The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05060434420732), received between Dec 12, 2024 and Feb 3, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Loose or Intermittent Connection1
- Pumping Stopped1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2026Unintended Movement
THIS ISSUE RELATES TO AN OCCLUDER OPENING TWICE DURING SURGERY WITHOUT REQUEST. THE DEVICE IS BEING RETURNED AND AN INVESTIGATION IS ONGOING INTO THIS PROBLEM. ALTHOUGH, THERE WAS NO PATIENT HARM OR DELAY TO SURGERY.THIS EVENT IS DEEMED REPORTABLE BY SPECTRUM MEDICAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24244558 ↗MalfunctionDec 12, 2024Loose or Intermittent Connection; Pumping Stopped; Loose or Intermittent Connection; Pumping Stopped
USER NOTICED ARTERIAL FLOW WAS UNAVAILABLE. THIS PROMPTED THE USER TO MOVE BUBBLE AND FLOW SENSORS. USER ATTEMPTED TO DISABLE BUBBLE ALARM PRIOR TO MOVING BUT DUE TO A SUSPECTED CONNECTION ISSUE BUBBLE ALARM REMAINED PRIMED. ON MOVING THE BUBBLE SENSOR, BUBBLE ALARM TRIGGERED CAUSING OCCLUDER TO CLOSE AND PUMP TO STOP. USER FOLLOWED RISK COTROLS AND RESUMED BYPASS 40 SECONDS AFTER CESSATION. THERE WAS NO HARM TO THE PATIENT. SPECTRUM MEDICAL CONSIDERS PUMP STOP A REPORTABLE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20914129 ↗
Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.