Quantum ® QDM1Recalls & reported problems

Spectrum Medical Limited · Quantum Diagnostic Module · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05060434420701), received between Jul 5, 2024 and Feb 4, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Sensing Problem1
  • Energy Output Problem1
  • High Readings1
  • Incomplete or Inadequate Connection1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Low Oxygen Saturation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2026Energy Output Problem; Output Problem

    THIS ISSUE RELATES TO INACCURATE SWEEP FLOW, WHICH IS ONLY DISPLAYING ROUGHLY HALF OF THE VALUE SET BY THE GAS BLENDER. THERE WAS NO PATIENT HARM AND NO DELAY TO OPERATION. THIS DEVICE IS BEING RETURNED FOR CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256539 ↗
  • MalfunctionAug 26, 2025High Readings; High Readings

    USER REPORTED THAT WHEN CHECKING THE GAS FLOW ACCURACY, THE READING VALUE WAS NOT WITHIN ALLOWABLE TOLERANCES. READING VALUE FOR FLOW RATE IS 7.1LPM RATHER THAT THE EXPECTED 5LPM. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22894581 ↗
  • MalfunctionJan 31, 2025Device Sensing Problem; Device Sensing Problem

    USER REPORTED ISSUES RELATED TO DEVICE SENSING PROBLEMS USING SENSORS. INABILITY TO CORRECTLY SENSE IS CONSIDERED A REPORTABLE EVENT. THERE WAS NO PATIENT HARM AS A RESULT OF THESE MALFUNCTIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21282751 ↗
  • MalfunctionJul 5, 2024Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    USER REPORTED THAT THEY NOTICED DARK ARTERIAL BLOOD AND LOW SVO2 ON INITIAL OF BYPASS WHILST MANUAL GAS BLENDER 02 SWEEP WAS SET AT 2L AND FIO2 SET AT 100%. THE PATIENT WAS WEANED OFF BYPASS TO ALLOW FOR REOXYGENATION VIA THE VENTILATOR AND TO HAVE TIME TO TROUBLE SHOOT THE GAS ISSUE. ON INSPECTION, THE USER REPORTED THAT THEY NOTICED THAT THE GAS SUPPLY FROM MANUAL GAS BLENDER TO QDM WAS NOT COMPLETE ON THE QDM. THE GAS CIRCUIT WAS RECONNECTED, AND BYPASS WAS RE-INITIATED WITHOUT FURTHER ISSUES.

    Patient problems: Low Oxygen Saturation

    MAUDE report 19679460 ↗

FDA clearance and classification

Cleared through 510(k) K173591 on Mar 28, 2018 (decision: Substantially Equivalent). Submitted by Spectrum Medical , Ltd. as “Quantum Diagnostic Module, Quantum Diagnostic Module - No Gas”. The FDA review took 127 days. FDA record ↗

The FDA regulates product code DRY (monitor, blood-gas, on-line, cardiopulmonary bypass) under 21 CFR 870.4330 as a Class II device. 21 CFR 870.4330 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Quantum ® QDM1

Has the Quantum ® QDM1 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05060434420701), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Quantum ® QDM1?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 5, 2024 and Feb 4, 2026. In those reports, the most common specific device problems are “Device Sensing Problem” (1), “Energy Output Problem” (1) and “High Readings” (1); the most common patient problem is “Low Oxygen Saturation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Quantum ® QDM1?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Quantum ® QDM1 FDA approved?

It was cleared through 510(k) K173591 on Mar 28, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Quantum ® QDM1, and is it still sold?

The FDA's device database lists Spectrum Medical Limited as the company responsible for it (brand Quantum ®, model QDM1). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.