SARA FLEX HEB0000Recalls & reported problems
Arjohuntleigh AB · SARA FLEX - Standing and raising aid and transfer aid - Powered · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05056097354941), received between Mar 4, 2022 and Jun 3, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem2
- Insufficient Device Problem Information1
Patient problems
- Multiple Fractures2
- Bone Fracture(s)1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2026Insufficient Device Problem Information
A RESIDENT LOST BALANCE DURING A TRANSFER FROM BED TO AN ELECTRIC WHEELCHAIR USING ARJO SARA FLEX LIFT AND ARJO SLING AFTER HER LEFT KNEE SHIFTED WITHIN THE CALF SUPPORT DURING THE STANDING PHASE. THE CAREGIVER WAS ABLE TO CATCH THE RESIDENT AND GUIDE HER SAFELY TO THE GROUND; HOWEVER, THE LEGS REMAINED SECURED BEHIND THE CALVES IN THE SUPPORT, RESULTING IN PRESSURE BEING APPLIED TO THE RESIDENT¿S KNEE. THE RESIDENT SUSTAINED A KNEE FRACTURE THAT REQUIRED SURGERY. THE DEVICE INSPECTION DID NOT REVEAL ANY PRODUCT MALFUNCTION. THE DEVICE OPERATED CORRECTLY.
Patient problems: Bone Fracture(s)
MAUDE report 25370807 ↗InjuryMar 4, 2022Use of Device Problem; Use of Device Problem
ARJO RECEIVED A NOTIFICATION REGARDING THE EVENT WITH A SARA FLEX ACTIVE LIFT. IT WAS REPORTED THAT A RESIDENT OF THE NURSING HOME, WHO IS SUFFERING FROM SEVERE OSTEOPOROSIS, WAS BEING TRANSPORTED IN A STANDING AID FROM THE BED TO THE BATHROOM. FOLLOWING THE INFORMATION RECEIVED THE RESIDENT WAS HELPED TO STAND UP WITH THE STANDING AID AND IN THIS MOMENT BROKE HER ANKLE. THE RESIDENT WAS NOT ABLE TO STAY UP AND SINKS IN THE DEVICE. SHE COULD NOT HOLD THE FEET ON THE PLATFORM, BUT THE SAFETY BELT USED TO KEEP THE FEET CLOSE TO THE FOOTREST HAS BEEN REMOVED. THIS CAUSED THAT THE PERSON ROTATED IN THE DEVICE. THIS CONTRIBUTED TO A FRACTURE OF BOTH FEMURS. IT WAS REPORTED THAT THE RESIDENT IS STILL IN HOSPITAL.
Patient problems: Multiple Fractures
MAUDE report 13666419 ↗InjuryMar 4, 2022Use of Device Problem; Use of Device Problem
ARJO RECEIVED A NOTIFICATION REGARDING THE EVENT WITH A SARA FLEX ACTIVE LIFT. IT WAS REPORTED THAT A RESIDENT OF THE NURSING HOME, WHO IS SUFFERING FROM SEVERE OSTEOPOROSIS, WAS BEING TRANSPORTED IN A STANDING AID FROM THE BED TO THE BATHROOM. FOLLOWING THE INFORMATION RECEIVED THE RESIDENT WAS HELPED TO STAND UP WITH THE STANDING AID AND IN THIS MOMENT BROKE HER ANKLE. THE RESIDENT WAS NOT ABLE TO STAY UP AND SINKS IN THE DEVICE. SHE COULD NOT HOLD THE FEET ON THE PLATFORM, BUT THE SAFETY BELT USED TO KEEP THE FEET CLOSE TO THE FOOTREST HAS BEEN REMOVED. THIS CAUSED THAT THE PERSON ROTATED IN THE DEVICE. THIS CONTRIBUTED TO A FRACTURE OF BOTH FEMURS. IT WAS REPORTED THAT THE RESIDENT IS STILL IN HOSPITAL.
Patient problems: Multiple Fractures
MAUDE report 13666279 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SARA FLEX HEB0000
Has the SARA FLEX HEB0000 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05056097354941), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SARA FLEX HEB0000?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 4, 2022 and Jun 3, 2026. In those reports, the most common specific device problem is “Use of Device Problem” (2); the most common patient problems are “Multiple Fractures” (2) and “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SARA FLEX HEB0000?
None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SARA FLEX HEB0000 FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arjohuntleigh AB.
Who makes the SARA FLEX HEB0000, and is it still sold?
The FDA's device database lists Arjohuntleigh AB as the company responsible for it (brand SARA FLEX, model HEB0000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, Patient, Non-Ac-Powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Likorall™ overhead lift 3122010Liko AB · Lift, patient, non-ac-powered1 recall13 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page