Skip to content

LifeStream LSMU0800716Recalls & reported problems

Clearstream Technologies Limited · LifeStream™ Balloon Expandable Vascular Covered Stent, 7 mm x 16 mm, 80 cm catheter length · Class 3 · Implant

All models of LifeStream (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05391522081505), received between Nov 7, 2018 and Mar 5, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated3

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2026Device Dislodged or Dislocated

    A PATIENT PREPPED FOR A STENT PLACEMENT PROCEDURE BY USING LIFESTREAM. PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT THE HCP TOOK OFF THE COVER OF THE BALLOON, AND THE STENT ALLEGEDLY DISLODGED FROM THE BALLOON. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24515931 ↗
  • MalfunctionSep 10, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING PREPARATION FOR A COVERED STENT DEPLOYMENT PROCEDURE, THE BALLOON EXPANDABLE COVERED STENT ALLEGEDLY DISLODGED FROM THE BALLOON. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8984016 ↗
  • MalfunctionNov 7, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING POSITIONING OF THE BALLOON EXPANDABLE COVERED STENT IN THE CELIAC ARTERY, THE HEALTH CARE PROVIDER (HCP) DID NOT WANT TO DEPLOY THE COVERED STENT THROUGH THE TIGHT STENOSIS OF THE LESION WITHOUT THE PROTECTION OF THE SHEATH; THEREFORE, THE HCP RETRACTED THE PARTIALLY EXPOSED COVERED STENT BACK INTO THE INTRODUCER SHEATH AND PULLED THE BALLOON CATHETER BACK TO THE HEMOSTATIC VALVE OF THE SHEATH. AT THIS POINT, THE BALLOON CATHETER WAS COMPLETELY REMOVED AND THE HCP ALLEGEDLY IDENTIFIED THAT THE COVERED STENT WAS NO LONGER ATTACHED TO THE BALLOON CATHETER. FLUOROSCOPY WAS USED TO LOCATE THE COVERED STENT AT THE VALVE OF THE INTRODUCER SHEATH. IT WAS FURTHER REPORTED THAT THE SHEATH WAS REMOVED AND REPLACED WITH ANOTHER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8047007 ↗

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page