Quantum QSOL (REF 51-000022-00)Recalls & reported problems

Spectrum Medical Limited · Quantum Smart Occluder - Left hand · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05060434420732), received between Dec 12, 2024 and Feb 3, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loose or Intermittent Connection1
  • Pumping Stopped1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Unintended Movement

    THIS ISSUE RELATES TO AN OCCLUDER OPENING TWICE DURING SURGERY WITHOUT REQUEST. THE DEVICE IS BEING RETURNED AND AN INVESTIGATION IS ONGOING INTO THIS PROBLEM. ALTHOUGH, THERE WAS NO PATIENT HARM OR DELAY TO SURGERY.THIS EVENT IS DEEMED REPORTABLE BY SPECTRUM MEDICAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24244558 ↗
  • MalfunctionDec 12, 2024Loose or Intermittent Connection; Pumping Stopped; Loose or Intermittent Connection; Pumping Stopped

    USER NOTICED ARTERIAL FLOW WAS UNAVAILABLE. THIS PROMPTED THE USER TO MOVE BUBBLE AND FLOW SENSORS. USER ATTEMPTED TO DISABLE BUBBLE ALARM PRIOR TO MOVING BUT DUE TO A SUSPECTED CONNECTION ISSUE BUBBLE ALARM REMAINED PRIMED. ON MOVING THE BUBBLE SENSOR, BUBBLE ALARM TRIGGERED CAUSING OCCLUDER TO CLOSE AND PUMP TO STOP. USER FOLLOWED RISK COTROLS AND RESUMED BYPASS 40 SECONDS AFTER CESSATION. THERE WAS NO HARM TO THE PATIENT. SPECTRUM MEDICAL CONSIDERS PUMP STOP A REPORTABLE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20914129 ↗

Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices

All Console, Heart-Lung Machine, Cardiopulmonary Bypass devices →