ULTIMUM™ 407845Recalls & reported problems
St. Jude Medical, Inc. · HEMOSTASIS INTRODUCER6F-ACT · Class 2
All models of ULTIMUM (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 05414734011075), received between Sep 18, 2018 and Apr 2, 2025 — classified as 5 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Adverse Event Without Identified Device or Use Problem2
- Material Puncture/Hole2
- Material Separation2
- Dull, Blunt1
- Material Twisted/Bent1
- Patient Device Interaction Problem1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient2
- Stroke/CVA2
- Cardiac Perforation1
- Foreign Body In Patient1
- Hemorrhage/Blood Loss/Bleeding1
- No Known Impact Or Consequence To Patient1
- Thrombosis/Thrombus1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 2, 2025Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, A BLOOD LEAK OCCURRED. THERE WERE NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21751442 ↗InjuryJun 10, 2024Material Puncture/Hole; Material Puncture/Hole
DURING PAROXYSMAL PROCEDURE, THE STYLET BROKE THE PURGE LINE WHICH RESULTED IN PATIENT BLEEDING. THE DEVICE WAS EXCHANGED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19504734 ↗MalfunctionMar 11, 2024Material Puncture/Hole; Material Puncture/Hole
DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, IT WAS NOTED THAT THERE WAS A HOLE IN THE SHEATH. THE SHEATH WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. AT THE END OF THE PROCEDURE, EXOSEAL WAS ABOUT TO BE PERFORMED WHEN THE CLOSURE DEVICE WOULD NOT ADVANCE. THE SHEATH WAS EXCHANGED AND UPON REMOVAL A HOLE WAS OBSERVED. IT IS UNKNOWN IF THIS OCCURRED WHEN THE EXOSEAL WAS BEING INTRODUCED OR AT SOME OTHER POINT. THE PROCEDURE WAS COMPLETED WITH THE REPLACEMENT INTRODUCER WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. IT IS NOTED THAT THE AFFECTED SHEATH INSERTED WITHOUT ISSUE AND THE RIGHT COMMON FEMORAL ARTERY ANGIOGRAM LOOKED GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18874162 ↗MalfunctionAug 23, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
DURING A RIDE-SIDED PROCEDURE, THERE WAS A BLOOD CLOT NOTED IN THE SHEATH. THE SHEATH AND DILATOR WERE PREPPED, AND THE SIDE PORT WAS FLUSHED. THE SHEATH AND CATHETER WERE PLACED IN THE FEMORAL VEIN, BUT THE SHEATH WAS NOT CONSTANTLY IRRIGATED WITH HEPARINIZED SALINE AS ROUTINE PRACTICE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES. THERE IS NO FURTHER INFORMATION AVAILABLE.
Patient problems: Thrombosis/Thrombus
MAUDE report 17607971 ↗InjuryJun 5, 2023Material Separation; Material Separation
DURING A PAROXYSMAL ATRIAL FIBRILLATION PROCEDURE, A PORTION OF THE SHEATH REMAINED IN THE PATIENT AND AN EFFUSION WAS NOTED. AT THE END OF THE PROCEDURE, WHEN THE SHEATHS WERE BEING REMOVED, A PORTION OF ONE SHEATH REMAINED IN THE PATIENT'S RIGHT EXTERNAL ILIAC VEIN. THE PATIENT ALSO EXPERIENCED A PERICARDIAL EFFUSION. AN X-RAY IMAGE WAS TAKEN TO CONFIRM WHERE THE RETAINED SHEATH WAS, HOWEVER ATTEMPTS TO RETRIEVE IT WERE UNSUCCESSFUL AND MANUAL COMPRESSION WAS ADMINISTERED.
