Maxi Move KMCSUNRecalls & reported problems
Arjohuntleigh AB · Maxi Move & SCALE - Mobile Floor Lifter - Safe working load of 500 lb / 227 kg. · Class 1
All models of Maxi Move (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05055982766937), received between Oct 28, 2024 and Sep 30, 2026 — classified as 1 death, 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem3
- Device Tipped Over2
- Detachment of Device or Device Component1
Patient problems
- Bone Fracture(s)2
- Fall2
- No Clinical Signs, Symptoms or Conditions2
- Head Injury1
- Hematoma1
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
DeathSep 30, 2026Detachment of Device or Device Component
Based on information received from the customer (the Executive Director), during a transfer using a Maxi Move lift and clip sling, the right leg clip became detached, resulting in the resident falling from the sling and striking their head. A hematoma was identified during the initial assessment, and the resident was transferred to the hospital for further evaluation. The resident passed away the following day. The reported cause of death was intracranial bleeding.\The customer report indicated a possible right leg spasm at the time of the incident. Following the event, the lift and sling were inspected by an Arjo technician. No malfunctions were identified, and all functions were found to be operating as intended.
Patient problems: Fall; Head Injury; Hematoma
MAUDE report 50943292 ↗MalfunctionFeb 20, 2026Use of Device Problem
A CAREGIVER TRIPPED OVER A LOOSE PLASTIC COVER LOCATED ON THE LEG OF A MAXI MOVE LIFT WHILE WALKING ACROSS THE LIFT BASE. AS A RESULT, THE CAREGIVER FELL AND SUSTAINED A KNEE INJURY. THE CAREGIVER DID NOT ATTEND THE EMERGENCY DEPARTMENT BUT SUBSEQUENTLY SOUGHT MEDICAL EVALUATION, WHICH INCLUDED A RADIOLOGY EXAMINATION; HOWEVER, THE RESULTS OF THIS ASSESSMENT ARE NOT KNOWN TO ARJO. WE WERE INFORMED THAT THE CAREGIVER HAS BEEN OFF WORK AND RETURNED TO WORK WITH SLIGHTLY MODIFIED DUTIES THREE DAYS AFTER THE EVENT. NO ADDITIONAL CLINICAL INFORMATION REGARDING THE INJURY OR TREATMENT HAS BEEN PROVIDED TO DATE. NO PATIENT WAS PRESENT OR INVOLVED AT THE TIME THE REPORTED ISSUE OCCURRED.
Patient problems: Fall; Insufficient Information
MAUDE report 24399060 ↗InjuryAug 4, 2025Use of Device Problem; Use of Device Problem
THE RESIDENT WAS BEING TRANSFERRED USING A MAXI MOVE WHEN THEY FELL THROUGH THE SLING. THE RESIDENT SUSTAINED A PELVIC FRACTURE.
Patient problems: Bone Fracture(s)
MAUDE report 22697192 ↗InjuryAug 4, 2025Use of Device Problem; Use of Device Problem
THE RESIDENT WAS BEING TRANSFERRED USING A MAXI MOVE WHEN THEY FELL THROUGH THE SLING. THE RESIDENT SUSTAINED A PELVIC FRACTURE. NO MALFUNCTION WAS IDENTIFIED WITH EITHER THE MAXI MOVE LIFT OR THE LOOP SLING THAT COULD HAVE CAUSED THE PATIENT¿S FALL.
Patient problems: Bone Fracture(s)
MAUDE report 22697191 ↗MalfunctionOct 28, 2024Device Tipped Over; Device Tipped Over
FOLLOWING THE REPORTED INFORMATION, THE MAXI MOVE FLOOR LIFT TIPPED SIDEWAYS DURING THE TRANSFER OF THE RESIDENT TO THE WHEELCHAIR. THIS OCCURRED WHEN ONE OF THE TWO PERSONAL SUPPORT WORKERS (PSWS) PULLED THE SLING WITH THE RESIDENT TO ENSURE HIS BACK WAS AGAINST THE WHEELCHAIR¿S BACKREST, WHILE THE SECOND PSW WAS GUIDING THE LIFT. ONE OF THE LIFT¿S LEGS WAS POSITIONED UNDER THE WHEELCHAIR AND ANOTHER IN FRONT. THE FACILITY STAFF WAS NOT SURE WHETHER THE BRAKES WERE LOCKED AT THAT MOMENT OR NOT. THE RESIDENT WAS PLACED SAFELY IN THE WHEELCHAIR. TWO OTHER PSWS ASSISTED TO PREVENT THE LIFT FROM TIPPING OVER AND HELPED TO UNHOOK THE SLING. NO INJURY WAS CLAIMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20544231 ↗MalfunctionOct 28, 2024Device Tipped Over; Device Tipped Over
FOLLOWING THE REPORTED INFORMATION, THE MAXI MOVE FLOOR LIFT TIPPED SIDEWAYS DURING THE TRANSFER OF THE PATIENT FROM BED TO THE WHEELCHAIR. THIS OCCURRED WHEN ONE OF THE TWO PERSONAL SUPPORT WORKERS (PSWS) GUIDING RESIDENT LEGS AND PUSHING ON SLING WITH ONE HAND AND THE SECOND GUIDING THE LIFT. THE BRAKES WERE RELEASED AND THE LIFT'S LEGS WERE IN CLOSED POSITION DURING THE EVENT. NO INJURY WAS CLAIMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20544229 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Maxi Move KMCSUN
Has the Maxi Move KMCSUN been recalled?
No FDA recall record in our data names this exact device identifier (DI 05055982766937), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Maxi Move KMCSUN?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Oct 28, 2024 and Sep 30, 2026. In those reports, the most common specific device problems are “Use of Device Problem” (3), “Device Tipped Over” (2) and “Detachment of Device or Device Component” (1); the most common patient problems are “Bone Fracture(s)” (2), “Fall” (2) and “Head Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Maxi Move KMCSUN?
1 report is classified as a death, 2 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Maxi Move KMCSUN FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arjohuntleigh AB.
Who makes the Maxi Move KMCSUN, and is it still sold?
The FDA's device database lists Arjohuntleigh AB as the company responsible for it (brand Maxi Move, model KMCSUN). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, Patient, Non-Ac-Powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Likorall™ overhead lift 3122010Liko AB · Lift, patient, non-ac-powered1 recall13 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page