ULTIMUM™ 407687Recalls & reported problems
St. Jude Medical, Inc. · HEMOSTASIS INTRODUCER · Class 2
All models of ULTIMUM (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05414734020930), received between Aug 18, 2020 and Feb 26, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak7
Patient problems
- No Consequences Or Impact To Patient6
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2026Fluid/Blood Leak
DURING A TAVI PROCEDURE, WHILE ACCESSING THE FEMORAL ARTERY USING THE INTRODUCER AND A NON-ABBOTT GUIDEWIRE (LUNDERQUIST BY COOK MEDICAL), BLOOD LEAKED FROM THE HEMOSTASIS VALVE RESULTING IN SIGNIFICANT BLOOD LOSS. MANUAL COMPRESSION WITH GAUZE WAS HELD AT THE VALVE OF THE INTRODUCER. THE BLEEDING RESULTED IN A MARKED DECREASE IN THE PATIENT¿S HEMOGLOBIN LEVELS: ON (B)(6) 2026 HB: 13.8; ON (B)(6) 2026 HB: 11.9. NO INTERVENTION WAS ADMINISTERED TO TREAT THE DROP IN HEMOGLOBIN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24457559 ↗MalfunctionOct 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, THE HEMOSTATIC VALVE STARTED LEAKING ONCE THE WIRE WAS INSERTED. A 14F NON-ABBOTT SHEATH WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10623506 ↗MalfunctionOct 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, THE HEMOSTATIC VALVE OF THE SHEATH STARTED LEAKING AFTER TAKING OUT THE BALLOON FOR THE PRE-DILATION. THE PROCEDURE WAS COMPLETED WITH A NON-ABBOTT 14F SHEATH WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10623452 ↗MalfunctionSep 4, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING THE TRANSCATHETER AORTIC VALVE IMPLANTATION PROCEDURE, THE HEMOSTATIC VALVE STARTED LEAKING AFTER TAKING OUT THE BALLOON FOR THE PRE-DILATATION. THE PROCEDURE WAS COMPLETED WITH THE SAME SHEATH WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10494638 ↗MalfunctionSep 4, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, THE HEMOSTATIC VALVE STARTED LEAKING AFTER THE BALLOON WAS TAKEN OUT OF THE SHEATH. THE PROCEDURE WAS COMPLETED USING THE SAME SHEATH BY INSERTING A 6F SHEATH INSIDE THE 14F SHEATH TO STOP THE LEAK WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10494635 ↗MalfunctionSep 4, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, THE HEMOSTATIC VALVE STARTED LEAKING AFTER TAKING OUT THE BALLOON. THE PROCEDURE WAS COMPLETED WITH THE SAME SHEATH WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10494629 ↗MalfunctionAug 18, 2020Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, AFTER REMOVING THE BALLOON, THE VALVE OF THE SHEATH STARTED LEAKING. A NON-ABBOTT DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10423624 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.