YelloPort Plus - Laparoscopic Port Access System YA05VSS01Recalls & reported problems
Surgical Innovations Limited · 5mm Single Use Duckbill Valve · Class 2
All models of YelloPort Plus - Laparoscopic Port Access System (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05051986002347), received between Jul 23, 2018 and Jul 30, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K070712); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Gas/Air Leak1
- Material Fragmentation1
- Material Separation1
- Melted1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 30, 2026Material Fragmentation
Description: During the closure phase of the TAPP procedure, a missing valve was noticed; a search for the missing part was conducted but it was not found within the abdominal cavity, and the abdomen was closed.The patient is to be kept under observation. The valve has been discarded.Combined equipment: Forceps by STORZ.Confirmed: The valve was damaged during the surgery, and a part of the valve was missing. Also, the damaged valve was discarded in the hospital.Conclusion after investigation: Based on previous complaint investigations, this failure mode has only been observed where instrument tips were slightly open during insertion and subsequently pushed through the valve, resulting in a tear. Given the consistency of this failure mechanism with previous investigations, the root cause is concluded to be user error during instrument insertion. Given the frequency of reported events over the previous 12 months, a CAPA will be raised to further investigate this issue and implement appropriate measures to prevent future occurrences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50082146 ↗MalfunctionOct 21, 2020Material Separation
BEFORE THE LAP PROCEDURE, NURSE OPENED THE PACKAGE AS USUAL AND CHECKED THE VALVE BRIEFLY. HOWEVER, AFTER THE PROCEDURE COMPLETED, THE NURSE DISSEMBLED THE REUSABLE AND DISPOSABLE PARTS OF TROCAR, SHE FOUND THERE WAS MISSING PORTION ON THAT VALVE AND THEY COULDN'T FIND IT ANYWHERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10710626 ↗MalfunctionNov 5, 2019Break; Melted
DURING SURGERY, A PROBE WAS INSERTED, AND PULLED OUT THROUGH YA05VSS01 SILICONE VALVE, AND OUR CUSTOMER FOUND THE PRODUCT WAS MELTED, AND BROKEN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9279092 ↗MalfunctionJul 23, 2018Break; Gas/Air Leak
DURING A PROCEDURE, THE DOCTOR CONFIRMED AIR LEAKAGE AND THE VALVE WAS DAMAGED. IT HAPPENED 5MINS AFTER THE BEGINNING OF SURGERY. HE/ SHE USED A GRASPING FORCEPS AND INSERTED JUST ONCE. THE USER COULD NOT FIND THE BROKEN PARTS. DURING A PROCEDURE, THE DOCTOR CONFIRMED AIR LEAKAGE AND THE VALVE WAS DAMAGED. IT HAPPENED 5MINS AFTER THE BEGINNING OF SURGERY. HE/ SHE USED A GRASPING FORCEPS AND INSERTED JUST ONCE. THE USER COULD NOT FIND THE BROKEN PARTS.
Patient problems: No Information
MAUDE report 7709657 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070712. One submission often covers many models, so check the product and lot details in each record.
- Z-1853-2022 · Jan 20, 2022 · Open, ClassifiedProduct/Part: YelloPort Plus 5x70 Cannula + Luer/YC0507001. For use in laparoscopic procedures.
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page