COMPACT INTUITIV OPO80Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · PHACO PACK, APOLLO COMPACT, SINGLE USE · Class 2

All models of COMPACT INTUITIV (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Johnson & Johnson single-use phaco pack is used in cataract surgery to break up and remove the cloudy lens. Most of the 74 reports describe unsealed packaging or holes in the lid, with surgery centers concerned the tubing set was no longer sterile; a few mention suction problems, and one describes increased eye pressure. There are no deaths and 1 injury. One recall, open since December 2022, concerns a gap between the housing and pack cover that could affect priming and vacuum during surgery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
74
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 74 adverse event reports that list this exact device (DI 05050474573352), received between Aug 2, 2018 and May 19, 2020 — classified as 1 injury, 73 malfunctions.

The most frequently reported specific device problem is “Unsealed Device Packaging” (70 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Dec 7, 2022Class IIOpen, Classified

    Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

    Z-1007-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unsealed Device Packaging70
  • Suction Problem4

Patient problems

  • No Patient Involvement70
  • No Consequences Or Impact To Patient3
  • Intraocular Pressure Increased1

Recent adverse event reports

The 25 most recent of 74 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 5 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073262 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 3 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073256 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 4 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073245 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 2 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073237 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 1 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073231 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUITIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 3 OF 3 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073226 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUITIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 2 OF 3 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073224 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUITIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 1 OF 3 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073222 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE.

    Patient problems: No Patient Involvement

    MAUDE report 10073202 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE.

    Patient problems: No Patient Involvement

    MAUDE report 10073186 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 5 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073180 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 4 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073171 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 3 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073160 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 2 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073145 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 1 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073137 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 5 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073124 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 4 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073107 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 3 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073084 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 2 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073077 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 1 OF 5 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073061 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 4 OF 4 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073024 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 3 OF 4 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10073013 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 2 OF 4 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10072992 ↗
  • MalfunctionMay 19, 2020Unsealed Device Packaging; Unsealed Device Packaging

    A SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. THIS IS 1 OF 4 REPORTS.

    Patient problems: No Patient Involvement

    MAUDE report 10072972 ↗
  • MalfunctionMar 16, 2020Unsealed Device Packaging; Unsealed Device Packaging

    THE SURGERY CENTER REPORTED OBSERVING HOLES WORN INTO THE TYVEK LID OF A COMPACT INTUTIV PACK DISPOSABLE TUBING SET MODEL OPO80. THE SURGERY CENTER HAD CONCERNS THE PACKAGING HAD BEEN COMPRISED AND THE TUBING PACK WAS NO LONGER STERILE. A VISUAL INSPECTION CONFIRMED A PUNCTURE HOLE ON THE TYVEK LID.

    Patient problems: No Patient Involvement

    MAUDE report 9835139 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the COMPACT INTUITIV OPO80

Has the COMPACT INTUITIV OPO80 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1007-2023 (Class II), began on Dec 7, 2022: Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the COMPACT INTUITIV OPO80?

The FDA's MAUDE database lists 74 adverse event reports for this device, received between Aug 2, 2018 and May 19, 2020. In those reports, the most common specific device problems are “Unsealed Device Packaging” (70) and “Suction Problem” (4); the most common patient problem is “Intraocular Pressure Increased” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the COMPACT INTUITIV OPO80?

None of the reports are classified as deaths, 1 as an injury and 73 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the COMPACT INTUITIV OPO80, and is it still sold?

The FDA's device database lists Johnson & Johnson Surgical Vision, Inc. as the company responsible for it (brand COMPACT INTUITIV, model OPO80). Its record shows it as in commercial distribution.