Fast-Cath™ 406183Recalls & reported problems

St. Jude Medical, Inc. · Class 2

All models of Fast-Cath (71)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05414734014458), received between Aug 21, 2019 and Mar 19, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2
  • Fluid/Blood Leak1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2024Gas/Air Leak; Gas/Air Leak

    DURING A PERSISTENT ATRIAL FIBRILLATION PROCEDURE, AN AIR LEAK WAS NOTED. AFTER INSERTION INTO THE HEART BUT BEFORE INSERTING THE CATHETER, AIR WAS ASPIRATED INTO THE SYRINGE. AIR ASPIRATION WAS PERFORMED SEVERAL TIMES WITH NO RESOLUTION. THE SHEATH WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18935674 ↗
  • MalfunctionAug 11, 2022Gas/Air Leak; Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AIR WAS ASPIRATED WHEN THE SHEATH WAS INSERTED INTO THE PATIENT AND AIR WAS REMOVED. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15210625 ↗
  • MalfunctionJul 22, 2022Material Separation; Material Separation

    DURING AN ATRIAL FIBRILLATION PROCEDURE, THE JOINT BETWEEN THE SHEATH TUBE AND SHEATH HUB SEPARATED. UPON REMOVAL OF THE SHEATH FROM THE PATIENT AND ATTEMPTING TO PULL OUT THE SHEATH FROM THE RIGHT INGUINAL REGION, THE JOINT BETWEEN THE SHEATH TUBE AND SHEATH HUB SEPARATED. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15087004 ↗
  • MalfunctionAug 21, 2019Fluid/Blood Leak; Fluid/Blood Leak

    BLOOD WAS LEAKING FROM THE HEMOSTASIS VALVE DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE, PRIOR TO INSERTION OF CATHETERS. THE SHEATH SET WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8914890 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →