HEALON DUET 10290080Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · HEALON DUET US · Class 3
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05050474573574), received between Jan 10, 2017 and Apr 14, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Appropriate Device Problem Term/Code Not Available1
- Device Contamination with Chemical or Other Material1
- Device Difficult to Setup or Prepare1
- Mechanical Problem1
Patient problems
- Corneal Clouding/Hazing1
- Eye Burn1
- Eye Injury1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Information1
- Visual Impairment1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 14, 2026Mechanical Problem; Device Difficult to Setup or Prepare; Appropriate Device Problem Term/Code Not Available
THE DOCTOR REPORTED AN ISSUE WITH A NEW CANNULA, NOTING DIFFICULTY IN PASSING IT THROUGH A 1-MILLIMETER SIDE PORT INCISION DUE TO ITS LARGER SIZE COMPARED TO THE PREVIOUS VERSION AND AN UNPOLISHED SURFACE THAT CAUSED IT TO CATCH DURING USE. HE OBSERVED THAT IN TWO INSTANCES, THE CANNULA RELEASED OR POPPED THROUGH THE INCISION, WITH ONE OCCURRENCE INVOLVING CONTACT WITH THE CAPSULAR BAG; HOWEVER, THIS DID NOT RESULT IN ANY COMPLICATIONS. ALL CASES WERE COMPLETED AND NO PRODUCT TO RETURN. THE DOCTOR STATED THAT WHILE THE ISSUE WAS NOT CRITICAL FOR HIM PERSONALLY, IT MIGHT POSE CHALLENGES FOR HIS COLLEAGUES AND INQUIRED ABOUT THE AVAILABILITY OF SMOOTHER CANNULAS IF NECESSARY. HE ALSO TESTED THE HEALON PRO CANNULA, WHICH FUNCTIONED BUT REQUIRED ADDITIONAL FORCE AND SEPARATED INTO LAYERS DURING USE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24880288 ↗InjuryApr 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTED THAT THE NEEDLE STARTED SMOKING INSIDE THE PATIENT'S EYE, WHICH CAUSED THE IRIS TO FLUFF AND LED TO A CORNEAL BURN. CUSTOMER NOTED THAT SHE DID NOT HEAR THE BEEP THAT USUALLY SIGNALS A RISE IN TEMPERATURE. ADDITIONAL INFORMATION WAS RECEIVED, PATIENT REQUIRED EXTENSIVE SUTURING AT THE END OF THE CASE TO ACHIEVE WOUND CLOSURE. SUTURES WERE REMOVED. PATIENT RESULTED WITH A CORNEAL SCAR, ASTIGMATISM BUT CORRECTABLE TO 20/20. NO ADDITIONAL SURGERY REQUIRED. PHACO SETTINGS: VACUUM LOW: 200, HIGH: 350, POWER: 40%. WHITESTAR ON YES. MODE & SUBSETTINGS DESIGNATIONS, PERISTALTIC, CONTINUOUS, 8/4, 67%, 83PPS. BOTTLE HEIGHT: 100 IN/CM. INCISION SIZE: 2.4MM. DOCTOR'S PHACO TECHNIQUE: DIVIDE & CONCUR. MODEL & SERIAL NUMBER OF EQUIPMENT: CONSOLE: SIGNATURE PRO, SERIAL NUMBER: (B)(6) (INCORRECT NUMBER), HANDPIECE: ELLIPS FX, SN# NOT AVAILABLE. TUBING TYPE: OPO73, TIP: 21G BLUE. VISCOELASTIC: DUET. THIS REPORT IS FOR HEALON DUET. OTHER REPORTS WERE SUBMITTED FOR WHITESTAR, PHACO HANDPIECE, PHACO TIP, PHACO TUBING. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Corneal Clouding/Hazing; Eye Burn; Eye Injury; Visual Impairment
MAUDE report 21883702 ↗InjuryApr 14, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS INITIALLY REPORTED THAT THE DOCTOR ''HAD A ABOUT 3 PATIENTS WITH TOXIC ANTERIOR SEGMENT SYNDROME (TASS) ON (B)(6) 2018. IT WAS LATER CONFIRMED THAT AT LEAST 1 PATIENT WAS AFFECTED. IN ADDITION, THE DOCTOR BELIEVES THE HEALON DUET MAY NOT HAVE CAUSED OR CONTRIBUTED TO THE TASS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Code Available
MAUDE report 7431538 ↗MalfunctionJan 10, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT A BLACK THREAD WAS FOUND IN THE PATIENT'S EYE COMING FROM THE VISCOELASTIC HEALON ENDOCOAT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Information
MAUDE report 6239598 ↗
Other AID, SURGICAL, VISCOELASTIC devices
- HEALON ENDOCOAT PRO 10300012Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC1 recall1 report
- HEALON DUET PRO II 10300016Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC1 recall0 reports
- HEALON ENDOCOAT VT585UJohnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC22 reports
- HEALON 5 10290051Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC20 reports
- HEALON GV PRO 10240014Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC15 reports
- Amvisc 59081LBausch & Lomb Incorporated · Aid, Surgical, Viscoelastic10 reports
- HEALON DUET PRO 10240016Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC9 reports
- DUOVISC 8065183150Alcon Laboratories, Inc. · Aid, Surgical, Viscoelastic9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.