Skip to content

Fast-Cath™ 406120Recalls & reported problems

St. Jude Medical, Inc. · Class 2

All models of Fast-Cath (71)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05414734008266), received between Aug 8, 2020 and Jun 13, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Separation2
  • Fluid/Blood Leak1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    RELATED MANUFACTURING REF: 3005334138-2023-00265. UPON REMOVAL OF THE HEMOSTASIS INTRODUCER, THE VALVE TO THE INTRODUCER DID NOT SEAL CLOSE AS EXPECTED WHICH ALLOWED BLEEDING FROM THE INTRODUCER. THE HEMOSTASIS INTRODUCER WAS IMMEDIATELY EXCHANGED FOR A SECOND INTRODUCER OF THE SAME MODEL AND LOT. THE SECOND INTRODUCER VALVE WOULD NOT SEAL EITHER, ALLOWING FOR BLEEDING. THE SECOND SHEATH WAS THEN EXCHANGED FOR A NON ABBOTT INTRODUCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17118395 ↗
  • MalfunctionFeb 24, 2022Material Separation; Material Separation

    DURING THE PROCEDURE, THE SHEATH BROKE UPON REMOVAL FROM THE PATIENT AND APPROXIMATELY HALF THE SHEATH REMAINED IN THE THE VEIN (WITH A SMALL PORTION STICKING OUT). A CUT DOWN PROCEDURE WAS PERFORMED AND THE DETACHED PORTION WAS RETRIEVED WITH NO ADVERSE EVENT TO THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 13592395 ↗
  • InjuryAug 8, 2020Material Separation; Material Separation

    DURING THE TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE, THE SHEATH SPLIT INSIDE THE PATIENT'S ANATOMY. THE DEVICE WAS REMOVED OVER A WIRE. THE PATIENT EXPERIENCED A CARDIAC TAMPONADE AND A PERICARDIOCENTESIS WAS PERFORMED TO STABILIZE THE PATIENT. THE PROCEDURE WAS COMPLETED USING A LARGER SHEATH.

    Patient problems: Perforation

    MAUDE report 10382995 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →