Quantum QFS2-ST-200Recalls & reported problems
Spectrum Medical Limited · Quantum Flow Sensor - ID 1/4" x Wall 1/16" (OD 3/8") 2.0m · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05060434421852), received between Jan 26, 2023 and Feb 6, 2026 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem1
- Inappropriate or Unexpected Reset1
- Pumping Stopped1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2026Pumping Stopped
USER REPORTED THAT A BUBBLE WAS DETECTED WITHOUT THE PRESENCE OF AIR. THERE WAS NO PATIENT HARM, BUT THIS TYPE OF EVENT IS CONSIDERED REPORTABLE SINCE IT CAUSES A PUMP STOP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24278796 ↗MalfunctionJan 31, 2025Device Sensing Problem; Device Sensing Problem
USER REPORTED ISSUES RELATED TO DEVICE SENSING PROBLEMS USING SENSORS. INABILITY TO CORRECTLY SENSE IS CONSIDERED A REPORTABLE EVENT. THERE WAS NO PATIENT HARM AS A RESULT OF THESE MALFUNCTIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21282622 ↗DeathJan 26, 2023Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
FLOW PROBE SENSOR, ECMO COORDINATOR, REPORTED THAT THEY HAD A 1/4" FLOW PROBE THAT WAS READING ABOUT HALF OF THE ACTUAL FLOW. THE FLOW PROBE FS2000138 WAS THE ARTERIAL FLOW PROBE, AND THE ENTIRE CASE IT WAS READING AROUND .200 LPM, WHICH WAS THE TARGET FLOW THEY ALSO HAD A VENOUS FLOW PROBE (MEASURED PRE-SHUNT SO THISSHOULD BE VERY CLOSE TO THE ART FLOW) AND AN "ALT FLOW" PROBE (SHUNT FLOW). AN ECMO SPECIALIST EVENTUALLY NOTICED THE DISCREPANCY BETWEEN ART AND VEN FLOWS (NO ALARMS SET OR CHARTING OF VEN AND ALT ROUTINELY). THEY DID SOME TROUBLE SHOOTING BY ATTACHING ALL FLOW PROBES TO THE ARTERIAL LINE, AND THE ART FLOW WAS READING ABOUT HALF OF WHAT THE OTHER TWO FLOW PROBES WERE READING. THEY THEN REPLACED FS2000138 WITH A NEW FLOW PROBE, AND THIS WAS READING THE SAME AS THE VENOUS FLOW PROBE. THE SPECIALIST THAT SET UP THE PUMP REPORTED ZEROING ALL FLOWS BEFORE THE CASE. THEY DID NOT ZERO THE REPLACEMENT FLOW PROBE BECAUSE THEY WERE ON SUPPORT.
Patient problems: Insufficient Information
MAUDE report 16245369 ↗
Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page