Skip to content

Fast-Cath™ 406104Recalls & reported problems

St. Jude Medical, Inc. · Class 2

All models of Fast-Cath (71)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05414734008242), received between Jul 19, 2019 and May 14, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Fracture1
  • Material Separation1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN ANGIOGRAPHY PROCEDURE, A BLOOD LEAK WAS NOTED IN THE SHEATH WHICH RESULTING A HEMORRHAGE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22021283 ↗
  • InjuryDec 12, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT DURING A HEART CATHETERIZATION PROCEDURE, THE SIDEPORT TUBING BECAME DETACHED RESULTING IN MINOR BLEEDING. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20912522 ↗
  • InjuryOct 11, 2019Fracture; Fracture

    DURING THE PROCEDURE, THE SHEATH BROKE OFF FROM THE HUB INSIDE THE PATIENT'S FEMORAL ARTERY. THE DEVICE WAS RETRIEVED USING A SNARE. THE PATIENT WAS STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9179495 ↗
  • MalfunctionJul 19, 2019Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN INFROPOPLITEAL ANGIOPLASTY, THERE WAS BLOOD LEAKING OUT OF THE INTRODUCER. THE PROCEDURE WAS COMPLETED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8808251 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page