Quantum QSOR (REF 51-000074-00)Recalls & reported problems
Spectrum Medical Limited · Quantum Smart Occluder - Right Hand · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05060434421944), received between Dec 9, 2024 and Feb 12, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure1
- Obstruction of Flow1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2026Unintended Movement
THE CUSTOMER REPORTED A SUDDEN LOCKING OF THE VENOUS OCCLUDER; AND THAT OPENING THE OCCLUDER WAS NOT POSSIBLE. THE CUSTOMER HAD TO OPEN THE OCCLUDE IN A MANUAL WAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24320428 ↗MalfunctionJan 29, 2025Activation Failure; Activation Failure
LOANER OCCLUDER THAT WAS SENT IN RESPONSE TO FON #(B)(6) STOPPED RESPONDING TO OPEN/CLOSE COMMANDS FROM THE QWS. REPORTEES ARE UNSURE IF THERE WAS A "NO OCCLUDER LOCATION" MESSAGE AT THE TIME OF THIS EVENT. THE OCCLUDER WAS ALSO REPORTED TO "EVENTUALLY START WORKING" AGAIN. ISSUE OCCURED APPROX. 1000 PRIOR TO A CASE. RECOMMENDATION WAS GIVEN TO TRY A NEW CABLE AND THEN A NEW PORT ON THE FRAME IF THE CABLE WAS DAMAGED FROM THE FRAME. NO FURTHER INFORMATION. SAME REP CONTACTED AGAIN AROUND 1700 STATING AS BIOMED ATTEMPTED TO DOWNLOAD LOGS FOR THE ISSUE ABOVE, ONLY THE QWS WOULD BOOT UP AND NOT THE REST OF THE SUBMODULES. REST OF THE PUMP WAS TURNED ON VIA THE SOLARIS POWER BUTTON BUT PUMP MANAGER WAS NOT PRESENT. SAP CABLE WAS UNPLUGGED FROM THE QWS AND REPLUGGED IN BUT PUMP MANAGER WAS STILL NOT PRESENT. QWS WAS THEN TURNED OFF WITH A LONG PRESS, THEN SAP CABLE UNPLUGGED AGAIN AND REPLUGGED IN. UPON BOOT UP PUMP MANAGER CAME BACK. MEDTRONIC REP WAS ON FACETIME WITH BIOMED AND VERIFIES THAT SOLARIS WAS PLUGGED INTO THE WALL DURING THIS TIME. HE ALSO SENT A PICTURE OF A QWS WITH A GREEN POWER LED IN THE BOTTOM RIGHT INSTEAD OF BLINKING BLUE, SUGGESTING THAT THE HLM WAS PLUGGED INTO TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21264089 ↗MalfunctionDec 9, 2024Obstruction of Flow; Obstruction of Flow
VENOUS OCCLUDER ENGAGED DURING A CASE WITH A PATIENT WHEN NOT ACTIVATED. PERFUSIONIST MANUALLY RELEASED THE CLAMP. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878470 ↗
Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.