LifeStream LSMU1350938Recalls & reported problems

Clearstream Technologies Limited · LifeStream™ Balloon Expandable Vascular Covered Stent, 9 mm x 38 mm, 135 cm catheter length · Class 3 · Implant

All models of LifeStream (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05391522081260), received between Dec 6, 2017 and Dec 12, 2023 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Migration or Expulsion of Device2
  • Bent1
  • Failure to Advance1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • Calcium Deposits/Calcification1
  • Injury1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2023Migration or Expulsion of Device; Device Dislodged or Dislocated; Migration or Expulsion of Device; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE SUBCLAVIAN ARTERY VIA BRACHIAL ACCESS, THE STENT GRAFT ALLEGEDLY DISLODGED FROM THE BALLOON. IT WAS FURTHER REPORTED THAT THE STENT GRAFT ALLEGEDLY FLOATED INTO THE CAROTID ARTERY. REPORTEDLY, THE STENT WAS UNABLE TO BE RETRIEVED THUS DEPLOYED IN THE CAROTID ARTERY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18308458 ↗
  • MalfunctionAug 1, 2022Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A STENT GRAFT PLACEMENT PROCEDURE, THE PRODUCT BOX WAS ALLEGEDLY DAMAGED AND FOUND WET UPON OPENING. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15137431 ↗
  • MalfunctionSep 25, 2020Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING TREATMENT IN THE ILIAC LESION, THE STENT ALLEGEDLY FAILED TO DEPLOY FROM THE DELIVERY SYSTEM. IT WAS FURTHER REPORTED THAT THE STENT ALLEGEDLY CAME OFF WHEN THE DEVICE WAS PULLED OUT OF THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10585366 ↗
  • MalfunctionFeb 19, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE BALLOON EXPANDABLE VASCULAR COVERED STENT ALLEGEDLY DISLODGED FROM THE BALLOON AS IT WAS PASSED OVER THE GUIDEWIRE. IT WAS FURTHER REPORTED THAT THE COVERED STENT WAS RECANNULATED IN ANOTHER AREA OF THE VESSEL. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8352287 ↗
  • InjuryDec 6, 2017Bent; Migration or Expulsion of Device; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE STENT DISLODGED AND MIGRATED REQUIRING A SURGICAL CUT DOWN TO REMOVE IT. THERE WERE NO ANOMALIES NOTED DURING THE PREP OF THE DEVICE. THERE WAS NO DAMAGE NOTED TO THE BOX, TRAY OR STENT COVER AND THERE WAS NO DIFFICULTY REMOVING THE STENT COVER. THE PATIENT WAS TO BE TREATED FOR SUBCLAVIAN STENOSIS. THE COMPLAINT DEVICE WAS TO BE USED IN THE LEFT OSTIAL SUBCLAVIAN LESION. THE LESION WAS 40% CALCIFIED AND THERE WAS STENOSIS. THE LESION WAS TORTUOUS. FEMORAL ACCESS WAS MADE. A 7FR SHUTTLE, 90CM INTRODUCER SHEATH WAS USED. THE COMPLAINT DEVICE WAS PREPPED CORRECTLY AND DELIVERED OVER A 035 GUIDEWIRE THROUGH THE SHEATH TO THE TARGET LESION. THE TRACKING PATH WAS 40% CALCIFIED. THE STENT GRAFT WAS PROTECTED BY THE SHEATH DURING ALL STEPS OF TRACKING TO THE TARGET LESION. THERE WAS NO RESISTANCE FELT WHEN TRACKING THE DEVICE. THE 9X38 STENT WAS DECIDED TO BE TOO LONG TO TREAT THE LESION AND IT WAS THEREFORE DECIDED TO BE REMOVED. THE DEVICE WAS REMOVED OVER THE WIRE BACK THROUGH THE SHEATH. WHEN RETRACTING THE DEVICE THE STENT DISLODGED FROM THE BALLOON NEAR THE EDGE OF THE SHEATH. THERE WAS A BEND IN THE WIRE ALSO WHICH WAS THOUGHT MAY HAVE CONTRIBUTED TO THE DI

    Patient problems: Calcium Deposits/Calcification; Injury

    MAUDE report 7085972 ↗

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →