Skip to content

Celt ACD PLUS Vascular Closure Device 7F CLT-07Recalls & reported problems

Vasorum Limited · Celt ACD PLUS Vascular Closure Device 7F (US) · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05391530280679), received between May 22, 2025 and Jun 11, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Malposition of Device1
  • Off-Label Use1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    A MEDWATCH REPORT FORM WAS RECEIVED WHICH STATED THAT THE 6F CELT ACD CLOSURE DEVICE WAS MAL-DEPLOYED AND INTRALUMINAL IN THE RIGHT COMMON FEMORAL ARTERY. THE MAL-DEPLOYED CLOSURE DEVICE WAS IMMEDIATELY RETRIEVED SUCCESSFULLY. AN UPDATE WAS RECEIVED ON (B)(6) 2025 INFORMING THAT THE EMBOLIZED IMPLANT WAS RETRIEVED SUCCESSFULLY BY A SNARE. ATTEMPTS TO OBTAIN FINAL PATIENT STATUS WERE UNSUCCESSFUL.

    Patient problems: Insufficient Information

    MAUDE report 22183532 ↗
  • InjuryMay 22, 2025Off-Label Use; Malposition of Device; Off-Label Use; Malposition of Device

    IT WAS REPORTED THAT PHYSICIAN USED A 7F CELT ACD PLUS DEVICE TO CLOSE AN 8F ARTERIOTOMY AFTER A 4-HOUR CORONARY CTO PROCEDURE. THE DEVICE WAS RETRACTED TO WHAT WAS THOUGHT TO BE THE ARTERIOTOMY AFTER THE DISTAL WING WAS DEPLOYED. UPON RAISING TO 90 DEGREES THE BLOOD SIGNAL DID NOT COMPLETELY STOP SO PHYSICIAN MADE MULTIPLE ATTEMPTS TO RETRACT FURTHER AND ELEVATE AGAIN. THE BLOOD SIGNAL APPEARED TO STOP. UPON DEPLOYMENT OF THE PROXIMAL WING THE DEVICE APPEARED TO JUMP UPWARD SLIGHTLY. IMPLANT WAS EJECTED, BLEEDING WAS NOTED AND MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS FOR APPROXIMATELY 1 HOUR. THE PHYSICIAN ELECTED TO GIVE A BLOOD TRANSFUSION SINCE THE PATIENT WAS BECOMING HYPOTENSIVE, AND THE PATIENT WAS THEN REASSESSED AND AN ATTEMPT TO ACHIEVE HAEMOSTATIC CONTROL WITH AN ANGIOPLASTY BALLOON WAS NOT SUCCESSFUL. PATIENT WAS THEN BROUGHT TO SURGERY TO REPAIR THE ARTERIOTOMY. PATIENT WAS REPORTED TO BE OK AND SURGERY SUCCESSFUL.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22068514 ↗

Other Device, Hemostasis, Vascular devices

All Device, Hemostasis, Vascular devices →