Maxi Twin KTBB4BSU2WWRecalls & reported problems
Arjohuntleigh AB · Maxi Twin Dedicated Powered DPS - with Scales - Mobile Floor Lifter - Safe working load of 400 lb / 182 kg · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05055982793360), received between Nov 2, 2020 and Nov 29, 2024 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Device Tipped Over1
Patient problems
- Head Injury3
- Abrasion2
- Fall2
- Skin Tears2
- Bruise/Contusion1
- Muscle/Tendon Damage1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT DURING RESIDENT TRANSFER THE LEFT LEG SLING CLIP DETACHED. AS A CONSEQUENCE OF THE EVENT THE RESIDENT FELL OUT FROM THE DEVICE. THE RESIDENT SUSTAINED FRACTURE TO THE SKULL AND BRAIN BLEED,
Patient problems: Head Injury
MAUDE report 20809491 ↗MalfunctionJul 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE RESIDENT WAS TRANSFERRED FROM A WHEELCHAIR TO A BED. WHEN THE RESIDENT WAS LIFTED FROM THE WHEELCHAIR THE SLING LEG CLIP DETACHED. THE RESIDENT SUSTAINED A BRUISING.
Patient problems: Bruise/Contusion
MAUDE report 19791940 ↗MalfunctionDec 21, 2023Device Tipped Over; Device Tipped Over
FOLLOWING THE INFORMATION REPORTED, THE MAXI MOVE PASSIVE FLOOR LIFT TIPPED OVER. A CERTIFIED NURSING ASSISTANT (CNA) SUSTAINED THE LEFT SHOULDER STRAIN WHILE STOPPING THE LIFT FROM HURTING THE RESIDENT. X-RAYS WERE TAKEN AND NO FRACTURE WAS NOTICED. NO MAGNETIC RESONANCE IMAGING (MRI) WAS NEEDED. CNA WAS RELEASED FROM THE HOSPITAL. ALTHOUGH IT WAS CLAIMED BY THE CUSTOMER THAT THE FLOOR LIFT LEGS WOULD NOT OPEN, SUCH ISSUE WAS NOT FOUND DURING THE DEVICE INSPECTION.
Patient problems: Muscle/Tendon Damage
MAUDE report 18370603 ↗InjuryNov 2, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
ARJO WAS MADE AWARE ABOUT AN EVENT INVOLVING ARJO MAXI TWIN PASSIVE FLOOR LIFT AND CLIP SLING. IT WAS STATED THAT WHEN REMOVING THE PATIENT FROM THE BATH, JUST ABOVE THE CHAIR, THE RESIDENT MADE A SUDDEN MOVE AND A LEFT LEG CLIP DETACHED FROM THE SPREADER BAR ATTACHMENT POINT. AS THE CONSEQUENCE OF THE EVENT THE PATIENT SLIPPED OUT OF THE SLING. THE PATIENT SUSTAINED A DEEP CUT ON THE LEG REQUIRING STITCHES (CALF AREA), ABRASIONS AND BUMP ON THE HEAD (CAT SCAN WAS NEEDED).
Patient problems: Abrasion; Fall; Head Injury; Skin Tears
MAUDE report 10768757 ↗InjuryNov 2, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
ARJO WAS MADE AWARE ABOUT AN EVENT INVOLVING ARJO MAXI TWIN PASSIVE FLOOR LIFT AND CLIP SLING. IT WAS STATED THAT WHEN REMOVING THE PATIENT FROM THE BATH, JUST ABOVE THE CHAIR, THE RESIDENT MADE A SUDDEN MOVE AND A LEFT LEG CLIP DETACHED FROM THE SPREADER BAR ATTACHMENT POINT. AS THE CONSEQUENCE OF THE EVENT THE PATIENT SLIPPED OUT OF THE SLING. THE PATIENT SUSTAINED A DEEP CUT ON THE LEG REQUIRING STITCHES (CALF AREA), ABRASIONS AND BUMP ON THE HEAD (CAT SCAN WAS NEEDED).
Patient problems: Abrasion; Fall; Head Injury; Skin Tears
MAUDE report 10768755 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Maxi Twin KTBB4BSU2WW
Has the Maxi Twin KTBB4BSU2WW been recalled?
No FDA recall record in our data names this exact device identifier (DI 05055982793360), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Maxi Twin KTBB4BSU2WW?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Nov 2, 2020 and Nov 29, 2024. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (4) and “Device Tipped Over” (1); the most common patient problems are “Head Injury” (3), “Abrasion” (2) and “Fall” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Maxi Twin KTBB4BSU2WW?
None of the reports are classified as deaths, 3 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Maxi Twin KTBB4BSU2WW FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arjohuntleigh AB.
Who makes the Maxi Twin KTBB4BSU2WW, and is it still sold?
The FDA's device database lists Arjohuntleigh AB as the company responsible for it (brand Maxi Twin, model KTBB4BSU2WW). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, Patient, Non-Ac-Powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Likorall™ overhead lift 3122010Liko AB · Lift, patient, non-ac-powered1 recall13 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page