Skip to content

Fast-Cath™ 406112Recalls & reported problems

St. Jude Medical, Inc. · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05414734013918), received between Sep 13, 2021 and Aug 13, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Material Separation2
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Insert1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Cardiac Perforation1
  • Foreign Body Embolism1
  • Foreign Body In Patient1
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 13, 2025Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem

    DURING A TAVI PROCEDURE IN THE RIGHT FEMORAL ARTERY, A PERFORATION OCCURRED REQUIRING A TRANSFUSION ALONG WITH FURTHER SURGICAL INTERVENTION.

    Patient problems: Cardiac Perforation; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22794378 ↗
  • InjuryApr 9, 2024Material Separation; Material Separation

    DURING THE PROCEDURE, THE TIP OF THE SHEATH WAS SHEARED OFF. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. THERE IS NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 19075476 ↗
  • InjuryAug 16, 2022Difficult to Insert; Difficult to Insert

    DURING THE PROCEDURE, THE DEVICE WAS INSERTED FROM THE RIGHT GROIN TO THE FEMORAL VEIN. FOLLOWING REMOVAL OF THE DILATOR, THE CANNULA WAS SWOLLEN IN TWO PLACES. ANOTHER DEVICE WAS USED AND INSERTED THROUGH AN ADDITIONAL PUNCTURE SITE TO COMPLETE THE PROCEDURE WITH NO CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15231872 ↗
  • MalfunctionSep 24, 2021Material Separation; Material Separation

    DURING AN ABLATION PROCEDURE, THE DISTAL PORTION THE SHEATH DETACHED FOLLOWING INSERTION INTO THE FEMORAL VEIN DURING THE ACCESS PROCESS. THE PIECE BROKE DUE TO A NEEDLE BEING INADVERTENTLY PUSHED THRU THE SIDE WALL OF THE SHEATH AND THEN PULLED. THE DETACHED PORTION WAS RETRIEVED USING AN ENDOMYOCARDIAL BIOPSY FORCEPS THRU A LARGER 14F SHEATH. A REPLACEMENT SHEATH WAS THEN USED TO COMPLETE THE PROCEDURE WITH NO CONSEQUENCES TO THE PATIENT.

    Patient problems: Foreign Body Embolism

    MAUDE report 12527147 ↗
  • MalfunctionSep 20, 2021Fluid/Blood Leak; Fluid/Blood Leak

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2021-00557. DURING PERIPHERAL PROCEDURE, A BLOOD LEAK OCCURRED. AFTER PLACING THE GUIDEWIRE, BLOOD WOULD LEAK FROM THE INTRODUCER. THE SHEATH WAS REPLACED BUT THE SAME ISSUE OCCURRED, HOWEVER, THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SECOND SHEATH WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12494730 ↗
  • MalfunctionSep 13, 2021Fluid/Blood Leak; Fluid/Blood Leak

    RELATED MANUFACTURER REPORT NUMBER: DURING PERIPHERAL PROCEDURE, A BLOOD LEAK OCCURRED. AFTER PLACING THE GUIDEWIRE, BLOOD WOULD LEAK FROM THE INTRODUCER. THE SHEATH WAS REPLACED BUT THE SAME ISSUE OCCURRED, HOWEVER, THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SECOND SHEATH WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12462960 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →