Skip to content

FemoStop™ GoldRecalls & reported problems

St. Jude Medical, Inc. · Femoral Compression System · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05414734050036), received between Oct 17, 2019 and May 12, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method3
  • Inflation Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Battery Problem1
  • Component Missing1
  • Deflation Problem1
  • Device Displays Incorrect Message1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Bradycardia1
  • Hematoma1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING A PROGLIDE DEVICE AFTER A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE USING A 7F SHEATH. REPORTEDLY, THE KNOT DEPLOYED IN THE TISSUE TRACT AND NOT IN THE VESSEL. MANUAL COMPRESSION WAS APPLIED BUT HEMOSTASIS WAS NOT ACHIEVED. A FEMOSTOP GOLD WAS USED TO ACHIEVE HEMOSTASIS WHICH RESULTED IN A HEMATOMA AND BRUISING AT THE ACCESS SITE. THE SIZE OF THE HEMATOMA IS UNKNOWN. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Hematoma

    MAUDE report 25151290 ↗
  • MalfunctionMar 31, 2025Inflation Problem; Improper or Incorrect Procedure or Method; Inflation Problem; Improper or Incorrect Procedure or Method

    IT WAS REPORTED WHEN THE FEMOSTOP WAS APPLIED, THE PRESSURE GAUGE DOES NOT WORK AND THE DOME CANNOT INFLATE. THEREFORE, THE FEMOSTOP WAS REPLACED. THE DEVICE WAS USED PAST THE EXPIRATION DATE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21734160 ↗
  • MalfunctionOct 7, 2024Improper or Incorrect Procedure or Method; Battery Problem; Improper or Incorrect Procedure or Method; Battery Problem

    SUBSEQUENT TO THE INITIALLY FILED REPORT, THE FOLLOWING INFORMATION WAS RECEIVED: THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20384114 ↗
  • MalfunctionJul 12, 2024Improper or Incorrect Procedure or Method; Component Missing; Improper or Incorrect Procedure or Method; Component Missing

    IT WAS REPORTED THAT THE "STRAP" [BELT] WAS MISSING WHEN OPENING A FEMOSTOP GOLD. ADDITIONALLY, BASED ON THE LOT NUMBER, THE DEVICE WAS USED AFTER EXPIRATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19734162 ↗
  • InjuryOct 17, 2019Deflation Problem; Inflation Problem; Device Displays Incorrect Message; Deflation Problem; Inflation Problem; Device Displays Incorrect Message

    DURING THE PROCEDURE, THE PATIENT EXPERIENCED BRADYCARDIA. WHEN ATTEMPTING TO DEFLATE THE DOME, NO PRESSURE OR AIR WOULD RELEASE. THE DIGITAL MANOMETER READ 73 BUT THE DOME APPEARED FULLY INFLATED. THE DOME WAS MANUALLY PRESSED WHICH DEFLATED THE DOME. THE DOME WAS ATTEMPTED TO BE RE-INFLATED, BUT IT WOULD NOT EASILY RE-INFLATE. EVENTUALLY THE DOME DID RE-INFLATE BUT THE MANOMETER STILL DID NOT APPEAR TO BE WORKING CORRECTLY. THE PATIENT WAS BRADYCARDIC, ATROPINE WAS ADMINISTERED BUT INEFFECTIVE. SHORTLY AFTER THE AIR IN THE DOME WAS RELEASED, THE PATIENT'S HEART RATE RETURNED TO BASELINE. THERE WAS NO PATIENT HARM.

    Patient problems: Bradycardia

    MAUDE report 9203875 ↗

Other CLAMP, VASCULAR devices

All CLAMP, VASCULAR devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.