Fast-Cath™ 406123Recalls & reported problems
St. Jude Medical, Inc. · Class 2
All models of Fast-Cath (71)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05414734014038), received between Aug 17, 2018 and Feb 9, 2023 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Perforation1
- Material Puncture/Hole1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 9, 2023Material Puncture/Hole; Material Puncture/Hole
DURING THE PROCEDURE, IT WAS NOTED THAT THERE WERE SMALL HOLES IN THE SHEATH DURING FEMORAL VEIN ACCESS. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16337243 ↗MalfunctionJan 5, 2022Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, A BLOOD LEAK WAS NOTED FROM THE HEMOSTASIS VALVE WHEN THE NON ABBOTT CATHETER WAS EXCHANGED FOR AN ABLATOR CATHETER. THERE WERE NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13170045 ↗MalfunctionJan 29, 2019Material Perforation; Material Perforation
THE FAST-CATH HEMOSTASIS INTRODUCER WAS BENT DURING INSERTION IN THE FEMORAL VEIN. AN ABLATION CATHETER WAS INSERTED INTO THE SHEATH AND THE CATHETER PUNCTURED THROUGH THE SHEATH WALL. THE SHEATH WAS REMOVED AND PROCEDURE WAS COMPLETED BY PERFORMING AN ADDITIONAL FEMORAL PUNCTURE AND INSERTING AN SLO. THERE WERE NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8287870 ↗InjuryAug 17, 2018Material Separation; Material Separation
THE FAST-CATH HEMOSTASIS INTRODUCER WAS ADVANCED THROUGH THE RIGHT FEMORAL VEIN. WHEN THE OPERATOR TWISTED THE HEMOSTASIS VALVE ON THE INTRODUCER TO MOVE IT INTO PLACE, THE VALVE SHEARED OFF THE INTRODUCER JUST DISTALLY TO THE END OF THE VALVE, LEAVING THE REMAINDER OF THE INTRODUCER WITHIN THE VEIN. IN ORDER TO RETRIEVE THE REMAINDER OF THE INTRODUCER, A SMALL INCISION WAS MADE IN THE GROIN AND A SET OF SURGICAL FORCEPS WAS UTILIZED TO GAIN TRACTION OF THE INTRODUCER. THE REMAINDER WAS RETRIEVED, AND THE PROCEDURE WAS CONTINUED USING A NEW SHEATH. THE PATIENT INFORMATION IS NOT AVAILABLE DUE TO HOSPITAL PRIVACY POLICY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7793606 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page