CONCERTO BAB1000-01Recalls & reported problems
Arjohuntleigh AB · CONCERTO SHOWER TROLLEY, 1900MM HYDRAULIC · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05055982788083), received between Aug 28, 2019 and Mar 18, 2025 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Fell” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Fell3
- Fracture1
- Insufficient Device Problem Information1
- Unintended System Motion1
- Use of Device Problem1
Patient problems
- Fall3
- Limb Fracture2
- Abrasion1
- Head Injury1
- No Patient Involvement1
- Swelling/ Edema1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2025Device Fell; Device Fell
ARJO WAS NOTIFIED OF AN EVENT RELATED TO A CONCERTO SHOWER TROLLEY. IT WAS REPORTED THAT A PATIENT FELL FROM THE DEVICE AS THE SIDE SUPPORT OPENED. THE PATIENT SUSTAINED HEAD INJURY BUT NO MORE DETAILS REGARDING THE PATIENT'S TREATMENT WERE RELEASED BY THE FACILITY.
Patient problems: Fall; Head Injury
MAUDE report 21629246 ↗MalfunctionNov 16, 2022Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
ARJO WAS NOTIFIED ABOUT AN EVENT WITH INVOLVEMENT OF THE CONCERTO/BASIC SHOWER TROLLEY. IT WAS REPORTED THAT A PATIENT FELL FROM THE DEVICE WHILE SHOWERING. AS A CONSEQUENCE THE PATIENT SUSTAINED A BUMP. IT WAS INDICATED THAT ONE OF THE CATCHES DOES NOT HOLD THE SIDE SUPPORT IN PLACE. THIS CAUSES THE SIDE SUPPORT TO BEND AT THE BOTTOM. AS A RESULT, THE SIDE PANEL IS ONLY WELL FIXED ON ONE SIDE (IN THE UPPER PART).
Patient problems: Fall
MAUDE report 15803765 ↗InjurySep 8, 2021Unintended System Motion; Unintended System Motion
ARJO WAS NOTIFIED ABOUT AN EVENT WITH INVOLVEMENT OF THE CONCERTO/BASIC SHOWER TROLLEY. IT WAS REPORTED THAT WHEN THE CAREGIVER TURNED HER BACK TO THE PATIENT TO THROW AWAY THE DIAPER THE STRETCHER TILTED AND THE PATIENT FELL TO THE GROUND. THE PATIENT'S FALL WAS DAMPENED BY THE SIDE RAIL BEFORE FALL TO THE GROUND. THE PATIENT SUSTAINED SCRATCHES ON HIS KNEES AND WRISTS. HE WAS CHECKED BY A DOCTOR AND THERE WAS NO FRACTURE NOR SPRAIN FOUND. THE STRETCHER WAS IN VERTICAL POSITION BEFORE USE. THE CAREGIVER PUT THE STRETCHER TO THE HORIZONTAL POSITION BEFORE THE PATIENT WAS PLACED ON IT. THE PATIENT WAS LATERALIZED ON HIS LEFT SIDE TO CHANGE HIM.
Patient problems: Abrasion; Fall
MAUDE report 12437605 ↗InjuryJul 26, 2021Device Fell; Device Fell
ARJO WAS INFORMED ABOUT AN INCIDENT WITH INVOLVEMENT OF A CONCERTO SHOWER TROLLEY. IT WAS REPORTED THAT THE RESIDENT FELL OUT ON THE FLOOR, WHEN THE SIDE SUPPORT UNLOCKED AND MOVED TO LOWER POSITION. THE PATIENT WAS IN PAIN, HAD A SWOLLEN KNEE AND WAS TAKEN TO THE EMERGENCY ROOM FOR AN X-RAY. IT WAS CONFIRMED THAT PATIENT SUSTAINED THE FEMORAL FRACTURE.
Patient problems: Limb Fracture
MAUDE report 12224664 ↗InjuryJul 26, 2021Device Fell; Device Fell
ARJO WAS INFORMED ABOUT AN INCIDENT WITH INVOLVEMENT OF A CONCERTO SHOWER TROLLEY. IT WAS REPORTED THAT THE RESIDENT FELL OUT ON THE FLOOR, WHEN THE SIDE SUPPORT UNLOCKED AND MOVED TO LOWER POSITION. THE PATIENT WAS IN PAIN, HAD A SWOLLEN KNEE AND WAS TAKEN TO THE EMERGENCY ROOM FOR AN X-RAY. IT WAS CONFIRMED THAT PATIENT SUSTAINED THE FEMORAL FRACTURE.
Patient problems: Limb Fracture; Swelling/ Edema
MAUDE report 12224653 ↗MalfunctionAug 28, 2019Fracture; Fracture
ARJO WAS REQUESTED BY THE CUSTOMER FACILITY TO PERFORM SERVICE DUE TO FINDING 2 BROKEN SIDE SUPPORT SAFETY CATCHES. NO PATIENT INVOLVEMENT OR INJURY OCCURRENCE WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 8939886 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CONCERTO BAB1000-01
Has the CONCERTO BAB1000-01 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05055982788083), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CONCERTO BAB1000-01?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Aug 28, 2019 and Mar 18, 2025. In those reports, the most common specific device problems are “Device Fell” (3), “Fracture” (1) and “Unintended System Motion” (1); the most common patient problems are “Fall” (3), “Limb Fracture” (2) and “Abrasion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CONCERTO BAB1000-01?
None of the reports are classified as deaths, 3 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CONCERTO BAB1000-01 FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arjohuntleigh AB.
Who makes the CONCERTO BAB1000-01, and is it still sold?
The FDA's device database lists Arjohuntleigh AB as the company responsible for it (brand CONCERTO, model BAB1000-01). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, Patient, Non-Ac-Powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Likorall™ overhead lift 3122010Liko AB · Lift, patient, non-ac-powered1 recall13 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page