Skip to content

LifeStream LSMU1350616Recalls & reported problems

Clearstream Technologies Limited · LifeStream™ Balloon Expandable Vascular Covered Stent, 6 mm x 16 mm, 135 cm catheter length · Class 3 · Implant

All models of LifeStream (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05391522081468), received between Aug 16, 2017 and Jul 9, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated6
  • Physical Resistance2
  • Sticking2
  • Device-Device Incompatibility1
  • Failure to Advance1
  • Retraction Problem1

Patient problems

  • No Consequences Or Impact To Patient3
  • Foreign Body In Patient1
  • Ischemia1
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • No Known Impact Or Consequence To Patient1
  • Perforation of Vessels1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 9, 2026Device Dislodged or Dislocated

    ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING A LIFESTREAM DEVICE FOR LEFT RENAL ARTERY STENOSIS. PRE DEPLOYMENT EVALUATION DEMONSTRATED THAT THE LESION WAS HIGHLY FOCAL, WITH A SIGNIFICANT PORTION OF THE STENT EXPECTED TO EXTEND INTO THE AORTA. AT THAT POINT, A DECISION WAS MADE TO REMOVE THE DEVICE AND USE ANOTHER STENT. THE STENT WAS WITHDRAWN, AND INSPECTION REVEALED THAT IT HAD DISLODGED FROM THE BALLOON WITHIN THE LEFT RENAL ARTERY. MULTIPLE ATTEMPTS WERE MADE TO SNARE THE STENT; HOWEVER, THESE WERE UNSUCCESSFUL. THE STENT ULTIMATELY BECAME LODGED IN A SEGMENTAL BRANCH OF THE RENAL ARTERY, WHICH HAD SPASMED AROUND IT, RENDERING IT UNRETRIEVABLE. THE CURRENT CONDITION OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 25769149 ↗
  • MalfunctionJul 1, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, A PATIENT UNDERWENT A STENT GRAFT PLACEMENT PROCEDURE USING THE LIFESTREAM. IT WAS REPORTED THAT DURING TO THE PROCEDURE, THE SENT ALLEGEDLY DISLODGED PRIOR TO INSERTING INTO SHEATH. THERE WAS NO PATIENT REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22368605 ↗
  • MalfunctionJul 13, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT, THE STENT ALLEGEDLY CAME OFF THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10263066 ↗
  • MalfunctionJul 13, 2020Device-Device Incompatibility; Device Dislodged or Dislocated; Device-Device Incompatibility; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING THE STENT PROCEDURE IN CELIAC CURVE THROUGH COMMON FEMORAL ACCESS, THE STENT GRAFT ALLEGEDLY DISLODGED WITHIN THE INTRODUCER SHEATH FROM THE BALLOON DELIVERY SYSTEM. ANOTHER DEVICE AND SAME SHEATH WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 10263040 ↗
  • MalfunctionSep 10, 2019Retraction Problem; Failure to Advance; Device Dislodged or Dislocated; Retraction Problem; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING OFF-LABEL TREATMENT OF THE CALCIFIC CELIAC ARTERY VIA RIGHT GROIN ACCESS, THE BALLOON EXPANDABLE VASCULAR COVERED STENT DELIVERY SYSTEM ALLEGEDLY FAILED TO CROSS THE LESION. THEREFORE, THE HEALTH CARE PROVIDER (HCP) DECIDED TO REMOVE THE DELIVERY SYSTEM THROUGH THE INTRODUCER SHEATH, CAUSING THE STENT GRAFT TO ALLEGEDLY PARTIALLY DISLODGE FROM BALLOON AT THE HEMOSTATIC VALVE. REPORTEDLY, THE STENT GRAFT AND THE DELIVERY SYSTEM WERE REMOVED AS ONE UNIT THROUGH THE INTRODUCER SHEATH WITH SOME DIFFICULTY. ANOTHER PTA BALLOON CATHETER AND ANOTHER DELIVERY SYSTEM WERE USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8982405 ↗
  • InjuryAug 21, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT AS THE DEVICE WAS PULLED BACK TO REPOSITION IT, THE STENT GRAFT ALLEGEDLY DISLODGED FROM THE CATHETER UNINTENTIONALLY IN THE ILIAC VIA A CONTRALATERAL APPROACH. IT WAS FURTHER REPORTED THAT THE VESSEL WAS PERFORATED AND A SECOND STENT WAS PLACED TO TAMPONADE THE VESSEL AND SUCCESSFULLY COMPLETE THE PROCEDURE. THE PATIENT WAS REPORTEDLY DOING WELL POST PROCEDURE.

