LifeStream LSMU1350958Recalls & reported problems
Clearstream Technologies Limited · LifeStream™ Balloon Expandable Vascular Covered Stent, 9 mm x 58 mm, 135 cm catheter length · Class 3 · Implant
All models of LifeStream (37)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05391522081277), received between Jul 2, 2019 and Nov 26, 2022 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated6
- Device-Device Incompatibility1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Hemorrhage/Blood Loss/Bleeding1
- Injury1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryNov 26, 2022Device-Device Incompatibility; Device Dislodged or Dislocated; Device-Device Incompatibility; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING THE STENT GRAFT PLACEMENT PROCEDURE, THE DEVICE WAS ALLEGEDLY DISLODGED AND HAD DEVICE INCOMPATIBILITY ISSUES. IT WAS FURTHER REPORTED THAT THE STENT WAS REMOVED AFTER SURGERY. FURTHERMORE BLEEDING WAS NOTED IN THE SITE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Hemorrhage/Blood Loss/Bleeding
MAUDE report 15866769 ↗InjuryNov 18, 2022Migration or Expulsion of Device; Device Dislodged or Dislocated; Migration or Expulsion of Device; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY DISLODGED IN THE PATIENT. IT WAS FURTHER REPORTED THAT THE STENT GOT STUCK IN THE ARTERY WALL AND THE PATIENT HAD TO BE TAKEN TO THE OPERATION ROOM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 15823932 ↗MalfunctionNov 17, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT PRIOR TO A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY DISLODGED FROM THE BALLOON. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15812425 ↗MalfunctionMar 24, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE OF A HIGHLY CALCIFIED TARGET LESION OF CHRONIC TOTAL OCCLUSION IN THE COMMON ILIAC ARTERY VIA RIGHT BRACHIAL APPROACH, THE STENT WAS ALLEGEDLY FOUND TO BE DISLOCATED ON THE BALLOON. IT WAS FURTHER REPORTED THAT IT WAS REMOVED AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13887343 ↗InjuryOct 8, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A STENT GRAFT TREATMENT OF A THORACIC TYPE B DISSECTION IN THE CELIAC ARTERY USING THE CHIMNEY TECHNIQUE VIA BRACHIAL ARTERY, THE STENT GRAFT ALLEGEDLY DISLODGED FROM THE BALLOON. REPORTEDLY, THE HEALTHCARE PROVIDER WAS ABLE TO CATCH THE STENT GRAFT BY BRINGING IT BACK INTO THE SHEATH, BUT ALLEGEDLY DISLODGED IN THE SHEATH. IT WAS FURTHER REPORTED THAT THE SHEATH WAS REMOVED FROM THE PATIENT AND THE STENT GRAFT REMAINED IN THE BRACHIAL ARTERY. THE HEALTHCARE PROVIDER ALLEGEDLY WAS AWARE THE PRODUCT WAS BEING USED OFF LABEL. THE PATIENT'S ARTERY WAS OPENED TO REMOVE THE STENT AND THREE VEIN GRAFTS WERE REQUIRED TO REPAIR THE ARTERY.
Patient problems: Injury
MAUDE report 9169088 ↗MalfunctionJul 2, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING TREATMENT OF THE ILIAC ARTERY VIA CALCIFIED COMMON FEMORAL ACCESS, THE STENT GRAFT ALLEGEDLY DISLODGED FROM THE DELIVERY SYSTEM. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED BY REPOSITIONING THE BALLOON AND IMPLANTING THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8755624 ↗
Other Iliac covered stent, arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page