Quantum QFS4-ST-075Recalls & reported problems
Spectrum Medical Limited · Quantum Flow Sensor - ID 3/8" x Wall 3/32" (OD 9/16") 0.75m · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05060434422002), received between Jan 31, 2025 and Mar 18, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem3
- Electrical /Electronic Property Problem1
- False Alarm1
- False Positive Result1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2026False Alarm
THIS EVENT RELATES TO THE ARTERIAL FLOW CHANNEL BECOMING UNAVAILABLE DURING WEANING. THE PERFUSIONIST MENTIONED THAT HE HAD AN ALARM OF ARTERIAL FLOW CHANNEL BEING UNAVAILABLE AND THAT THE MACHINE WAS ALARMING AND HE COULD NOT CLEAR NOR OVERRIDE THE ERROR. THE PERFUSIONIST HAD TO GO OUT OF WEANING MODE AND FINISH THE CASE MANUALLY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24625672 ↗MalfunctionSep 10, 2025Device Sensing Problem; Device Sensing Problem
USER REPORTED ERRONEOUS BUBBLE DETECTION. THERE WAS NO PATIENT HARM; HOWEVER, SPECTRUM MEDICAL CONSIDERS THIS EVENT TO BE REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23018971 ↗MalfunctionJul 17, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
BUBBLE ALARM WAS TRIGGERED AFTER THE CROSS-CLAMP WAS REMOVED DURING A CLINICAL CASE. ACCORDING TO THE PERFUSIONIST, THERE WAS NO VISIBLE AIR IN THE CIRCUIT, AND ALL OTHER SAFETY SYSTEMS FUNCTIONED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22530402 ↗MalfunctionJul 10, 2025Device Sensing Problem; Device Sensing Problem
USER REPORTED A DIFFERENCE IN FLOW BETWEEN THE SAFE FLOW PROBE AND ARTERIAL FLOW PROBE NOTED TO BE 0.5 L/MIN AT 4.0L/MIN OF FLOW. THIS FLOW PROBE WAS ABLE TO BE IDENTIFIED AND SWAPPED WITH A BACK UP DURING THE CASE. THERE WAS NO HARM TO THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22466413 ↗MalfunctionJun 17, 2025False Positive Result; False Positive Result
IT WAS REPORTED THAT THE THE BUBBLE DETECTOR TRIGGERED AFTER THE USER HAD FINISHED PRIMING. AS NO AIR WAS VISIBLE, THIS WAS TREATED AS A FALSE POSITIVE. THERE WAS NO HARM TO THE PATIENT INVOLVED. WE CONSIDER INCORRECT MEASUREMENT FROM BUBBLE SENSOR TO BE REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22235924 ↗MalfunctionJan 31, 2025Device Sensing Problem; Device Sensing Problem
THE LEVEL SENSOR'S LIGHTS WERE ON AND QWS SET TO SAFE FLOW LEVEL. THE PUMP RPM KICKED INTO FLOW REGULATION DURING PRIMING SEVERAL TIMES. THE PERFUSIONISTS REPLACED THE TAPE AND RESOLVED THIS ISSUE. HOWEVER, THIS HAPPENED AGAIN WHILE THEY WERE ON BYPASS. THE ALARM IS SHOWING FLASHING "S" BUT RPM NEVER DECREASED. ARTERIAL AND VENOUS FLOW STAYED STABLE DURING THESE EVENTS. THEY CHECKED THE CONNECTIONS AND MADE SURE EVERYTHING WAS CONNECTED TIGHTLY. THE PROBE DOESN'T HAVE VISIBLE DAMAGE WHEN THEY INSPECTED. THE LEVEL SENSOR WAS BEING USED FOR THE REST OF THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21281084 ↗
FDA clearance and classification
Cleared through 510(k) K190282 on May 24, 2019 (decision: Substantially Equivalent). Submitted by Spectrum Medical , Ltd. as “Quantum Smart Occluder LH (left-hand), Quantum Smart Occluder LH (right-hand)”. The FDA review took 102 days. FDA record ↗
The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Quantum QFS4-ST-075
Has the Quantum QFS4-ST-075 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05060434422002), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Quantum QFS4-ST-075?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jan 31, 2025 and Mar 18, 2026. In those reports, the most common specific device problems are “Device Sensing Problem” (3), “Electrical /Electronic Property Problem” (1) and “False Alarm” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Quantum QFS4-ST-075?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Quantum QFS4-ST-075 FDA approved?
It was cleared through 510(k) K190282 on May 24, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Quantum QFS4-ST-075, and is it still sold?
The FDA's device database lists Spectrum Medical Limited as the company responsible for it (brand Quantum, model QFS4-ST-075). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls449 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths59 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls16 deaths99 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass298 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass228 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death111 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall77 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page