    Patient problems: Perforation of Vessels

    MAUDE report 7803102 ↗
  • MalfunctionAug 22, 2017Sticking; Physical Resistance

    IT WAS REPORTED THAT THE DEVICE BECAME STUCK IN A 6FR COOK ANSEL SHEATH DURING ADVANCEMENT INTO THE LESION. THE DOCTOR PREPPED THE DEVICE. THERE WERE NO ANOMALIES NOTED DURING THE PREP OF THE DEVICE. THERE WAS NO DAMAGE NOTED TO THE BOX, TRAY OR STENT COVER AND NO DIFFICULTY REMOVING THE STENT COVER. THE PATIENT WAS BEING TREATED FOR SMA STENOSIS MESENTERIC ISCHEMIA. A 6F, 45CM ANSEL 2 INTRODUCER SHEATH AND A 0.035, 260CM ROSEN GUIDEWIRE WERE USED. THE DEVICE WAS ATTEMPTED TO BE INSERTED THROUGH THE 6F SHEATH HOWEVER RESISTANCE WAS FELT JUST PAST THE VALVE OF THE SHEATH. AN ATTEMPT WAS MADE TO RETRACT THE DEVICE FROM THE SHEATH HOWEVER THIS WAS UNSUCCESSFUL AS THE STENT WAS STUCK. THE STENT REMAINED STUCK IN THE INTRODUCER SHEATH. AIR EVACUATION WAS NOT PERFORMED PRIOR TO INSERTING THE DEVICE. MULTIPLE ATTEMPTS TO INSERT THE DEVICE WERE NOT MADE. THERE WAS NO REPORTED PATIENT INJURY. HOWEVER, THE ACCESS SITE WAS LOST.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6810362 ↗
  • MalfunctionAug 16, 2017Sticking; Physical Resistance

    IT WAS REPORTED THAT THE DEVICE BECAME STUCK IN A 6FR COOK ANSEL SHEATH DURING ADVANCEMENT INTO THE LESION. THE DOCTOR PREPPED THE DEVICE. THERE WERE NO ANOMALIES NOTED DURING THE PREP OF THE DEVICE. THERE WAS NO DAMAGE NOTED TO THE BOX, TRAY OR STENT COVER AND NO DIFFICULTY REMOVING THE STENT COVER. THE PATIENT WAS BEING TREATED FOR SMA STENOSIS MESENTERIC ISCHEMIA. A 6F, 45 CM ANSEL 2 INTRODUCER SHEATH AND A 0.035, 260 CM ROSEN GUIDEWIRE WERE USED. THE DEVICE WAS ATTEMPTED TO BE INSERTED THROUGH THE 6F SHEATH HOWEVER RESISTANCE WAS FELT JUST PAST THE VALVE OF THE SHEATH. AN ATTEMPT WAS MADE TO RETRACT THE DEVICE FROM THE SHEATH HOWEVER THIS WAS UNSUCCESSFUL AS THE STENT WAS STUCK. THE STENT REMAINED STUCK IN THE INTRODUCER SHEATH. AIR EVACUATION WAS NOT PERFORMED PRIOR TO INSERTING THE DEVICE. MULTIPLE ATTEMPTS TO INSERT THE DEVICE WERE NOT MADE. THERE WAS NO REPORTED PATIENT INJURY. HOWEVER, THE ACCESS SITE WAS LOST.

    Patient problems: Ischemia

    MAUDE report 6797689 ↗

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page