGEON CONSIDERED TASS AS THE DIAGNOSIS. THE CLEANING AND STERILIZATION PRACTICES WERE REVIEWED. PREVIOUSLY, ULTRASONIC CLEANING WAS NOT USED, AND A VACUUM BOILING TYPE WASH WAS USED FOR DECONTAMINATING INSTRUMENTS. IN THE FUTURE, THE TIP AND SLEEVE WILL BE SWITCHED FROM REUSE TO SINGLE USE. THE ACCESSORIES WILL BE CLEANED AND STERILIZED. IN ADDITION, THE HOSPITAL WILL BEGIN THOROUGHLY FLUSHING OUT INSTRUMENTS AFTER EACH CASE. IT WAS REPORTED THE PATIENT OUTCOME IS WELL AND HAS RECOVERED. THE HOSPITAL INDICATED THE HANDPIECE, TIPS AND SLEEVE WILL NOT BE RETURNED. THIS IS 1 OF 4 REPORTS. THIS REPORT IS FOR THE SLEEVE. SEPARATE REPORTS WILL BE SUBMITTED FOR THE HANDPIECE AND TIPS.
Patient problems: No Code Available
MAUDE report 9455757 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page