DISPOSABLE 25GA VITRECTOMY CUTTER NGP0025Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · DISPOSABLE 25GA VITRECTOMY CUTTER · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05050474534117), received between Mar 24, 2017 and Aug 9, 2018 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (7 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Sep 24, 2018Class IITerminated
Packaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.
Z-1893-2019 · Root cause: Packaging process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Markings/Labelling Problem7
- Incorrect Device Or Component Shipped3
Patient problems
- No Patient Involvement7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 9, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25-GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20-GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 2 OF 2 REPORTS.
Patient problems: No Patient Involvement
MAUDE report 7767518 ↗MalfunctionAug 9, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25-GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20-GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 1 OF 2 REPORTS.
Patient problems: No Patient Involvement
MAUDE report 7767478 ↗MalfunctionAug 9, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25-GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20-GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 2 OF 2 REPORTS.
Patient problems: No Patient Involvement
MAUDE report 7767455 ↗MalfunctionAug 9, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25-GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20-GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 1 OF 2 REPORTS.
Patient problems: No Patient Involvement
MAUDE report 7767390 ↗MalfunctionJun 14, 2017Device Markings/Labelling Problem; Incorrect Device Or Component Shipped; Device Markings/Labelling Problem; Incorrect Device Or Component Shipped
A SURGERY CENTER REPORTED MISLABELING OF A DISPOSABLE VITRECTOMY CUTTER. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25 GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20 GAUGE ENGRAVED ON THE DEVICE. NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 6641789 ↗MalfunctionMar 24, 2017Device Markings/Labelling Problem; Incorrect Device Or Component Shipped; Device Markings/Labelling Problem; Incorrect Device Or Component Shipped
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25 GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20 GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 2 OF 2 REPORTS.
Patient problems: No Patient Involvement
MAUDE report 6430698 ↗MalfunctionMar 24, 2017Device Markings/Labelling Problem; Incorrect Device Or Component Shipped; Device Markings/Labelling Problem; Incorrect Device Or Component Shipped
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25 GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20 GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 1 OF 2 REPORTS.
Patient problems: No Patient Involvement
MAUDE report 6430697 ↗
Other VITRECTOMY, INSTRUMENT CUTTER devices
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752450Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls1,160 reports
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752437Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls1,023 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065000095Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls616 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065753106Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls546 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065000096Alcon Laboratories, Inc. · Vitrectomy, instrument cutter1 recall554 reports
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752451Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls258 reports
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752438Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls203 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065753109Alcon Laboratories, Inc. · Vitrectomy, instrument cutter1 recall223 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page