Patient problems: Cardiac Perforation; Foreign Body In Patient
MAUDE report 17067088 ↗MalfunctionFeb 5, 2023Fluid/Blood Leak; Fluid/Blood Leak
RELATED MANUFACTURING REF: 3005334138-2023-00035. DURING THE LEFT HEART CATHETER PROCEDURE, THERE WAS A CONSTANT TRICKLE OF BLOOD OUT OF THE HEMOSTATIC VALVE. IT WAS ATTEMPTED TO REINSERT THE INNER STYLET TO SEE IF IT WOULD RESEAL WITHOUT SUCCESS. THE INTRODUCER WAS REPLACED, BUT THE BLEEDING WAS NOT SIGNIFICANT ENOUGH TO REQUIRE ANOTHER REPLACEMENT. THE PROCEDURE WAS COMPLETED AND THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16302103 ↗MalfunctionFeb 5, 2023Fluid/Blood Leak; Fluid/Blood Leak
RELATED MANUFACTURING REF: 3005334138-2023-00036. DURING THE LEFT HEART CATHETER PROCEDURE, THERE WAS A CONSTANT TRICKLE OF BLOOD OUT OF THE HEMOSTATIC VALVE. IT WAS ATTEMPTED TO REINSERT THE INNER STYLET TO SEE IF IT WOULD RESEAL WITHOUT SUCCESS. THE INTRODUCER WAS REPLACED, BUT THE BLEEDING WAS NOT SIGNIFICANT ENOUGH TO REQUIRE ANOTHER REPLACEMENT. THE PROCEDURE WAS COMPLETED AND THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16302102 ↗MalfunctionOct 4, 2022Fluid/Blood Leak; Fluid/Blood Leak
AFTER A DIAGNOSTIC PROCEDURE, THERE WAS BLEEDING FROM THE GREEN VALVE ON THE INTRODUCER. IT WAS NOT THOUGHT THAT THE VALVE WAS DEFECTIVE. THE PROCEDURE WAS COMPLETED WITHOUT DEVICE REPLACEMENT AND WITHOUT PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15541396 ↗MalfunctionApr 13, 2022Material Separation; Material Separation
DURING THE ATRIAL FIBRILLATION PROCEDURE, THE SIDE PORT/STOPCOCK DETACHED FROM THE SHEATH. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14102951 ↗MalfunctionJun 23, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING A CORONARY ANGIOGRAPHY FOR NSTEMI, A LEAKAGE AND REFLUX OF BLOOD OCCURRED FROM THE HEMOSTASIS VALVE OF THE INTRODUCER. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10184812 ↗InjuryAug 2, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURING REF: 3005334138-2019-00425, 3005334138-2019-00426. FOLLOWING AN ATRIAL FIBRILLATION PROCEDURE, THE PATIENT EXPERIENCED A STROKE. FOLLOWING THE PROCEDURE, THE PATIENT WAS OBSERVED IN THE HOSPITAL FOR 8 DAYS AND THEN DISCHARGED. A FEW DAYS LATER, THE PATIENT PRESENTED TO THE HOSPITAL FOR A STROKE. THE PATIENT WAS ALERT ON (B)(6) 2019. PARALYSIS OF ONE SIDE OF THE BODY SEEMED TO BE REGRESSING. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.
Patient problems: Stroke/CVA
MAUDE report 8854749 ↗InjuryAug 2, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURING REF: 3005334138-2019-00425, 3005334138-2019-00427 FOLLOWING AN ATRIAL FIBRILLATION PROCEDURE, THE PATIENT EXPERIENCED A STROKE. FOLLOWING THE PROCEDURE, THE PATIENT WAS OBSERVED IN THE HOSPITAL FOR 8 DAYS AND THEN DISCHARGED. A FEW DAYS LATER, THE PATIENT PRESENTED TO THE HOSPITAL FOR A STROKE. THE PATIENT WAS ALERT ON (B)(6) 2019. PARALYSIS OF ONE SIDE OF THE BODY SEEMED TO BE REGRESSING. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.
Patient problems: Stroke/CVA
MAUDE report 8854745 ↗MalfunctionApr 4, 2019Dull, Blunt; Positioning Problem
A ST JUDE ULTIMUM 6F SHEATH WAS OPENED FOR PLACEMENT IN THE PATIENT'S FEMORAL ARTERY; THE PATIENT WAS UNDERGOING CARDIAC CATHETERIZATION. THE INTRODUCER, INCLUDED IN THE PACKAGING WITH THE SHEATH, COULD NOT BE INTRODUCED INTO THE SHEATH (THROUGH THE PROXIMAL DIAPHRAGM AT THE HUB). UPON FURTHER INSPECTION, IT WAS NOTED THE DISTAL END WAS NOT TAPERED AS INTENDED BUT INSTEAD BLUNT. THE SHEATH WAS REMOVED FROM THE FIELD AND ANOTHER SHEATH WAS OPENED AND UTILIZED. THE INVENTORY, WITHIN THE CATH LAB, WAS INSPECTED AND ALL OTHER SHEATHS WITH THE LOT NUMBER OF THIS DEVICE APPEARED TO BE TAPERED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8481969 ↗InjurySep 18, 2018Material Twisted/Bent; Material Twisted/Bent
DURING INSERTION TO THE FEMORAL ARTERY, THE ULTIMUM HEMOSTASIS INTRODUCER BECAME KINKED. DUE TO THE KINK, THE WIRE WAS LOST IN THE ARTERY. BEFORE A NEW ATTEMPT, THE ARTERY HAD TO BE COMPRESSED AND THERE WAS A 15 MINUTE DELAY. AFTER COMPRESSION, A NEW PUNCTURE WAS MADE. IN ORDER TO AVOID ANOTHER KINK, 3 OTHER INTRODUCERS WERE USED TO DILATE THE ARTERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7887325 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page